IM 01C25R03-01E
9-11
9. PARAMETER LISTS
T0903-2.EPS
Write
Mode
Explanation
Factory
Default
Relative
Index
Parameter Name
Index
21
2521
BLOCK_TAG2
2014 (PRIMARY_
VALUE)
–
Block tag which includes a parameter to be displayed on
display2
31
2531
DISPLAY_UNIT3
Null
O/S
User specified unit to be displayed on display3, which will be
available when "Custom" is selected at UNIT SEL3.
32
2532
EXP_MODE3
0
O/S
Selection of the displayed value in exponent such as x1, x10,
x100, and x1000.
33
2533
BLOCK_TAG4
2047 (CAP_
TEMP_VAL)
–
Block tag which includes a parameter to be displayed on
display4
25
2525
DISPLAY_UNIT2
Null
O/S
User specified unit to be displayed on display2, which will be
available when "Custom" is selected at UNIT SEL2.
26
2526
EXP_MODE2
0
O/S
Selection of the displayed value in exponent such as x1, x10,
x100, and x1000.
27
2527
BLOCK_TAG3
2029 (SECONDARY_
VALUE)
–
Block tag which includes a parameter to be displayed on
display3
29
2529
DISPLAY_TAG3
Null
O/S
Name of block tag to be displayed on display3; up to six
alphanumeric plus a slash [/] and a period [.]
35
2535
DISPLAY_TAG4
Null
O/S
Name of block tag to be displayed on display4; up to six
alphanumeric plus a slash [/] and a period [.]
23
2523
DISPLAY_TAG2
Null
O/S
Name of block tag to be displayed on display2; up to six
alphanumeric plus a slash [/] and a period [.]
24
2524
UNIT_SEL2
0 (Auto)
O/S
Selection of unit to be displayed. The unit of the parameter
which is selected at PARAMETER SEL2 will be displayed
when "Auto" is selected; user-specified unit at DISPLAY
UNIT2 will be displayed when "Custom" is selected.
30
2530
UNIT_SEL3
0 (Auto)
O/S
Selection of unit to be displayed. The unit of the parameter
which is selected at PARAMETER SEL3 will be displayed
when "Auto" is selected; user-specified unit at DISPLAY
UNIT3 will be displayed when "Custom" is selected.
22
2522
PARAMETER_
SEL2
0 (PRIMARY_
VALUE)
O/S
Selection of a parameter to be displayed on display2.
Select a parameter from Table 6.1
28
2528
PARAMETER_
SEL3
1 (SECONDARY_
VALUE)
O/S
Selection of a parameter to be displayed on display3.
Select a parameter from Table 6.1
34
2034
PARAMETER_
SEL4
3 (CAP_TEMP_
VAL)
O/S
Selection of a parameter to be displayed on display4.
Select a parameter from Table 6.1
37
2537
DISPLAY_UNIT4
Null
O/S
User specified unit to be displayed on display4, which will be
available when "Custom" is selected at UNIT SEL4.
38
2038
EXP_MODE4
0
O/S
Selection of the displayed value in exponent such as x1, x10,
x100, and x1000.
39
2039
BAR_GRAPH_
SELECT
1 (display)
O/S
Selection of bar graph indicator
36
2536
UNIT_SEL4
0 (Auto)
O/S
Selection of unit to be displayed. The unit of the parameter
which is selected at PARAMETER SEL4 will be displayed
when "Auto" is selected; user-specified unit at DISPLAY
UNIT4 will be displayed when "Custom" is selected.
40
2540
DISPLAY_CYCLE
3 (2.4sec)
O/S
Duration of display cycle. (Time unit: 1=800msec)
41
2541
TEST40
0
–
Not used for EJX
20
2520
EXP_MODE1
0
O/S
Selection of the displayed value in exponent such as x1, x10,
x100, and x1000.