background image

 

IM MY40-S01E 

 

2nd Edition:Feb.2009  (YMI)

 

MY10, MY40 

Insulation Resistance Tester 

User’s  

Supplements 

Manual  

 

For EMC Standards 

 

This paper supplements the User’s Manual of MY10 and MY40 Insulation Resistance Tester. 

This is to advise the additional explanations for EMC standards and the update postal address   

for Yokogawa Europe.

 

 
 

Model 

User’s Manual 

Description  

before the revision 

Revised Content 

MY10 

 

IM MY10-E 
6th Edition

 

EMC standard 

EN55011 Class B Group 1 

 EN61326 Class B 
 
 
 
Effect of radiation immunity 
At the strength of    radio-frequency 
electromagnetic field of 3V/m 
•Insulation resistance measurement 
  1st effective measuring range: ±10% of rdg 

2nd effective measuring range: ±20% of rdg 

Infinity and zero indications:   

±1.4% or less of scale length 

•AC Voltage measurement: ±10% of    FS 
 

EMC standards: 
Compliant with 

EN55011 Class B Group 1, 
EN61326-1 Class B,   
EN61326-2-2 

 
Effect of radiation immunity 
At the strength of    radio-frequency 
electromagnetic field of 3V/m 
•Insulation resistance measurement 
  1st effective measuring range: ±10% of rdg 

2nd effective measuring range: ±20% of rdg 
Infinity and zero indications:   

±1.4% or less of scale length 

•AC Voltage measurement: ±10% of    F.S. 

MY40 

 

IM MY40-E 
6th Edition

 

EMC standard 

EN55011 Class B Group 1 

 EN61326 Class B 
 
 
 
Effect of radiation immunity 
At the strength of    radio-frequency 
electromagnetic field of 3V/m 
•Insulation resistance measurement 
    1st effective measuring range:   

±(5% of rdg+12dgt) 

2nd effective measuring range:   

±(10% of rdg+12dgt) 

•AC Voltage measurement:   

±(5% of rdg+12dgt) 

•Conductor resistance measurement: 

±(10% of range) 

 

EMC standards: 
Compliant with 

EN55011 Class B Group 1, 
EN61326-1 Class B,   
EN61326-2-2 

 
Effect of radiation immunity 
At the strength of    radio-frequency 
electromagnetic field of 3V/m 
•Insulation resistance measurement 
    1st effective measuring range:   

±(5% of rdg+12dgt) 

2nd effective measuring range:   

±(10% of rdg+12dgt) 

•AC Voltage measurement:   

±(5% of rdg+12dgt) 

•Conductor resistance measurement: 

±(10% of range) 

 

 
 
 
 
 
 
 

YOKOGAWA EUROPE B. V. (THE NETHERLANDS) 

Euroweg 2, 3225HD, Amersfoort THE NETHERLANDS 

Phone: 31-88-4641000    Facsimile: 31-88-4641111 

 

Отзывы: