Confirming and Optimizing the Maximum Capturing Count
(Optimize Count)
Press the
Optimize Count
soft key to display the following screen.
Maximum capturing count (0 to the maximum number of captures)
* The maximum number of times that data can be captured depends on the number of numeric
data items that you have set to be saved and the free space at the save destination.
For instructions on how to set the save destination and the numeric data items to be saved, see
section 15.2.
Even if you have specified a USB memory device as the save destination, if you remove the USB
memory device, the save destination switches automatically to the internal RAM disk. If you close
this screen and then open it again by pressing the Optimize Count soft key, the maximum
capturing count changes to the value determined by the internal RAM disk,s free space.
Set the number of data captures.
The capturing count is set to the maximum capturing count
displayed to the left.
Configuring Capture Control Settings (Control Settings)
Press the
Control Settings
soft key to display the following menu.
Turn external signal synchronization on or off.
• OFF: This instrument captures data every 5 ms.
• ON: This instrument captures data in sync with an external signal.
Configure trigger settings.
Configure the voltage and current measurement modes.
Set the HS Filter.
• Turn the HS filter on or off.
• Set the cutoff frequency (1 Hz to 1000 Hz in steps of 1 Hz)
Configuring the Voltage and Current Measurement Modes
Press the
U/I Measuring Mode
soft key to display the following menu.
Set the measurement mode (rms, mean, dc, r-mean).
Select the measurement mode setup method (Each,
All).
When you select Each, you can set the measurement
mode separately for the current and voltage of each input
element.
Note
If the voltage and current measurement mode settings differ for elements assigned to the same wiring unit,
the measurement data (Σ function) for the wiring unit is displayed as “-------” (no data).