background image

5-76

IM 701730-17E

:MEASure:USER<x>?

Function

Queries all settings related to the automated
measurement of waveform parameters of the
Dual Area.

Syntax

:MEASure:USER<x>?

<x>=1 to 4

Example

:MEASURE:USER1? -> :MEASURE:USER1:
AREA1:TRACE 1;ITEM PTOPEAK;:
MEASURE:USER1:AREA2:TRACE 2;
ITEM MAXIMUM;:MEASURE:USER2:AREA1:
TRACE 3;ITEM MINIMUM;:MEASURE:
USER2:AREA2:TRACE 4;ITEM RMS;:
MEASURE:USER3:AREA1:TRACE MATH1;
ITEM PERIOD;:MEASURE:USER3:AREA2:
TRACE MATH2;ITEM AVGPERIOD;:
MEASURE:USER4:AREA1:TRACE CONST1;
ITEM TY1INTEG:MEASURE:USER4:
AREA2:TRACE NONE;ITEM XY2INTEG

:MEASure:USER<x>:AREA<y>?

Function

Queries all settings related to the specified
computation parameter/area of the automated
measurement of waveform parameters of the
Dual Area.

Syntax

:MEASure:USER<x>:AREA<y>?

<x>=1 to 4
<y>=1, 2

Example

:MEASURE:USER1:AREA1? ->
:MEASURE:USER1:AREA1:TRACE 1;
ITEM PTOPEAK

:MEASure:USER<x>:AREA<y>:ITEM

Function

Sets the measurement item of the specified
computation parameter/area of the automated
measurement of waveform parameters of the
Dual Area or queries the current setting.

Syntax

:MEASure:USER<x>:AREA<y>:
ITEM {<Parameter>}
:MEASure:USER<x>:AREA<y>:ITEM?

<x>=1 to 4
<y>=1, 2
<Parameter>=

{AVERage|AVGFreq|

AVGPeriod|BWIDth1|BWIDth2|DUTYcycle|
FALL|FREQuency|HIGH|LOW|MAXimum|
MINimum|NOVershoot|NWIDth|PERiod|
PNUMber|POVershoot|PTOPeak|PWIDth|
RISE|RMS|SDEViation|TY1Integ|TY2Integ|
XY1Integ|XY2Integ}

Example

:MEASURE:USER1:AREA1:ITEM PTOPEAK
:MEASURE:USER1:AREA1:ITEM? ->
:MEASURE:USER1:AREA1:ITEM PTOPEAK

:MEASure:USER<x>:AREA<y>:TRACe

Function

Sets the measurement trace of the specified
computation parameter/area of the automated
measurement of waveform parameters of the
Dual Area.

Syntax

:MEASure:USER<x>:AREA<y>:TRACe
{<NRf>|MATH<z>|NONE|CONSt<w>}
:MEASure:USER<x>:AREA<y>:TRACe?

<x>=1 to 4
<y>=1 or 2
<NRf>=1 to 4 (1 to 2 on the DL1720E)
<z>=1 or 2 (1 on the DL1720E)
<w>=1 to 4

Example

:MEASURE:USER1:AREA1:TRACE 1
:MEASURE:USER1:AREA1:TRACE ->
:MEASURE:USER1:AREA1:TRACE 1

:MEASure:USER<x>:COUNt?

Function

Queries the statistical processing count of the
automated measurement of waveform
parameters of the Dual Area.

Syntax

:MEASure:USER<x>:COUNt?

<x>=1 to 4

Example

:MEASURE:USER1:COUNT? ->
:MEASURE:USER1:COUNT 100

:MEASure:USER<x>:{MAXimum|MEAN|
MINimum|SDEViation}?

Function

Queries the statistical value of the automated
measurement of waveform parameters of the
Dual Area.

Syntax

:MEASure:USER<x>:{MAXimum|MEAN|
MINimum|SDEViation}?

<x>=1 to 4

Example

(The following is an example for the maximum
value.)

:MEASURE:USER1:MAXIMUM? ->
:MEASURE:USER1:MAXIMUM 1.000E+00

Description

If the statistics cannot be determined, “NAN
(Not A Number)” is returned.

:MEASure:USER<x>:OPERation

Function

Sets the computing operation of the automated
measurement of waveform parameters of the
Dual Area or queries the current setting.

Syntax

:MEASure:USER<x>:OPERation {PLUS|
MINus|MULTiple|DIVision}
:MEASure:USER<x>:OPERation?

<x>=1 to 4

Example

:MEASURE:USER1:OPERATION PLUS
:MEASURE:USER1:OPERATION ->
:MEASURE:USER1:OPERATION PLUS

:MEASure:USER<x>:STATe

Function

Turns ON/OFF the automated measurement of
waveform parameters of the Dual Area or
queries the current setting.

Syntax

:MEASure:USER<x>:STATe {<Boolean>}
:MEASure:USER<x>:STATe?

<x>=1 to 4

Example

:MEASURE:USER1:STATE ON
:MEASURE:USER1:STATE ->
:MEASURE:USER1:STATE 1

:MEASure:USER<x>:UNIT

Function

Sets the display unit of the automated
measurement of waveform parameters of the
Dual Area or queries the current setting.

Syntax

:MEASure:USER<x>:UNIT <Character
string>
:MEASure:USER<x>:UNIT?

<x>=1 to 4
<Character string>=Up to 3 characters

Example

:MEASURE:USER1:UNIT "UNIT"
:MEASURE:USER1:UNIT ->
:MEASURE:USER1:UNIT "UNIT"

5.17 MEASure Group

Содержание DL1720E

Страница 1: ...Digital Oscilloscope Communication Interface IM 701730 17E 1st Edition DL1720E DL1740E DL1740EL...

Страница 2: ...ions Operation Guide and basic operating procedures of the DL1720E DL1740E DL1740EL DL1740E DL1740EL IM 701730 51E Describes the serial bus signal analysis Serial Bus Signal Analisis function option F...

Страница 3: ...f Adobe Systems incorporated For purposes of this manual the TM and symbols do not accompany their respective trademark names or registered trademark names The TCP IP software and TCP IP related docum...

Страница 4: ...f the Ethernet Interface Option Describes the functions and specifications of the Ethernet interface Chapter 4 Before Programming Describes formats used when sending a command Chapter 5 Commands Descr...

Страница 5: ...eatures in descriptions Procedure Carry out steps in the order shown The operating procedures are given with the assumption that you are not familiar with the operation Thus it may not be necessary to...

Страница 6: ...mbols are referred to as BNF notation Backus Naur Form For detailed information refer to page 4 5 to 4 6 Symbol Description Example Example of Input Defined value CHANnel x x 1 to 4 CHANNEL2 One of th...

Страница 7: ...ce Option 3 1 Names of the Parts and their Functions 3 1 3 2 Ethernet Interface Functions and Specifications 3 2 3 3 Connection via the Ethernet Interface 3 4 3 4 Setting up this Instrument Network 3...

Страница 8: ...rt 6 1 Overview of the Status Report 6 1 6 2 Status Byte 6 2 6 3 Standard Event Register 6 3 6 4 Extended Event Register 6 4 6 5 Output Queue and Error Queue 6 5 Chapter 7 Sample Program 7 1 Before Pr...

Страница 9: ...20pF 40Vpk _ MISC Key Page 1 5 Press to enter the communication settings such as the address and the timeout SHIFT Key CLEAR TRACE Key Press to switch from remote mode to local mode which allows key o...

Страница 10: ...However it is not possible to connect more than 15 items of equipment including the controller to a single bus If you connect the instrument to more than one item of equipment make sure that a differ...

Страница 11: ...al to Remote Mode Remote mode is activated when a REN Remote Enable message is received from a controller while local mode is active REMOTE is displayed on the upper right of the screen All front pane...

Страница 12: ...sten Address no talk only capability Listener L4 Basic listener capability unlisten on MTA My Talk Address no listen only capability Service Request SR1 Full service request capability Remote Local RL...

Страница 13: ...ey 5 Turn the jog shuttle to set the desired address Explanation When you make settings which can be made using the front panel keys of the instrument or when you output setup data or waveform data us...

Страница 14: ...T Take Control are not supported Response to a Multi Line Message Universal Command LLO Local LockOut Invalidates SHIFT CLEAR on the front panel to disable switching to local mode DCL Device CLear Sam...

Страница 15: ...ut DCL Device CLear PPU Parallel Poll Unconfigure SPE Serial Poll Enable SPD Serial Poll Disable In addition to the above commands a listener address talker address on secondary command can be sent in...

Страница 16: ...pk _ MISC Key Page 2 5 Press to enter the communication settings such as the ID value SHIFT Key CLEAR TRACE Key Press to switch from remote mode to local mode which allows key operation However this i...

Страница 17: ...is required for connection to the PC Data Transfer Rate The table below shows approximate response times for output of waveform data assuming the following configuration Model 701730 Controller PC Pe...

Страница 18: ...When the DL1720E DL1740E DL1740EL receives a COMMunicate REMote OFF command from the PC the DL1720E DL1740E DL1740EL switches to the local mode regardless of the local lock condition REMOTE indication...

Страница 19: ...multiple devices using USB hubs connect the instrument to the USB hub that is closest to the controller Do not mistakenly insert the USB cable to the GO NO GO output terminal This may cause damage to...

Страница 20: ...key 5 Turn the jog shuttle to set the ID value Explanation Carry out the following settings when using a controller to set information that can be specified through key operation on the instrument or...

Страница 21: ...k _ _ 1M 20pF 40Vpk _ MISC Key Page 3 5 Press to confirm the user name and password SHIFT Key CLEAR TRACE Key Press to switch from remote mode to local mode which allows key operation However this is...

Страница 22: ...s approx 1 9 s 100000 approx 182 1 ms approx 401 5 ms approx 19 6 s 1000000 approx 1908 8 ms approx 3006 3 ms approx 197 2 s Switching between Remote and Local Mode When Switched from Local to Remote...

Страница 23: ...User Verification Function The user must enter a user name and password for the Ethernet interface when connecting to the network The user name and password is set on the instrument in the Remote Cntl...

Страница 24: ...ble or STP Cable Straight Cable Network Card Ethernet Personal Computer or Workstation DL1720E DL1740E DL1740EL 100BASE TX Compatible Hub or Router PERIPHERAL USB USB ISM 1 A N200 LINK ACT ETHERNET 10...

Страница 25: ...the User Name If you do not wish to limit access use the anonymous setting If access is to be limited enter a user name of 15 characters or less 6 Use the jog shuttle SELECT to enter the Password of...

Страница 26: ...ay be used The default value is anonymous Password Setting 15 characters or less may be used Setting the Timeout Time If the unit is not being accessed at the specified timeout time the connection to...

Страница 27: ...nd this instrument during communications Messages sent from the controller to this instrument are called program messages and messages sent back from this instrument to the controller are called respo...

Страница 28: ...end another program message until a response message has been received An error will occur if a program message is sent before a response message has been received in its entirety A response message w...

Страница 29: ...ble to omit a colon Example ACQuire MODE NORMal STARt PMT When Concatenating Common Commands Common commands defined in IEEE 488 2 1987 are independent of hierarchical level Thus it is not necessary t...

Страница 30: ...y levels When Separating Commands with PMT If a terminator is used to separate two commands each command is a separate message Therefore the common header must be typed in for each command even when c...

Страница 31: ...ROBe 100 Voltage Time Physical value Frequency Current Example Time axis range TIMebase TDIV 1US Register Register value expressed as either binary octal decimal or hexadecimal Example Extended event...

Страница 32: ...F 1 0 When Boolean is expressed in NRf form OFF is selected if the rounded integer value is 0 and ON is selected if the rounded integer is Not 0 A response message is always 1 if the value is ON and 0...

Страница 33: ...ter string 1 CASE CASE If you input an NRf value the system converts the value after rounding to the nearest integer to the corresponding 8 character ASCII string If you set the value to 1 the name be...

Страница 34: ...p EXECute CASE1 WAI CHANnel1 VDIV PMT The COMMunicate OPSE command is used to designate which commands are to be subject to the WAI command In the above example only auto setup is designated Since a W...

Страница 35: ...o judge whether acquisition is in progress or not by reading bit 0 of the condition register Bit 0 is 1 if acquisition is in progress and 0 if acquisition is stopped Example TRIGger MODE SINGle STARt...

Страница 36: ...ted to the vertical axis of the channel 5 20 CHANnel x BWIDth Sets the input filter of the channel or queries the current setting 5 20 CHANnel x COUPling Sets the input coupling of the channel or quer...

Страница 37: ...r cursors 5 29 CURSor TY MARKer M x Queries all settings related to the specified marker cursor 5 29 CURSor TY MARKer M x JUMP Jumps the marker cursor onto the zoom waveform display frame 5 29 CURSor...

Страница 38: ...eform CHANnel x BACK MATH x Sets the colors for waveform and computation channels or queries the current setting 5 36 DISPlay FORMat Sets the display format or queries the current setting 5 36 DISPlay...

Страница 39: ...O and also the determination criteria 5 47 GONogo ACTion BUZZer Sets whether or not to sound an alarm when the determination is NO GO or queries the current setting 5 47 GONogo ACTion HCOPy Sets wheth...

Страница 40: ...ent setting 5 51 HCOPy Group HCOPy Queries all settings related to the print of the screen image data 5 53 HCOPy ABORt Aborts printing or paper feeding 5 53 HCOPy COMMent Sets comments at the lower se...

Страница 41: ...elated to the history search zone 5 59 HISTory ZONE EDIT x Queries all settings related to the search zone of the history search zone 5 59 HISTory ZONE EDIT x CONDition Sets the search condition of th...

Страница 42: ...surement of waveform parameters 5 69 MEASure AMODe Sets the normal automated measurement of parameters NORMal and the automated measurement of parameters on dual areas AREA2 or queries the current set...

Страница 43: ...re CYCLe ABORt Aborts the periodic statistical processing of the automated measurementof waveform parameters Cycle Statistics 5 75 MEASure CYCLe EXECute Executes the periodic statistical processing of...

Страница 44: ...setting 5 80 PRESet CHANnel x USERdefine Queries all preset values when using user type presets 5 80 PRESet CHANnel x USERdefine LEVel Sets the preset value of the user type trigger level or queries...

Страница 45: ...aveform number 5 89 SEARch SELect MAXimum Queries the detection count of the search zoom function 5 89 SEARch SPATtern Queries all settings related to the serial pattern search of the search zoom func...

Страница 46: ...5 94 STATus SPOLl Executes serial polling 5 94 STOP Group STOP Stops the waveform acquisition 5 94 STORe Group STORe Queries information about the setting parameters that are stored to the internal m...

Страница 47: ...orts are sent as an ACTion 5 105 TRIGger ACTion SAVE Sets whether or not to save the waveform data when the action is enabled or queries the current setting 5 105 TRIGger ACTion STARt Starts action on...

Страница 48: ...ries the current setting 5 109 TRIGger WIDTh CONDition Sets the trigger condition or queries the current setting 5 110 TRIGger WIDTh TIME x Sets the pulse width or queries the current setting 5 110 TR...

Страница 49: ...setting 5 117 ZOOM MODE Sets the zoom waveform display mode or queries the current setting 5 117 ZOOM PLINkage Turns the link ON OFF when moving the zoom box or queries the current setting 5 117 ZOOM...

Страница 50: ...about acquisition This allows you to make the same settings that you can make using ACQ on the front panel ACQuire AVERage COUNt Space NRf INFinity EWEight Space NRf CLOCk Space INTernal EXTernal COU...

Страница 51: ...ACQuire CLOCk Example ACQUIRE CLOCK INTERNAL ACQUIRE CLOCK ACQUIRE CLOCK INTERNAL ACQuire COUNt Function Sets the number of waveform acquisitions during normal mode envelope mode and box average or q...

Страница 52: ...ple CALIBRATE DESKEW CALIBRATE DESKEW MODE 0 TARGET 1 TIME 0 0E 00 CALibrate DESKew MODE Function Turns the skew compensation ON OFF or queries the current setting Syntax CALibrate DESKew MODE Boolean...

Страница 53: ...te Function Executes calibration Syntax CALibrate EXECute Example CALIBRATE EXECUTE CALibrate MODE Function Sets queries the ON OFF of the auto calibration Syntax CALibrate MODE AUTO OFF CALibrate Exa...

Страница 54: ...n make using VERTICAL CH V DIV CHANnel x BWIDth Space FULL COUPling Space AC DC DC50 GND DISPlay Space OFF ON NRf LABel Space Character string Character string Frequency Voltage Voltage Voltage LSCale...

Страница 55: ...d symbols displayed on the keyboard on the screen can be used In ASCII code and are expressed as 1EH and 1FH respectively CHANnel x LSCale Function Queries all settings related to the linear scaling o...

Страница 56: ...s the probe attenuation current to voltage conversion ratio of the channel or queries the current setting Syntax CHANnel x PROBe NRf C10 C100 CHANnel x PROBe x 1 to 4 1 to 2 on the DL1720E NRf 1 10 10...

Страница 57: ...roup The CLEar command is used to cler the trace This allows you to perform the same operation that you can perform using CLEAR TRACE on the front panel CLEar 5 6 CLEar Group CLEar Function Clears tra...

Страница 58: ...ean COMMunicate LOCKout Example COMMUNICATE LOCKOUT ON COMMUNICATE LOCKOUT COMMUNICATE LOCKOUT 1 Description This is an exclusive commandfor the USB interface or Ethernet interface option 5 7 COMMunic...

Страница 59: ...ets remote local It is in remote mode when it is set to ON Syntax COMMunicate REMote Boolean COMMunicate REMote Example COMMUNICATE REMOTE ON COMMUNICATE REMOTE COMMUNICATE REMOTE 1 Description This i...

Страница 60: ...t you can make using CURSOR on the front panel CURSor TY DEGRee DX DY JUMP Space C1_Z1 C1_Z2 C2_Z1 C2_Z2 POSition x Space NRf REFerence x Space NRf RVALue Space NRf TRACe Space ALL NRf MATH x X x Y x...

Страница 61: ...ion x Space NRf TRACe Space ALL NRf MATH x X x Y x XY HORizontal DY POSition x Space NRf TRA Ce Space XY1 XY2 Y x MARKer T POSition Space NRf TRACe Space XY1 XY2 X Y TYPE Space OFF HORizontal VERTical...

Страница 62: ...90 CURSor TY DEGRee DX Function Queries the angle difference D between angle cursors in the T Y display Syntax CURSor TY DEGRee DX Example CURSOR TY DEGREE DX CURSOR TY DEGREE DX 180 00000E 00 CURSor...

Страница 63: ...TY HORIZONTAL DY CURSOR TY HORIZONTAL DY 3 0000000E 00 Description When the linear scaling is ON the scaled value is queried CURSor TY HORizontal POSition x Function Sets the horizontal cursor positio...

Страница 64: ...ARKER M1 POSITION 1 CURSOR TY MARKER M1 POSITION CURSOR TY MARKER M1 POSITION 1 0000000 CURSor TY MARKer M x TRACe Function Sets the waveform on which to use the marker cursor in the T Y display or qu...

Страница 65: ...ICAL POSITION1 2 00E 00 CURSo TY VERTical TRACe Function Sets the target waveform of the vertical cursor of the T Y display or queries the current setting Syntax CURSor TY VERTical TRACe NRf MATH x PO...

Страница 66: ...00E 03 CURSor XY MARKer TRACe Function Sets the waveform on which to use the marker cursor in the X Y display or quries the current setting DL1740E DL1740EL only Syntax CURSor XY MARKer TRACe XY1 XY2...

Страница 67: ...y Syntax CURSor XY VERTical TRACe XY1 XY2 CURSor XY VERTical TRACe Example CURSOR XY VERTICAL TRACE XY1 CURSOR XY VERTICAL TRACE CURSOR XY VERTICAL TRACE XY1 CURSor XY VERTical X x Function Queries th...

Страница 68: ...tence COLor PERSistence Space Time INFinity COLor Space RGB INTensity RGB MENU BACK Space NRf NRf NRf DIALog Space NRf NRf NRf FOCus Space NRf NRf NRf FORE Space NRf NRf NRf LINE Space NRf NRf NRf SEL...

Страница 69: ...ce NRf MENU Space NRf GRID Space NRf SNAP Space NRf WAVeforn Space NRf INTERpolate Space OFF SINE LINE PULSe MAPPing Space AUTO FIXed USERdefine SMAPping CHANnel x Space NRf MATH x Space NRf SVALue Sp...

Страница 70: ...S SNAP 15 15 15 ZONE1 0 0 15 ZONE2 15 4 11 ZONE3 11 15 4 ZONE4 0 7 15 GRID 8 8 8 CURSOR 13 13 13 MARKER 15 15 15 5 9 DISPlay Group DISPlay Function Queries all display settings Syntax DISPay Example D...

Страница 71: ...HEXa DISPlay FORMat Example DISPLAY FORMAT QUAD DISPLAY FORMAT DISPLAY FORMAT QUAD Description QUAD and HEXa are not available on the DL1720E 5 9 DISPlay Group DISPlay GRATicule Function Sets queries...

Страница 72: ...lects queries whether scaling values are displayed Syntax DISPlay SVALue Boolean DISPlay SVALue Example DISPLAY SVALUE OFF DISPLAY SVALUE DISPLAY SVALUE 0 DISPlay TLABel Trace LABEL Funcion Selects qu...

Страница 73: ...Filename JPEG EXECute Space Filename MEASure EXECute Space Filename PNG EXECute Space Filename POSTscript EXECute Space Filename SETup EXECute Space Filename SNAP EXECute Space Filename TIFF EXECute...

Страница 74: ...FF ON NRf ASCii EXECute ABORt HISTory Space ONE ALL RANGe Space MAIN Z1 Z2 Z1_Z2 TRACe Space ALL NRf MATH x BINary ABORt HISTory Space ONE ALL RANGe Space MAIN Z1 Z2 Z1_Z2 COMPression Space OFF ON NRf...

Страница 75: ...ilename SETup EXECute ABORt SNAP EXECute ABORt NRf MATH x SNAP EXECute PLANe Space NRf ALL UNLoad BINary EXECute TRACe Space ALL 1 FLOPpy can be used when the built in floppy disk drive is available P...

Страница 76: ...nd NRf FLOPpy can used when the built in floppy disk drive is available PCCard can be used when the built in PC card drive is available NETWork can be used when the Ethernet interface option is instal...

Страница 77: ...ASCii BINary FLOat HISTory Function Sets whether to save all the waveform data or a selected waveorm or queries the current setting Syntax FILE SAVE ASCii BINary FLOat HISTory ONE ALL FILE SAVE ASCii...

Страница 78: ...EXECute Function Executes the unloading of the various data that has been loaded This is an overlap command Syntax FILE UNLoad BINary SNAP EXECute Example FILE UNLOAD BINARY EXECUTE FILE UNLoad BINar...

Страница 79: ...nd inquiries as when GO NO GO on the front panel is pressed GONogo ABORt ACTion BUZZer Space OFF ON NRf HCOPy Space OFF ON NRf MAIL MODE SAVE Space OFF ON NRf COUNt EXECute LOGic Space AND OR MODE Spa...

Страница 80: ...e DONTcare HIGH Space DONTcare DONTcare LOW MAXimum MINimum NOVershoot Space NRf NRf DONTcare DONTcare NWIDth Space DONTcare DONTcare PERiod PNUMber Space NRf NRf DONTcare DONTcare POVershoot PTOPeak...

Страница 81: ...n Space OUT IN TRACe Space MATH MODE Space Boolean SEND Space Block data TRANge Space NRf NRf NRf TLINkage Space ON OFF WAIT SEQuence Space SINGle CONTinue STOPcount Space NRf INFinity 1 Can be used w...

Страница 82: ...print the screen image to the built in printer optional during NO GO determination or queries the current setting Syntax GONogo ACTion HCOPy Boolean GONogo ACTion HCOPy Example GONOGO ACTION HCOPY OF...

Страница 83: ...TEM1 CONDITION OUT GONogo PARameter ITEM x CAUSe Function Queries whether or not the waveform parameter is the cause of the NO GO result Syntax GONogo PARameter ITEM x CAUSe x 1 to 4 Example GONOGO PA...

Страница 84: ...on Selects single or continuous action or queries the current setting Syntax GONogo SEQuence CONTinuous SINGle GONogo SEQuence Example GONOGO SEQUENCE SINGLE GONOGO SEQUENCE GONOGO SEQUENCE SINGLE GON...

Страница 85: ...H1 and 2 7 to 12 are determination zones 1 to 6 On the DL1720E X 1 2 5 7 8 or 9 1 to 2 are traces 5 is MATH1 7 to 9 are determination zones 1 to 3 NRf 0 to 5 00 div 0 01 steps left and right 0 to 8 00...

Страница 86: ...ontain ASCII codes 0AH or 00H that stand for NL Hence care must be taken when programming the controller This command is used to save the zone data to the PC and set the data as zone data on the instr...

Страница 87: ...tional built in printer You can make the same settings that you can make by pressing SHIFT then pressing COPY HCOPy ABORt COMMent Space Character string DIRection Space EXTPrinter PRINter NETPrint EXE...

Страница 88: ...Rf 1 to 127 Example HCOPY EXTPRINTER SELECT 3 HCOPY EXTPRINTER SELECT HCOPY EXTPRINTER SELECT 3 Description If you specify an address at which no printer is connected an error occurs In this case NAN...

Страница 89: ...E HCOPY NETPRINT TYPE PSCRIPT Description NRf is specified only when the type is set to BJ Can be used when the Ethernet interface option is installed HCOPy PRINter Function Queries all settings relat...

Страница 90: ...PARameter ITEM x CONDition IN OUT OFF SOURce NRf MATH x TYPE AVERage Space NRf NRf AVGFreq Space AVGPeriod Space BWIDth x Space DELay Space DUTYcycle Space NRf NRf FALL Space FREQuency Space HIGH Spa...

Страница 91: ...RECord Space NRf MINimum TIME Space NRf MINimum ZONE EDIT x CONDition Space IN OUT OFF HORizontal Space NRf NRf SOURce Space NRf MATH x VERTical Space NRf NRf LOGic Space AND OR TRANge Space NRf NRf S...

Страница 92: ...ls see the User s Manual HISTory EXECute Function Executes the history search This is an overlap command Syntax HISTory EXECute Example HISTORY EXECUTE HISTory DMODe Display Mode Function Sets the dis...

Страница 93: ...ameter or queries the current setting Syntax HISTory PARameter LOGic AND OR HISTory PARameter LOGic Example HISTORY PARAMETER LOGIC AND HISTORY PARAMETER LOGIC HISTORY PARAMETER LOGIC AND HISTory PARa...

Страница 94: ...ion Use EDIT x to specify the desired search zone HISTory ZONE EDIT x HORizontal Function Sets the horizontal position of the zone of the history search zone or queries the current setting Syntax HIST...

Страница 95: ...Ment Space SEND EXECute TONE Space COLor OFF GRAY REVerse COMPression Space OFF ON NRf FORMat Space BMP TIFF PSCript Space NRf SAVE ANAMing ON OFF Space CDIRectory Space Filename NAME Space DRIVe NETW...

Страница 96: ...ames or queries the current setting Syntax IMAGe ANAMing Boolean IMAGe SAVE ANAMing Example IMAGE ANAMING ON IMAGE SAVE ANAMING IMAGE SAVE ANAMING 1 IMAGe SAVE CDIRectory Function Changes the director...

Страница 97: ...itializing You can make the same settings using the Initialize menu of SETUP on the front panel INITialize EXECute UNDO INITialize EXECute Function Executes initializing Syntax INITialize EXECute Exam...

Страница 98: ...ON NRf FFT POINts Space NRf SPOint Space NRf WINDow Space RECTangle HANNing FLATtop SMOothing Space OFF ON NRf LABel Space Character string INTegral SPOint Space NRf OPERation Space PLUS MATH1 NRf NRf...

Страница 99: ...SPlay Function Sets whether or not to display computation waveforms or queries the current setting Syntax MATH x DISPlay Boolean MATH x DISPlay x 1 or 2 1 on the DL1720E Example MATH1 DISPLAY ON MATH1...

Страница 100: ...ng Syntax MATH x SCALe CENTer NRf MATH x SCALe CENTer x 1 or 2 1 on the DL1720E NRf 9 9999E 30 to 9 9999E 30 Example MATH1 SCALE CENTER 1 MATH1 SCALE CENTER MATH1 SCALE CENTER 1 00000E 00 MATH x SCALe...

Страница 101: ...een may not match the names used by communication commands A correspondence table is given in Appendix 4 MEASure CHANnel MATH x x ALL Space OFF ON NRf COPY DELay COUNt SDEViation MAXimum MEAN MEASure...

Страница 102: ...STATistics CYCLe HISTory TRANge Space NRf NRf CONSt X Space NRf X COUNt SDEViation MAXimum MEAN MINimum STATe Space OFF ON NRf VALue CYCLe ABORt EXECute x TRACe OWN MATH Space NRf HISTory ABORt EXECut...

Страница 103: ...le MiNus PLUS DIVision SDEVitation STATe Space ON OFF Character string UNIT Space USER X Space NRf AREA X TRACe COUNt MAXimum MEAN MINimum NONE x MATH Space Parameter ITEM x CONSt When the target chan...

Страница 104: ...MEASURE CHANNEL2 RMS STATE 0 MEASURE CHANNEL2 SDEVIATION STATE 0 MEASURE CHANNEL2 TY1INTEG STATE 0 MEASURE CHANNEL2 TY2INTEG STATE 0 MEASURE CHANNEL2 XY1INTEG STATE 0 MEASURE CHANNEL2 XY2INTEG STATE 0...

Страница 105: ...MEASURE MATH2 BWIDTH2 STATE 0 MEASURE MATH2 DUTYCYCLE STATE 0 MEASURE MATH2 FALL STATE 0 MEASURE MATH2 FREQUENCY STATE 0 MEASURE MATH2 HIGH STATE 0 MEASURE MATH2 LOW STATE 0 MEASURE MATH2 MAXIMUM STA...

Страница 106: ...ction Copies the ON OFF state of all measurement parameters of one waveform to all other waveforms Syntax MEASure CHANnel x MATH x COPY x of CHANnel x 1 to 4 1 to 2 on the DL1720E x of MATH x 1 or 2 1...

Страница 107: ...ANNEL1 DELAY REFERENCE SLOPE MEASURE CHANNEL1 DELAY REFERENCE SLOPE RISE MEASure CHANnel x MATH x DELay REFerence SOURce Function Sets the reference of the delay measurement between waveforms to the t...

Страница 108: ...e CHANnel x MATH x DPRoximal UNIT Function Sets the distal mesial and proximal values of the automated measurement of waveform parameters in terms of a voltage or current or queries the current settin...

Страница 109: ...on the DL1720E Parameter AVERage AVGFreq AVGPeriod BWIDth1 BWIDth2 DUTYcycle FALL FREQuency HIGH LOW MAXimum MINimum NOVershoot NWIDth PERiod PNUMber POVershoot PTOPeak PWIDth RISE RMS SDEViation TY1I...

Страница 110: ...HISTory EXECute Function Executes the statistical processing within the historical data of the automated measurement of waveform parameters History Statistics Syntax MEASure HISTory EXECute Example ME...

Страница 111: ...f MATH z NONE CONSt w MEASure USER x AREA y TRACe x 1 to 4 y 1 or 2 NRf 1 to 4 1 to 2 on the DL1720E z 1 or 2 1 on the DL1720E w 1 to 4 Example MEASURE USER1 AREA1 TRACE 1 MEASURE USER1 AREA1 TRACE ME...

Страница 112: ...omitted If omitted the newest parameter in history is queried If NRf is appended queries the parameters from the newest waveform in history memory to the NRf waveform MEASure WAIT Function Waits for...

Страница 113: ...00 CHANNEL8 0 0E 00 PHASe CHANnel x Function Sets the amount of phase shift of the waveform or queries the current setting Syntax PHASe CHANnel x time PHASe CHANnel x x 1 to 4 1 to 2 on the DL1720E ti...

Страница 114: ...the front panel is pressed PRESet ALL PROBe Space NRf C10 C100 TYPE Space CMOS5 CMOS3 ECL USERdefine USERdefine LEVel Space OFFSet Space VDIV Space CHANnel x PROBe Space NRf C10 C100 TYPE Space CMOS5...

Страница 115: ...hannels Syntax PRESet ALL USERdefine VDIV voltage voltage 2 mV to 10 V when the probe attenuation is 1 1 See the User s Manual Example PRESET ALL USERDEFINE VDIV 2V Description When set to the current...

Страница 116: ...value or queries the current setting Syntax PRESet CHANNEL x USERdefine VDIV voltage PRESet CHANNEL x USERdefine VDIV x 1 to 4 1 to 2 on the DL1720E voltage 2 mV to 10 V when the probe attenuation is...

Страница 117: ...These commands can be used to carry out the same operations as when SETUP on the front panel is pressed RECall SETup X EXECute RECall SETup x EXECute Function Recalls the setting parameters from the...

Страница 118: ...o make the same settings and inquiries as when SHIFT ZOOM on the front panel is pressed SEARch EDGE COUNt Space NRf HYSTeresis Space NRf LEVel Space Voltage SLOPe Space RISE FALL SOURce Space NRf MATH...

Страница 119: ...oltage Voltage NRf CLOCk HYSTeresis Space NRf LEVel Space SLOPe Space FALL RISE SOURce Space NONE NRf SPATtern CLEar CLOCk HYSTeresis Space NRf LEVel Space SLOPe Space FALL RISE SOURce Space NONE NRf...

Страница 120: ...e SPATtern EDGE WIDTh PPATtern ASCRoll WIDTh HYSTeresis Space NRf NRf TIME x Space Time LEVel Space POLarity Space HIGH LOW MATH OUT IN SOURce Space NRf x Voltage TYPE Space TIMeout BETWeen When the t...

Страница 121: ...on Example SEARCH ASCROLL DIRECTION LEFT SEARCH ASCROLL DIRECTION SEARCH ASCROLL DIRECTION LEFT SEARch ASCRoll SPEed Function Sets the speed for moving the zoom box of the auto scroll when using the s...

Страница 122: ...POSition x x 1 or 2 NRf 5 to 5 div 10 div displayed record length steps Example SEARCH POSITION1 2 5 SEARCH POSITION1 SEARCH POSITION1 2 5000000 SEARch PPATtern Function Queries all settings related t...

Страница 123: ...channel of the parallel pattern search using the search zoom function or queries the current setting Syntax SEARch PPATtern CLOCk HYSYeresis NRf SEARch PPATtern CLOCk HYSYeresis NRf 0 3 to 4 div 0 1...

Страница 124: ...he clock channel of the serial pattern search using the search zoom function or queries the current setting Syntax SEARch SPATtern CLOCk HYSTeresis NRf SEARch SPATtern CLOCk HYSTeresis NRf 0 3 to 4 di...

Страница 125: ...32 to 39 x 6 character string bits 40 to 47 x 7 character string bits 48 to 55 x 8 character string bits 56 to 63 The pattern is assigned the pattern number that is selected with the SEARch SPATtern...

Страница 126: ...LOW SEARch WIDTh POLarity Example SEARCH WIDTH POLARUTY HIGH SEARCH WIDTH POLARITY SEARCH WIDTH POLARITY HIGH SEARch WIDTh SOURce Function Sets the target waveform of the pulse width search of the sea...

Страница 127: ...TART and wait 0 START only No wait 36000 to 1 100 ms resolution wait period START and wait Example SSTART 100 SSTART 0 Description If the specified time period is positive data acquisition is started...

Страница 128: ...TER14 NEVER FILTER15 NEVER FILTER16 NEVER QENABLE 1 QMESSAGE 1 STATus CONDition Function Queries the contents of the condition register and clears the register Syntax STATus CONDition Example STATUS C...

Страница 129: ...STATus Group 5 26 STOP Group STATus QMESsage Function Selects queries whether message content is appended to responses to STATus ERRor Syntax STATus QMESsage Boolean STATus QMESsage Example STATUS QME...

Страница 130: ...are to be stored to the internal memory at the specified number or queries the current setting Syntax STORe SETup x COMMent Character string STORe SETup x COMMent x 1 to 3 Character string Within 8 c...

Страница 131: ...F ON NRf LANGuage Space CHINese ENGLish MLANGuage Space CHINese ENGLish JAPANese LCD AUTO BRIGhtness Space NRf MODE Space Time Space OFF ON NRf OCANcel Space OFF ON NRf BATTery MODE TIME Space OFF ON...

Страница 132: ...MTTIME 09 00 SYSTem CLOCk SNTP EXECute Functions Execute date and time setting per SNTP server Syntax SYSTem CLOCk SNTP EXECute Example SYSTEM CLOCK SNTP EXECUTE SYSTem CLOCk SNTP GMTTime Functions Se...

Страница 133: ...se SYSTem MLANguage Example SYSTEM MLANGUAGE ENGLISH SYSTEM MLANGUAGE SYSTEM MLANGUAGE ENGLISH SYSTem OCANcel Offset Cancel Function Sets whether or not to reflect the offset voltage to the measured r...

Страница 134: ...se Example TIMEBASE TIMEBASE SOURCE INTERNAL TDIV 500 0E 06 TIMebase SOURce Function Sets the time base or queries the current setting Syntax TIMebase SOURce EXTernal INTernal TIMebase SOURce Example...

Страница 135: ...oup keys SIMPLE and ENHANCED and LEVEL nob and POSITION DELAY TRIGger ABN COUNt Space NRf ACTion BUZZer Space OFF ON NRf STARt STOP SAVE Space OFF ON NRf HCOPy Space OFF ON NRf IMAGe Space OFF ON NRf...

Страница 136: ...FALL IN OUT OFF WINDow Space OFF ON NRf HOLDoff TIME Space Time Time BTRigger CHANnel x Space DONTcare HIGH LOW CONDition Space ENTer EXIT DELay Space LOW CLOCk SLOPe Space FALL RISE SOURce Space NRf...

Страница 137: ...LEVel Space SLOPe Space BOTH FALL RISE SOURce Space NRf EXTernal LINE SOURce CHANnel x Voltage Frequency Frequency COUPling Space AC DC HFRejection Space OFF PROBe Space NRf HYSTeresis Space HIGH LOW...

Страница 138: ...I1080 24 I1080 50 I1080 60 P1080 24 P1080 25 P1080 60 P480 60 P720 60 WIDTh CHANnel x Space HIGH LOW IN OUT DONTcare CONDition Space TRUE FALSe TIME x Space OUT BETWeen TIMeout WINDow Space OFF ON NR...

Страница 139: ...OUNt Function Sets queries the number of times the pattern for the A B n trigger should become true Syntax TRIGger ABN COUNt NRf TRIGger ABN COUNt NRf 1 to 100000000 Example TRIGGER ABN COUNT 1 TRIGGE...

Страница 140: ...ns to 5 s Example TRIGGER ADB DELAY 1NS TRIGGER ADB DELAY TRIGGER ADB DELAY 1 0E 09 TRIGger ATRigger Function Queries all the pattern A settings Syntax TRIGger ATRigger Example TRIGGER ATRIGGER TRIGG...

Страница 141: ...EL1 RISE Description Select from IN OUT when TRIGger OR WINDow ON is specified and from FALL RISE otherwise TRIGger OR WINDow Function Turns the OR trigger window ON OFF or queries the current setting...

Страница 142: ...TRIGger SIMPle HFRejection OFF Frequency TRIGger SIMPle HFRejection Frequency 20 MHz 15 kHz Example TRIGGER SIMPLE HFREJECTION 20MHz TRIGGER SIMPLE HFREJECTION TRIGGER SIMPLE HFREJECTION 20 0E 06 Des...

Страница 143: ...RCE CHANNEL1 TRIGGER SOURCE CHANNEL1 LEVEL 0 0E 00 COUPLING DC HFREJECTION OFF HYSTERESIS LOW CENTER 0 0E 00 WIDTH 2 000E 00 TRIGger SOURce CHANnel x COUPling Function Sets the trigger coupling of eac...

Страница 144: ...s the current setting Syntax TRIGger TV LINE NRf TRIGger TV LINE NRf 5 to 1054 for NTSC 2 to 1251 for PAL or SECam 2 to 2251 other than NTSC PAL or SECam 8 to 2251 for 480 60P Example TRIGGER TV LINE...

Страница 145: ...DTh TLINkage Function Turns the link ON OFF when changing the pulse width for the pulse width trigger or queries the current setting Syntax TRIGger WIDTh TLINkage Boolean TRIGger WIDTh TLINkage Exampl...

Страница 146: ...RMat Space ASCii BYTE WORD LENGth OFFSet RANGe RECord RECord Space MINimum NRf SEND SIGN SRATe STARt Space NRf Space NRf TRACe Space NRf MATH x TRIGger TYPE MINimum WAVeform Function Queries all infor...

Страница 147: ...form LENGth Function Queries the number of records total number of data points in the waveform designated by WAVeform TRACe Syntax WAVeform LENGth Example WAVEFORM LENGTH WAVEFORM LENGTH 1002 WAVeform...

Страница 148: ...e record number specified by the WAVeform RECord command NRf in order WAVeform SIGN Function Queries the presence of the sign when querying the waveform specified with WAVeform TRACe in binary data fo...

Страница 149: ...XAXIS SINGLE DISPLAY 1 TRANGE 5 00 5 00 XTRACE 1 TLINKAGE 0 YTRACE 2 XY x DISPlay Function Sets whether or not to display the X Y waveform or queries the current setting Syntax XY x DISPlay Boolean X...

Страница 150: ...tion Sets the channel that is to be assigned to the X axis in the X Y display or queries the current setting Syntax XY x XTRace NRf MATH x XY x XTRace x of XY x 1 or 2 1 on the DL1720E x of MATH x 1 o...

Страница 151: ...OOM key on the front panel ZOOM ALLOcation ALL Space OFF ON NRf CHANnel x Space OFF ON NRf MATH x Space OFF ON NRf FITMeasure Space Z1 Z2 FORMat Space MAIN SINGle DUAL TRIad QU AD HEXa MLINkage Space...

Страница 152: ...MEAURE Z1 ZOOM FITMEAURE ZOOM FITMEAURE Z1 ZOOM FORMat Function Sets queries the zoom display format Syntax ZOOM FORMat MAIN SINGle DUAL TRIad QUAD HEXa ZOOM FORMat Example ZOOM FORMAT SINGLE ZOOM FOR...

Страница 153: ...Example CLS Description The output queue will also be cleared if a CLS command is appended after the program message terminator For details of the registers and queue refer to Chapter 6 ESE standard E...

Страница 154: ...E MODE 0 AVALUE 1 00000E 00 BVALUE 0 00000E 00 UNIT CHANNEL3 LABEL CH3 CHANNEL4 DISPLAY 1 VDIV 50 000E 00 POSITION 0 00 COUPLING DC PROBE 10 OFFSET 0 0E 00 BWIDTH FULL VARIABLE 50 000E 00 LSCALE MODE...

Страница 155: ...ires about the current setting Syntax SRE NRf SRE NRf 0 to 255 Example SRE 239 SRE 239 Description NRf is the sum of the bits expressed as a decimal number For example if ESE 239 is set the service re...

Страница 156: ...12 11 10 9 8 7 6 5 4 3 2 1 0 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 Condition Filter Transit Filter Extended Event Register 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0...

Страница 157: ...ESE query is used to check whether bits in that register are set to 1 or 0 For details of these commands refer to Chapter 5 6 2 Status Byte Overview of Status Byte 7 6 ESBMAVEES EAV 1 0 RQS MSS Bits 0...

Страница 158: ...n the status byte are cleared except bit 4 MAV since the output queue cannot be emptied by the CLS command However the output queue will also be cleared if the CLS command is received just after a pro...

Страница 159: ...MES 7 5 4 3 2 1 0 PRN TST CAL TRG HLD RUN HST 8 SUP 9 NGO 10 SCH 11 NSG 12 INI 13 FFT 14 0 15 6 7 5 4 3 2 1 0 8 9 10 11 12 13 14 15 Condition Register STATus CONDition Transition Filter 0 14 15 Exten...

Страница 160: ...ored record by record in the output queue and is read out oldest item first newest item last The output queue is emptied in the following cases in addition to when read out is performed When a new mes...

Страница 161: ...rface Conforms to USB Rev 1 0 Setting up the Visual Basic Standardmodule tmctl bas tmval bas Setting up the DL1720E DL1740E DL1740EL GP IB All the sample programs given in this chapter use address 1 f...

Страница 162: ...7 2 IM 701730 17E 7 2 Image of Sample Program 7 2 Image of Sample Program...

Страница 163: ...ot Terminator LF EOI Enable If sts CTL_OK Then Call DisplayError Dev InitGpib 1 GoTo PostProcces End If sts TmSetTimeout Dev timeout Timeout 10s If sts CTL_OK Then Call DisplayError Dev InitGpib 1 GoT...

Страница 164: ...8192 Then ers Illegal device ID Else ers End If MsgBox Error No Str err Chr 13 ers vbExclamation Error End Sub Private Sub Command1_Click Dim sts As Long If StartFlag 1 Then Exit Sub End If StartFlag...

Страница 165: ..._Click Dim sts As Long If StartFlag 1 Then Exit Sub End If StartFlag 1 Text1 Text START List1 Clear sts UsbGetWord Run Sample5 USB Get Word Data If sts 0 Then List1 AddItem END End If Text1 Text END S...

Страница 166: ...UsbTdiv 1 GoTo UsbTdivEnd End If UsbTdiv 0 UsbTdivEnd Call CloseDevice Device Close End Function Private Function GetTdiv As Long Dim msg As String Command buffer Dim qry As String Query biffer Dim st...

Страница 167: ...Sample Program 7 7 IM 701730 17E 7 GP IB USB 7 4 Sets Queries the T Div...

Страница 168: ...nd If UsbGetWord 0 UsbGetWordEnd Call CloseDevice Device Close End Function Private Function GetWord As Long Dim msg As String Command buffer Dim qry As String Query biffer Dim sts As Long Dim vdv As...

Страница 169: ...End If sts TmReceive Dev qry Len qry rlen If sts CTL_OK Then Call DisplayError Dev GetWord 1 Exit Function End If vdv Val qry msg WAVEFORM OFFSET Get Offset value sts TmSend Dev msg If sts CTL_OK Then...

Страница 170: ...30 17E msg COMMUNICATE HEADER ON Query Header On sts TmSend Dev msg If sts CTL_OK Then Call DisplayError Dev GetWord 1 Exit Function End If GetWord 0 End Function GP IB USB 7 5 Data Output in Word For...

Страница 171: ...1 Exit Function End If msg COMMUNICATE HEADER OFF Query Header Off for Get V div sts TmSend Dev msg If sts CTL_OK Then Call DisplayError Dev GetMeasure 1 Exit Function End If msg MEASURE MODE OFF Meas...

Страница 172: ...v GetMeasure 1 Exit Function End If msg MEASURE TRANGE 5 5 Measure Time Range 5 5 sts TmSend Dev msg If sts CTL_OK Then Call DisplayError Dev GetMeasure 1 Exit Function End If msg SSTART 500 Start Sin...

Страница 173: ...1 Exit Function End If sts TmReceive Dev qry Len qry rlen If sts CTL_OK Then Call DisplayError Dev GetMeasure 1 Exit Function End If Query 1 Average Left qry rlen 1 msg MEASURE CHANNEL1 FREQUENCY VAL...

Страница 174: ...70 71 72 73 74 75 76 77 78 79 7A 7B 7C 7D 7E 7F 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53...

Страница 175: ...ge Check the header Chapter 5 120 Numeric data error Numeric value must be entered for NRf format 4 5 123 Exponent too large Use a smaller exponent for NR3 format 4 5 Chapter 5 124 Too many digits Lim...

Страница 176: ...88 2 278 Macro header not found Does not conform to the macro function specified in IEEE488 2 Error in Communications Query 400 to 499 Code Message Action Reference Page 410 Query INTERRUPTED Check tr...

Страница 177: ...urn Multiple Response Messages Refer to Chapter 5 Command List c Queries which Generate Response Data during Analysis of the Syntax Every query generates a response data when analysis of the syntax is...

Страница 178: ...Size of Storage Area for Protected User Data for PUD and PUD PUD and PUD are not supported 18 Length of RDT and RDT Resource Name RDT and RDT are not supported 19 Change in Status Due to RST LRN RCL...

Страница 179: ...ured Results P P PTOPeak P P Max MAXimum Max Min MINimum Min Avg AVERage Avg Rms RMS Rms Sdev SDEViation SDv High HIGH High Low LOW Low OShot POVershoot Ovr OShot NOVershoot Ovr Freq FREQuency Freq Pe...

Страница 180: ...tus 5 88 current setting 5 119 current to voltage conversion ratio 5 21 cursor X Y display 5 30 CURSor Group 5 25 cursor measurement 5 27 cycle statistics 5 75 D data 4 1 4 5 data saving 5 42 data tra...

Страница 181: ...roup 5 79 print external printer 5 53 print of screen image data 5 53 print operation 5 53 probe 5 107 probe attenuation 5 21 program message 4 1 program message unit 4 1 pulse width search 5 90 pulse...

Страница 182: ...zoomed waveform 5 117 Command List ACQuire Group ACQuire 5 16 AVERage COUNt 5 16 AVERage EWEight 5 16 AVERage 5 16 CLOCk 5 16 COUNt 5 16 INTerleave 5 16 MODE 5 16 REPetitive 5 16 RLENgth 5 16 ASETup G...

Страница 183: ...RGB MENU BACK DIALog FOCus FORE LINE SELect SHADow 5 35 RGB MENU 5 35 RGB OTHers CURSor GRID MARKer SNAP ZONE x 5 36 RGB OTHers 5 35 RGB WAVeform 5 36 RGB 5 35 SMAPping CHANnel x MATH x 5 37 SMAPping...

Страница 184: ...5 59 ZONE LOGic 5 59 ZONE 5 59 IMAGe Group ABORt 5 61 COMMent 5 61 COMPression 5 61 EXECute 5 61 FORMat 5 61 IMAGe 5 61 SAVE ANAMing 5 61 SAVE CDIRectory 5 61 SAVE DRIVe 5 61 SAVE NAME 5 61 SAVE 5 61...

Страница 185: ...8 PPATtern CHANnel x MATH x PATTern 5 88 PPATtern CHANnel x MATH x 5 87 PPATtern CLOCk HYSYeresis 5 88 PPATtern CLOCk LEVel 5 88 PPATtern CLOCk SLOPe 5 88 PPATtern CLOCk SOURce 5 88 PPATtern CLOCk 5 8...

Страница 186: ...Rejection 5 108 SOURce CHANnel x HYSTeresis 5 108 SOURce CHANnel x LEVel 5 108 SOURce CHANnel x WIDTh 5 109 SOURce CHANnel x 5 108 SOURce 5 108 TRIGger 5 104 TV FIELd 5 109 TV FRAMe 5 109 TV LEVel 5 1...