IM 1C22H1-01E
7-11
7. BRAIN TERMINAL BT200 OPERATION
b. Setting User-set Engineering Unit
(D21: DISP UNIT)
This parameter allows entry of the engineering units to
be displayed on the BT200. When the instrument is
shipped, this is set as specified in the order.
Follow the procedure below to change this setting.
Since these units are not displayed on the integral
indicator, use the adhesive labels provided. This
parameter need not be set for % display.
• Example: Set an engineering unit M.
CODE
CAPS
CLR
ESC
SET
D21:DISP UNIT
M
FEED
NO
OK
SET
D21:DISP UNIT
M_
Set “M.”
Press the key twice to
enter the setting.
Press the (OK) key.
F0721.EPS
c. Lower and Higher Range Value Setup
in Engineering Unit (D22: DISP LRV,
D23: DISP HRV)
These parameter items are used to set the lower and
higher range values for the engineering unit display.
When the instrument is shipped, these are set as
specified in the order. Follow the procedure below to
change these settings. Note that these parameters need
not be set for % display.
• Example: Set lower range value (LRV) to –50
and higher range value (HRV) to 50.
FEED
NO
OK
SET
D23:DISP HRV
50M
Press the (OK) key.
F0722.EPS
DEL
CLR
ESC
DATA
DIAG
PRNT
ESC
PARAM
D21:DISP UNT
M
D22:DISP LRV
– 50M
D23:DISP HRV
50M
DEL
CLR
ESC
SET
D22:DISP LRV
0M
- 50
Set “–50.”
Press the key twice to
enter the setting.
Setting LRV
Setting HRV
Set “50.”
Press the key twice to
enter the setting.
SET
D23:DISP HRV
100M
+ 50
(7) Unit Setup for Displayed Temperature
(D30: TEMP UNIT)
When the instrument is shipped, the temperature units
are set to
degC
. Follow the procedure below to
change this setting. Note that changing the unit here
changes the unit for
A20: AMP TEMP
(amplifier
temperature) and
A21: CAPSULE TEMP
(capsule
temperature).
• Example: Change the unit for the temperature
display.
ESC
SET
D30:TEMP UNIT
deg C
< deg C >
< deg F >
Use the or key to
select “deg F.”
Press the key twice to
enter the setting.
F0723.EPS
Содержание Dpharp EJA118N
Страница 2: ...Blank Page...
Страница 6: ...Blank Page...
Страница 70: ...Blank Page...
Страница 72: ...2 CMPL 1C22A1 01E Subject to change without notice Printed in Japan Blank Page...