9-12
IM 01F06A00-01E
9. GENERAL DESCRIPTION
9.4.4 Option Specifications (For Explosion Protected Type)
Code
Specification
Item
TIIS Certification
TIIS Flame proof Approval (Note 1)
Flame proof Ex d IIC T6 Certified by TIIS.
(TIIS is the abbreviation of Technology Institution of Industrial Safety.)
Amb. Temp: –20 to 60
°
C
Electrical connection: JIS G1/2 female
FM Intrinsically safe Approval (Note 2)
Applicable Standard: FM3600 1998, FM3610 1999, FM3611 2004, FM3810 2005,
Including Supplement 1 1995, ANSI/NEMA 250 1991, ISA-S12.0.01 1998
Type of Protection : Intrinsically Safe for Class I, II, III, DIV.1, Groups A, B, C, D, E, F and G, T4,
and Class I, Zone 0, AEx ia IIC T4
Nonincendive for Class I, II, Div.2, Groups A, B, C, D, F and G, Class III, DIV.1, T4,
and Class I, Zone 2, Groups IIC, T4
Ambient Temperature : –29 to +60
ⴗ
C
(Integral Type Flowmeter)
–29 to +80
ⴗ
C (Remote Type Flowmeter)
–40 to +60
ⴗ
C (Remote Type Converter)
Maximum Working Pressure : 16MPa (DY015 to DY200)
5MPa (DY250 and DY300)
Indoors and Outdoors : NEMA Type 4X
Electrical Parameter : Vmax=30Vdc, Imax=165mAdc, Pi=0.9W, Ci=12nF, Li=0.15mH
Electrical Connection : ANSI 1/2NPT female
FM Explosion proof Approval
Applicable Standard: FM3600 1998, FM3611 1999, FM3615 2006, FM3810 1989,
Including Supplement 1 1995, ANSI/NEMA 250 1991
Type of Protection : Explosion proof for Class I, Division 1, Groups A, B, C and D;
Dust-ignitionproof for Class II/III, Division 1, Groups E, F, and G.
"SEAL ALL CONDUITS WITHIN 18 INCHES." "WHEN INSTALLED IN DIV.2,
SEALS NOT REQUIRED."
Enclosure Rating : NEMA Type 4X
Temperature Code : T6
Ambient Temperature : –29 to +60
°
C (Integral Type Flowmeter and Remote Type Flowmeter)
–40 to +60
°
C (Remote Type Converter)
Maximum Working Pressure : 16MPa (DY015 to DY200)
5MPa (DY250 and DY300)
Electrical Connection : ANSI 1/2NPT female
Factory Mutual
(FM)
T090407.EPS
CENELEC ATEX(KEMA) Intrinsically safe Approval (Note 2)
Applicable Standard : EN50014: 1997+A1, A2, EN50020: 2002
EN50284: 1999
Type of protection : EEx ia IIC T4...T1(Integral Type Flowmeter and Remote Type Flowmeter)
EEx ia IIC T4(Remote Type Converter)
Group : II
Category : 1G
Degree of Protection of Enclosure : IP67
Maximum Working Pressure : 16MPa (DY015 to DY200)
5MPa (DY250 and DY300)
Tamb. : –29 to +60
°
C (Integral Type Flowmeter)
–29 to +80
°
C (Remote Type Flowmeter)
–40 to +60
°
C (Remote Type Converter)
Process temp.: T4;135
°
C, T3;200
°
C, T2 (*);300
°
C, T1(*); 450
°
C
(*: Use /HT version above 250
°
C)
Electrical data:
Supply and Output Circuit ( and -, PULSE + and -);
Maximum Input Voltage Ui: 30 V
Maximum Input Current Ii: 165 mA
Maximum Input Power Pi: 0.9 W
Internal Capacitance Ci: 6nF
Internal Inductance Li: 0.15mH
For connection of DYA and DY-N
Maximum capacitance of cable: 160nF
Electrical connection : ANSI 1/2NPT female, ISO M20
×
1.5 female.
CENELEC ATEX(KEMA) Explosion proof Approval
Applicable Standard : EN50014: 1997, EN50018: 2000, EN61010-1: 1993
Type of protection : EExd IIC T6...T1(Integral Type Flowmeter and Remote Type Flowmeter)
EExd IIC T6 (Remote Type Converter)
Group : II
Category : 2G
Process temp.: T6; 85
°
C, T5;100
°
C, T4;135
°
C, T3;200
°
C, T2 (*);300
°
C, T1 (*);450
°
C
(*: Use /HT version above 250
°
C)
Degree of Protection of Enclosure : IP67
Tamb: –29 to +60
°
C (Integral Type Flowmeter and Remote Type Flowmeter)
–40 to +60
°
C (Remote Type Converter)
–30 to +60
°
C (Remote Type Converter with indicator)
Maximum working Pressure : 16MPa (DY015 to DY200)
5MPa (DY250 to DY300)
Electrical Connection : ANSI 1/2 NPT female, ISO M20
×
1.5 female.
CENELEC ATEX
(KEMA)
KF1
KS1
FS1
FF1
JF3