9. GENERAL SPECIFICATIONS
d
Physical Specifications
Wetted Parts Materials:
Diaphragm and other wetted parts;
See ‘Model and Suffix Codes’
Non-wetted Parts Materials:
Capillary tube;
SUS316
Protection tube;
SUS304, PVC-sheathed [Max. operating tempera-
ture: 100
°
C (212
°
F) ]
Fill Fluid;
See Table 1.
Housing;
Low copper cast-aluminum alloy with polyurethane
paint (Munsell 0.6GY3.1/2.0)
Enclosure Classification;
JIS C0920 immersion proof (equivalent to NEMA
4X and IEC IP67)
Cover O-rings;
Buna-N
Data plate and tag;
SUS304
Weight:
21.7 kg (47.8 lb): Model EJA118W, 3-inch ANSI
Class 150 flange, without mounting bracket. Add
1.4 kg (3.1 lb) for JIS SCS14A stainless steel
amplifier housing.
Connections:
Refer to the ‘Model and Suffix Codes’ to specify
the process and electrical connection type.
Process temperature
for fill fluid code B
Process temperature
for fill fluid code A
Process temperature
for fill fluid code C
Flange max.
working
pressure
Atmospheric
pressure
Transmitter ambient
temperature range
(For fill fluid code A,B)
Working
pressure
2.7{20}
Process Temperature (
8
C)
kPa abs
{mmHg abs}
100{750}
0.1{0.75}
1{7.5}
10{75}
-50
0
50
100 150 200 250 300
F0901.EPS
Figure 1. Working Pressure and Process Temperature
Table 1. Process Temperature and Ambient Temperature
Process
Temperature
(Note 1)
Ambient
temperature
(Note 2)
Working pressure
Specific
gravity
(Note 3)
See Figure 1.
0.94
1.07
1.09
–10 to 250
8
C
(14 to 482
8
F)
–10 to 60
8
C
(14 to 140
8
F)
–30 to 180
8
C
(–22 to 356
8
F)
–15 to 60
8
C
(5 to 140
8
F)
10 to 300
8
C
(50 to 572
8
F)
10 to 60
8
C
(50 to 140
8
F)
–20 to 120
8
C
(–4 to 248
8
F )
–10 to 60
8
C
(–14 to 140
8
F )
–50 to 100
8
C
(–58 to 212
8
F)
–40 to 60
8
C
(–40 to 140
8
F)
Fill Fluid
Code ‘A’
Fill Fluid
Code ‘B’
Fill Fluid
Code ‘C’
Fill Fluid
Code ‘D’
Fill Fluid
Code ‘E’
51 kPa abs or more
{380 mmHg abs}
Vacuum pressure
not allowed
1.90 to 1.92
1.09
Silicone Oil
Fluorinated Oil
Ethylene Glycol
T0903.EPS
Note 1: See Figure 1. ‘Working Pressure and Process Temperature.’
Note 2: This ambient temperature is the transmitter ambient temperature.
Note 3: Approximate values at a temperature of 25
°
C(77
°
F)
Note 4: The differential pressure transmitter should be installed at least 600 mm below the high
pressure(HP) process connection. However, this value(600 mm) may be affected by ambient
temperature, operating pressure, fill fluid or material of the wetted diaphragm. Contact
YOKOGAWA when the transmitter can not be installed at least 600 mm below the HP process
connection.
<Settings When Shipped “
e
”>
Tag Number
Output Mode
Display Mode
Operation Mode
Damping Time
Constant
As specified in order
*1
‘Linear’ unless otherwise specified in order
‘2 sec.’
‘Normal’ unless otherwise specified in order
‘Linear’ unless otherwise specified in order
Calibration Range
Units
As specified in order
As specified in order
Calibration Range
Higher Range Value
Calibration Range
Lower Range Value
Selected from mmH
2
O, mmAq, mmWG,
mmHg, Pa, hPa, kPa, MPa, mbar, bar,
gf/cm
2
, kgf/cm
2
, inH
2
O, inHg, ftH
2
O, or
psi. (Only one unit can be specified)
T0909.EPS
Note 1: If Tag No. is no more than 16 alphanumeric characters (including - and ·), it will be written into
the tag plate and amplifier memory settings.
Содержание Dpharp EJA118N
Страница 2: ...Blank Page...
Страница 6: ...Blank Page...
Страница 70: ...Blank Page...
Страница 72: ...2 CMPL 1C22A1 01E Subject to change without notice Printed in Japan Blank Page...