<2. DO402G Specifications>
2-4
IM 12J05D02-01E
2.2
Operating specifications
l
Performance : DO (at t process = 25 °C)
Linearity:
± 0.03 mg/L or ± 0.5%FS, whichever is greater
Repeatability: ± 0.03 mg/L or ± 0.5%FS, whichever is greater
Accuracy:
± 0.05 mg/L or ± 0.5%FS, whichever is greater
l
Performance : Temperature (Pt1000, PB36, 22kNTC)
Linearity:
± 0.3 ºC
Repeatability: ± 0.1 ºC
Accuracy:
± 0.3 ºC
l
Performance : Temperature (Pt100)
Linearity:
± 0.4 ºC
Repeatability: ± 0.1 ºC
Accuracy:
± 0.4 ºC
Note on performance specifications:
The specifications are expressed with simulated inputs, because the DO402G can be
used with many different sensors with their unique characteristics.
The following tolerance is added to above performance.
mA output tolerance: ± 0.02 mA of “0/4 - 20 mA”
Digital display tolerance:
+1 digit
l
Response time
0 - 90%:
10 s
l
Ambient operating temperature
-10 to +55 °C (14 to 131 ºF)
l
Storage temperature
-30 to +70 °C (-22 to 158 ºF)
l
Humidity
10 to 90% RH non-condensing
l
Housing
Case:
Cast aluminium with chemically resistant coating
Cover:
flexible polycarbonate window.
Case color:
off-white (munsell 2.5Y8.4/1.2)
Cover color: moss green (munsell 0.6GY3.1/2.0)
Cable entry:
via six Pg13.5 nylon glands.
Cable terminals:
for up to 2.5 mm
2
finished wires.
Protection:
Weather resistant to IP65, NEMA 4X, CSA Type 3S
Mounting:
Pipe, wall or panel, using optional bracket.
l
Data protection
Non volatile memory for configuration and logbook, and lithium battery for clock support.
9th Edition: Aug. 05, 2015-00