<9 DATA FILES AND FORMAT> 9-7
IM 11Y01B02-01E-A 3rd Edition June 19, 2012-00
Determine whether or not the analyzer should capture spectrum
files under a
WARNING
condition.
Note: this may be useful to do so however, if the Warning alarm
conditions are not set correctly then there could be excessive files
created for less meaningful Warning alarm conditions.
An example is a low transmission warning alarm set at 70% for an
application that often runs at less than 70% transmission.
Determine whether or not the analyzer should capture spectrum
files under a
FAULT
condition.
Note: this may be useful to do so however, if the Fault alarm
conditions are not set correctly, then there could be excessive files
created for less meaningful Fault alarm conditions.
These files are often useful for Factory based diagnostics.