![YOKOGAWA DPharp EJAC50E Скачать руководство пользователя страница 58](http://html.mh-extra.com/html/yokogawa/dpharp-ejac50e/dpharp-ejac50e_user-manual_911099058.webp)
<7. General Specifications>
7-11
IM 01C25W01-01EN
■ for EJXC80A, EJAC80E Direct Mount Flanged Diaphragm Seal System
EJXC50A, EJAC50E Direct Mount Flanged Diaphragm Seal System
EJXC40A Digital Remote Sensor Flange Mounted Diaphragm Seal System
Process temperature
for fill fluid code F
Process temperature
for fill fluid code B
Process temperature
for fill fluid code A
Working
pressure
kPa abs
2.7
100
0.1
1
10
-50 0
50 100 150 200 250 300 350
-100
Temperature °C
Flange max.
working
pressure
Transmitter
ambient
temperature
range
F28E.ai
Figure 1e. Working Pressure and Process Temperature
(Fill fluid: silicone oil for general and low
temperature use)
Process temperature
for fill fluid code E
Process temperature
for fill fluid code P
Process temperature
for fill fluid code D
Flange max.
working pressure
Transmitter
ambient
temperature
range
Working
pressure
kPa abs
2.7
Temperature °C
100
0.1
1
10
-50 0
50 100 150 200 250 300 350
F29E.ai
-100
51
Figure 1f. Working Pressure and Process Temperature
(Fill fluid: silicone oil for oil-prohibited use,
ethylene glycol and propylene glycol)
0
340
-100
100
200
300
0
-20
-40
-60
-80
-100
100
80
60
40
20
Fill fluid code A
Fill fluid code F
Ambient temperature °C
Process temperature °C
Fill fluid code B
F30E.ai
Figure 1g. Ambient Temperature and Process
Temperature (Fill fluid: silicone oil for
general and low temperature use)
0
340
-100
100
200
300
0
-20
-40
-60
-80
-100
100
80
60
40
20
Fill fluid code E
Fill fluid code P
F31E.ai
Process temperature °C
Ambient temperature °C
Fill fluid code D
Figure 1h. Ambient Temperature and Process
Temperature (Fill fluid: silicone oil for
oil-prohibited use, ethylene glycol and