YOKOGAWA EXAxt ZR22G Скачать руководство пользователя страница 2

 

 

 

Tech Note: Unreleased 

 

 

 

 

 

 

 

Date: October 22, 2013 

may become damaged. Once the cell is removed, the nut (8) which holds the 
heater in place becomes visible.  
 

 
Heater Strut Removal

 

 
Open the terminal cover on the head of the probe by loosening the set screw 
using an allen wrench and then turn the cover counter-clockwise. Remove the 
positive cell wire and both thermocouple wires from the top terminal block (26) 
using a Philips-head screwdriver to loosen the screws (31). The top terminal block 
can now be removed by loosening the screws (28) in the upper-left and lower-
right corners, being careful not to lose the spacers (29) between the screws and 
terminal block. The two heater wires may now be removed using the Philips-
head screwdriver. Next, loosen the two screws (15) that hold the bolt cover (12) 
to the air connector block (13) and slide the cover back toward the flange. This 
will expose the four sets of bolts (10) and washers (11) which hold the terminal 
box to the air connector and probe. Remove (10) by turning them counter-
clockwise, taking care not to let the terminal box (16) fall. Gently slide the 
terminal box away from the probe without damaging the wiring. Loosen screw 
(19) to detach the air connector (13) from the heater strut (23).  Flip the probe 
over and using special service tool M1234HD-A, insert the tool’s raised end into 
the notches in the nut (8). Remove the nut by sticking a tool such as a 
screwdriver in the hole at the other end of the service tool and turn it counter-
clockwise. Once the nut is removed the heat strut (23) can be removed from 
the probe’s tube (24). The heater strut should be removed with care to not 
damage the calibration line. Remove the contactor spring (9), which is located 
in the groove near the back of where the heater nut was, with a pick or pair of 
tweezers. Now is a good time to label wiring if it is deemed necessary. 
 

 
Air Connector Servicing

 

 
Remove the translucent o-ring (14) from the air connector and discard it. Loosen 
the screw (25) holding the plate (17) to the air connector (13) by turning it 
counter-clockwise. Once the plate is removed the air caps (20) can be 
removed. Use a pick to remove o-ring (21) from the reference and calibration 
air caps (20). The other o-rings (22) are in the connector holes that the air caps 
were in. These can also be removed with a pick or by pushing them out from the 
back. Discard all old o-rings. Install an o-ring (22) in the groove in each air cap 
hole, ensuring it is seated at the bottom of the hole. Then install an o-ring (21) on 

Отзывы: