![YOKOGAWA WT5000 Скачать руководство пользователя страница 107](http://html1.mh-extra.com/html/yokogawa/wt5000/wt5000_features-manual_4140530107.webp)
11-1
IM WT5000-01EN
11 Graph Display
Graph Display (GRAPH)
Press GRAPH to display the waveforms of the following input signals.
•
•
•
•
Each time you press GRAPH, the screen switches, in order, between waveform display, trend display, bar graph
display, and vector display.
Waveform display (Wave)
The waveforms of the following input signals can be displayed.
• Input element’s voltage and current
• Motor evaluation function’s (option) speed, torque, Aux
Waveform Display Example
Time at the right edge of the screen
(amount of time contained in the
screen)
Time at the left edge
of the screen
(fixed at 0 seconds)
• The number of data points between the left and right edges of the display.
• “P-P” indicates that the displayed waveform has been P-P compressed.
Scale value
Voltage or current, element, and upper limit of the displayed waveform
Scale value
Voltage or current, element, and lower limit of the displayed waveform
Trigger level
Ground level
Waveform label
U1 3.000kV
U1
−
3.000kV
0.000s
50.000ms
<< 2002 (p
−p)
>>
U1