5-9. Input Contact Setup
General
Remote range switching to 10 times the
programmed range.
Factor 10 mA1 and/or mA2
This "ZOOM" function will cause the range
of the mA to be multiplied by 10 when the
binary input is closed. The range is divided
by 10 when the binary contact is opened.
5-10. Error Configuration
Errors 1/3 ~ 3/3
Error message configuration. Two different
types of failure mode can be set.
Hard fail gives a steady FAIL flag in the
display, and a continuous contact closure.
All the other contacts (controls) are
inhibited and a Fail signal is transmitted on
the outputs when enabled.
Soft fail gives a flashing FAIL flag in the
display, and the relay contact is pulsed.
The other contact (controls) is still
functional, and the controller continues to
work normally. A good example is a control
time-out for a soft fail. A warning that the
regular maintenance is due, should not be
used to shut down the whole
measurement.
5-11. Logbook Configuration
General
Logbook is available to keep an electronic
record of events such as error messages,
calibrations and programmed data
changes. By reference to this log, users
can for instance easily determine
maintenance or replacement schedules.
In configure Logbook the user can select
each item he is interested in to be logged
when the event occurs. This can be done
for three separate logbooks. Each logbook
can be erased individual.
IM 12D7B4-E-H
Содержание SC150
Страница 2: ......
Страница 20: ...IM 12D7B4 E H 14...
Страница 21: ...IM 12D7B4 E H 15 5 MENU STRUCTURE COMMISSIONING...
Страница 35: ...IM 12D7B4 E H 29 Time Span x 1 hour Temperature Concentration...
Страница 49: ...IM 12D7B4 E H 43...
Страница 50: ...IM 12D7B4 E H 44...
Страница 56: ...QIS 12D7B4 E H 50...