IM 12D08B02-01E
Parameter setting 5-15
Code Display
Function
Function detail
X
Y
Z
Default values
mA Outputs
30
Not
used
31
*OUTP.F
mA output functions
Linear
0
0 Linear
Table
1
32
*BURN
Burn function
No burnout
0
0 No Burn.
Burnout
downscale
1
Burnout
upscale
2
Pulse
burnout
3
33, 34
Not used
35
*TABLE
Output table for mA
*0%
Linearization table for mA in 5% steps.
*5%
The measured value is set in the main
*10%
display using the >, ^, ENT keys, for
...
each of the 5% interval steps.
...
Where a value is not known, that value
*95%
may be skipped, and a linear interpolation
*100%
will take place.
36-39
Not
used
mA