App-10
IM MV1000-01E
Manually Sampled Data
• The MV generates manually sampled data to a tab-separated text file that contains
numeric values and text.
• The MV does not generate values for measurement channels set to Skip or for
computation or external input channels set to Off.
• Data is added to the file each time you carry out a manual sampling operation.
Format
YREC
CRLF
Manual Sample Data Version 1.00.00
CRLF
Model
MV2000
CRLF
Language Code
shift-JIS
CRLF
File Status
ffffffff
CRLF
Serial No.
III•••I
CRLF
File Header
HHH•••H
CRLF
Ch
ccccc
ccccc
•••
ccccc
CRLF
Tag
ttt•••t
ttt•••t
•••
ttt•••t
CRLF
Unit
uuuuuu
uuuuuu
•••
uuuuuu
CRLF
yyyy/mo/dd hh:mi:ss nnn•••n
nnn•••n
•••
nnn•••n
CRLF
ffffffff
File status (eight characters)
Complete
Completed
Progress
Data being added
Decrease
Defective
III•••I
MV serial number (16 characters)
HHH•••H
File header (50 characters)
ccccc
Channel number (five characters)
ttt•••t
Tag name (16 characters)
uuuuuu
Unit (six characters)
yyyy/mo/dd hh:mi:ss
Year, month, day, and time of sampling (19 characters)
nnn•••n
Measured value (13 characters)
Appendx 3 Text Fle Formats