IM 12D08B02-01E
Parameter setting 5-13
Code Display
Function
Function
detail
X
Y
Z
Default
values
Temperature compensation functions
20
*T.R.°C
Set reference temp.
Use >, ^, ENT keys to set value
25 °C
21
*T.C.1
Set temp. coef. 1
Adjust compensation factor
2.1 %
if set to TC in section 5-2-5.
per °C
Set value with >, ^, ENT keys
*T.C.2
Set temp. coef. 2
Adjust compensation factor
2.1 %
if set to TC in section 5-2-5.
per °C
Set value with >, ^, ENT keys
22
*MATRX
Select matrix
Choose matrix if set to matrix comp.
in section 5-2-5, using >, ^, ENT keys
HCl (cation) pure water (0-80 °C)
1
1
HCI
Ammonia pure water (0-80 °C)
2
Morpholine pure water (0-80 °C)
3
HCl (0-5 %, 0-60 °C)
4
NaOH (0-5 %, 0-100 °C)
5
User programmable matrix
9
23
*T1 °C (°F) Set temp. range
Enter 1st (lowest) matrix temp. value
*T2..
Enter 2nd matrix temp. value
*T3..
Enter 3rd matrix temp. value
*T4..
Enter 4th matrix temp. value
*T5..
Enter 5th (highest) matrix temp. value
24
*L1xT1
Enter conductivity
Value for T1
*L1xT2
values for lowest
Value for T2
....
concentration
*L1xT5
Value for T5
25
*L2xT1
Concentration 2
Similar to code 24
26
*L3xT1
Concentration 3
Similar to code 24
27
*L4xT1
Concentration 4
Similar to code 24
28
*L5xT1
Concentration 5
Similar to code 24
29
Not
used
Содержание SC202G
Страница 8: ...IM 12D08B02 01E LT LV PL EST SLO H BG RO M CZ SK...
Страница 13: ...IM 12D08B02 01E 1 4 Introduction...
Страница 33: ...IM 12D08B02 01E 2 20 Specifications...
Страница 74: ...IM 12D08B02 01E Spare Parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Страница 92: ...IM 12D08B02 01E 3 4 Appendix...
Страница 96: ...IM 12D08B02 01E 3 8 Appendix...
Страница 99: ...IM 12D08B02 01E Appendix 3 11 3 3 QIS 12D08B02 61E Figure 1 Testing Circuit and Test Equipment...
Страница 100: ...IM 12D08B02 01E 3 12 Appendix...
Страница 104: ...IM 12D08B02 01E 3 16 Appendix...