SM 04L01A01-01E
4-5
4
ADJUSTMENTS
4.2.2 PC Controlled Adjustment
You can also adjust the DX using a PC.
Connection
UUA
When DC range
DC Voltage
standard
DC Voltage
standard
DECADE
RESISTANCE
BOX
DECADE
RESISTANCE
BOX
GROUP 1
CH2 CH1
+
-
A
H
L
H
L
b
B
A
b
B
+
-
UUA
When RTD or Cu range
GROUP 1
CH2
CH1
Commands
The commands below are used for adjustment.
Change mode
Adjustment
executing
Calibration value
saving
Calibration value
confirmation
Calibration value
editing
DSp1
(Ex) DS2
XZp1,p2,p3
(Ex) XZ01-02,CAL/EXEC,20mV
XZp1,p2,p3
(Ex) XZ01-02,END,STORE
XZp1,p2?
(Ex) XZ01-02,DISPLAY? or
XZp1,p2,p3?
(Ex) XZ01-02,DISPLAY,20mV?
XZp1,p2,p3,o4,p5
(Ex) XZ01-02,DISPLAY,Pt100,0,-32768
Mode*
Group*
Group*
Group*
Group*
CAL/
EXEC
END
DISPLAY
DISPLAY
Range*
Way*
Range*
Range*
Zero
calibrated
value
*1: 0= Change to normal operation, 1= Change to setup mode, 2= Change to adjustment mode
*2: 01-02, 03-04, 05-06, 07-08, 01-06, 01-10, 11-20, or 21-30.
*3: 20mV, 60mV, 200mV, 1V, 2V, 6V, 20V, Pt100, Cu10, or Cu25
*4: STORE (: saved), ABORT (: don’t saved)
FS
calibrated
value
Parameters
Command
contents
Syntax
P1
P2
P3
P4
P5
1
2
2
2
2
3
3
3
4
4.2 AD Board Offset and Gain Adjustment