Appendix 1 How to Make Accurate Measurements
Because the WT310/WT310HC/WT330 measures high frequencies, the effects of i
C
s cannot be
ignored.
If you connect the WT310/WT310HC/WT330 current input terminal to the side of the power supply
(SOURCE) that is close to its earth potential, the WT310/WT310HC/WT330 current measurement
circuit positive and negative terminals are close to the earth potential, so V
C
s becomes approximately
zero and very little i
C
s flows. This reduces the effect on measurement accuracy.