13
3.4 OUTPUT Terminal (For current ranges only)
WARNING
● Never use the instrument on a circuit above 750VAC or 1000VDC.
This may cause electrical shock hazard and damage to the instrument or
the circuit under test.
● Never apply voltage to the OUTPUT terminal.
When the plug is inserted into the output terminal, auto-range function is
cleared.
Set the range depending on the state of Data Hold Switch.
Data Hold Switch OFF
400A range
Data Hold Switch ON
2000A range
NOTE
After measurement, be sure to return Data Hold switch to "OFF" position.
(1) Attach the output plug cable (sold separately : 98076 or 98077) to a
connection lead so that the output voltage can be connected to a
recording device such as a chart recorder.
(2) Slide the terminal cover to the right to disclose the OUTPUT terminal
and insert the output plug into the terminal. Make connection to the
recording device.
(3) Set the Function Selector switch to the desired position (ACA or DCA)
and follow appropriate measurement instructions.