YOKOGAWA EXAxt ZR22G Скачать руководство пользователя страница 1

 

 

 

Tech Note: Unreleased 

 

 

 

 

 

 

 

Date: October 22, 2013 

Standard ZR22G Heater and Cell Replacement 

 

Necessary Parts to Order

 

 
ZR22A-###-A-A (Heater strut assembly with tool (M1234HD-A) provided) 
B1011ER (O-Ring for piping connector) 
B1013ER (O-Ring for piping connector to reference gas cap) 
B1013ER (O-Ring for piping connector to calibration gas cap)   
B1014ER (O-Ring for reference gas cap) 
B1014ER (O-Ring for calibration gas cap) 
E7042UD (If replacing the zirconia cell as well; otherwise only the K9470BJ metal 
o-ring, the E7042BS contacting spring, and E7042AY filter screen are needed) 
High-temperature grease 
 

Tools Needed 

 
#2 Philips-head screwdriver 
Pick set or tweezers (having both is optimal) 
8mm socket 
Torque wrench 
10mm combination wrench or an appropriately sized crescent wrench 
Allen wrench 
4.5mm pin spanner (if using the type one oil and gas dust filter) 
 

Verify the heater is off and has sufficient time to cool before attempting this 
procedure. Use Appendices A and B as a reference for this guide. 
 
Zirconia Cell Removal 
 

Remove (1) by twisting it counter-clockwise if you have type the type one filter; if 
you have type two keep reading. 
 
Remove the four bolts (2) and washers (3) on the heated end of the probe by 
turning them counter-clockwise with an 8mm socket wrench. Remove the 
contactor plate (4), calibration j-tube (5) and filter screen (6) from the tip of the 
probe. Remove the zirconia cell (6) by twisting it clockwise while simultaneously 
gently pulling it away from the tip of the probe. The metal o-ring (7) should fall 
out with the cell removed. You should not re-use this o-ring, otherwise the cell 