<2. INSTALLATION, PIPING, AND WIRING>
2-13
IM 12B01B30-02EN
1st Edition : Mar. 08, 2023-00
WARNING
Never let the power cables of TB830D (Figure 2.15) touch the Cable for Relay box control
(F), SENCOM cable for turbidity/chlorine (A), resulting in failure to comply with the safety
requirements.
WARNING
The wiring must meet the IP65 or higher. The four screws of detector cover are tightened to 1.5
N·m.
Ground Terminal
101
102
201
202
301
302
Solenoid valve cable (G)
FLXA402T power supply cable (B)
TB830D power supply cable (C)
View from the bottom
Relay box
control cable (F)
Relay Box power supply cable (D)
or power supply cable (E)
SV3 cable
SV2 cable
SV1 cable
SV4 cable
Figure 2.16
Wiring of Relay box (for-A5)
n
SENCOM cable for Turbidity/Chlorine (A)
Dedicated cables are included. Refer to Figure 2.14 and
for wiring. Refer to Section
2.6.1 of FLXA402T Installation Manual (IM12A01G01-02EN) for the wiring on FLXA402T.
n
FLXA402T power supply cable (B)
Dedicated cables are included. Refer to Figure 2.14 and Figure 2.16 for wiring. The lengths of
the terminal processes at both ends are different. Connect the longer side (approx. 80 mm) to
TB830D side and the shorter side (approx. 50 mm) to FLXA402T side. Cables have no terminal
names on them. Wire them by color. L:Black, N: White, G: Red. The tightening torque of M3
screws (L and N) is 0.6 N·m, and that for M4 screws (G) is 1.4 N·m.
n
TB830D power supply cable (C)
power supply cabling that connects the Relay box to TB830D.
Dedicated cables are included. Refer to Figure 2.14,
, and Figure 2.16 for wiring.
Cables have no terminal names on them. Wire them by color. L:Black, N: White, G: Red. M3 The
tightening torque for screws (L and N) is 0.6 N·m, and that for M4 screws (G) is 1.4 N·m .
n
Relay Box power supply cable (D)
This is a power supply cable that connects a Relay box to another Relay box when a Relay box is
attached.
Dedicated cables are included. Refer to Figure 2.14 and Figure 2.16 for wiring. Cables have
no terminal names on them. Wire them by color. L:Black, N: White, G: Red. M3 The tightening
torque for screws (L and N) is 0.6 N·m, and that for M4 screws (G) is 1.4 N·m .