IM 12B7K1-02E
<
2. Installation >
2-3
2.2.2
Installing Sensor in PH8HS Submersion Holder
To install the sensor in the submersion holder, proceed as follows:
(1) Pass the sensor cable through the sensor holder. If the submersion holder remains installed,
remove the sensor holder in any case.
For a pipe-mounting submersion holder without a cleaner, loosen the sensor holder nut to
remove the holder.
F2.4.ai
Right-angled Pipe Clamp
Sensor Holder
Sensor Holder
Sensor Holder Nut
Arm Pipe
Removal of Sensor Holder (Arm Pipe used with option /MS1 or /MS2)
Removal of Sensor Holder (Stainless Bracket used with option /MS3 or /MS4)
Bracket
Nut
Figure 2.4
Removal of Sensor Holder (for Pipe Mounting Sensor without Cleaner)
For a flange mounting submersion holder without a cleaner, remove the sensor holder by
loosening the two bolts securing the sensor holder to the flange (see Figure 2.5).
F2.5.ai
Flange
Fixing Bolt
Figure 2.5
Removal of Sensor Holder (for Flange Mounting Sensor without Cleaners)
11th Edition : Dec.14,2018-00