<2. INSTALLATION, PIPING, AND WIRING>
2-14
IM 12B01B30-02EN
1st Edition : Mar. 08, 2023-00
n
Power supply cable (E)
If a Terminal box is included, wire it to the Terminal box If there is no Relay box, wire it to the
Relay box.
Refer to Figure 2.14 for wiring. Cables for power supply and grounding must be provided by the
customer. Refer to “2.2.4 Power supply cable (E)”
for the specifications of the cable and the
requirements for power supply and grounding wiring.
n
Relay box control cable (F)
Dedicated cables are included. Connect to the specified connectors referring to
n
Solenoid valve cable (G)
Solenoid valve and its valve cabling must be provided by the customer. Refer to Table 2.2 and
for required specifications. Wire solenoid valve cabling referring to
Figure 2.16. Use a solenoid valve that matches the required power supply voltage.
Table 2.2
Solenoid valve Requirements
Port
Valve type
Max. working
pressure
Max. working
pressure difference
SV1 (Measure)
15A
Motor Valve 3 Terminals
(Open
,
Shut
,
Common)
1 MPa
1 MPa
SV2 (Wash)
RC3/8
Solenoid valve NOOC.
(*1)
1.5 MPa
0.5 MPa
SV3 (Zero)
RC3/8
Solenoid valve NOOC.
(*1)
1.5 MPa
0.5 MPa
SV4 (Drain)
Ø21 × Ø18
Silicone tube
Pinch valve N.C. (*1)
―
―
*1:
The valve opens when energized.
Table 2.3
Solenoid valve Cable Requirements (Recommended Cables)
Cable length
1.2 m
Rated voltage
300 V AC
Specifications
SV1
AWG 17 3 core cable, M3 round terminal
SV2, SV3, SV4
AWG 20 2 core cable, M3 round terminal
2.3
Desiccant mounting
High humidity in detector internal storage may affect the measurement. Desiccants must be
installed on detector internal storage for correct readings.
l
With sampling system (-A1, -A2, -A3)
The desiccant is shipped pre-installed on detector internal storage, so you do not need to install
the desiccant when starting operation.
l
Without sampling system (-NN, -A5)
Desiccant must be attached to detector internal storage. Perform mounting of the desiccant
according to the following.
To prevent electric shock, turn off the power supply of the detector before turning mounting the
desiccant.