
5-14
IM 01F06A00-01E
5. PARAMETER SETUP
[F20 PRESS UNIT] Pressure unit
Select pressure unit from MPa abs(0), kPa abs(1), bar abs(2),
kg/cm
2
a(3), psia(4).
[F21 PRESS f] Pressure value f
Set absolute pressure values at operating condition.
[F22 PRESS b] Pressure value b
Set absolute pressure values at normal/standard Condition.
[F23 DAVIATION] Daviation factor
Set the daviation factor.
[F24 STD/NOR UNIT] Standard/Normal unit
Select Volumetric unit at standard/normal condition From
Nm3(0), k Nm3(1), M Nm3(2), Nl(3), Sm3(4)
k Sm3(5), M Sm3(6), Sl(7), scf(8), k scf(9), M scf(10)
N: Normal
S: Standard
The following item should be done in case of
which F10 is LIQUID:MASS
[F26 DENSITY UNIT] Density unit
Select density unit from kg/m3(0), lb/cf(1), lb/Usgal(2), lb/
Ukgal(3).
[F27 DENSlTY b] Density b
Set density value at standard condition.
[F28 TEMP UNIT] Temperature unit
Select temperature unit from deg C(0), deg F(1).
[F29 TEMP b] Temperature b
Set temperature value at standard condition
[F30 1st coef] 1st coefficient
Set lst temperature coefficient using the density correction.
[F31 2nd coef] 2nd coefficient
Set lst temperature coefficient using the density correction.
[F32 MASS UNIT] Mass unit
Select mass flow rate unit from kg(0), t(1), lb(2), k lb(3).
[F35 TIME UNIT] Time unit
Select time unit from /s(0), /m(1), /h(2), /d(3).
[F40 FLOW SPAN] Flow span
Set span flow rate, 0 to 32000.
[F45 DAMPING] Flow damping
Set flow damping, 0 to 99sec.
[F50 TEMP DAMPING] Temperature damping
Set temperature damping, 0 to 99sec.
[F52 CABLE LENGTH] Cable length of signal
cable(DYC)
Set cable length(m) of signal cable for DYA.
When type of model is integral, cable length is set in 0m.
IMPORTANT
Be sure to set this parameter to correct temperature
measurement error, occured by cable length.
[F55 A/OUT SELECT] Analog out select
Select the analog output from FLOW(0), TEMP(1).
The following item should be done in case of
which F55 is TEMP
[F56 TEMP 0%] Temperature at 0%
Set temperature value at 0%.
[F57 TEMP 100%] Temperature at 100%
Set temperature value at 100%.
[F58 TEMP ERR OUT] Output selection of
thermometer error
Select output function when thermometer error from 0%(0),
OUTLIMIT(H)(1), TEMP f.
In case of OUT LIMIT(H), it is based on parameter “D30”