background image

IM CL255 

5

NOTE

●  During  current  measurement,  keep  the  transformer  jaws  fully  closed. 

Otherwise,  accurate  measurement  cannot  be  made.  The  maximum 
measurable conductor size is approx. 55mm in diameter.

●  When  the  current  flows  from  the  upside  (the  display  side)  to  the 

underside  of  the  instrument,  the  polarity  of  the  reading  is  positive  and 
vice versa.

●  The 

  button  may  not  completely  zero  adjust  the  output  voltage 

from  the  OUTPUT  terminal.  In  this  case,  make  zero  adjustment  on  the 
recording device.

●  Turing  the  Function  Selector  switch  to  a  position  other  than  DCA 

cancels the zero adjustment.

2.3 AC Current Measurement

 

WARNING

●  Never use the instrument on a circuit above 750VAC. This may cause 

electrical  shock  hazard  and  damage  to  the  instrument  or  the  circuit 
under test.

●  Do  not  make  measurement  with  the  battery  compartment  cover 

removed.

●  Keep your fingers and hands behind the barrier during measurement.

(1)  Set the Function Selector switch to the "  A " position. "AC" should be 

shown on the upper left corner of the display.

(2)  Press  the  open/close  lever  to  open  the  transformer  jaws  and  clamp 

them onto a single conductor and take the reading on the display. The 
most accurate reading will be obtained by keeping the conductor at the 
center of the transformer jaws.

Содержание CL255

Страница 1: ...IM CL255 2018 6 12 KYOU IM CL255 User s Manual CL255 Clamp on Tester...

Страница 2: ...Voltage Measurement 7 2 6 Resistance Measurement 8 2 7 Continuity Check 400 range fixed 8 2 8 Frequency Measurement 9 2 9 Peak Measurement 10 2 10 Average Measurement 11 3 Other Functions 12 3 1 Sleep...

Страница 3: ...anual in order to avoid the risk of personal injury or death and or damage to the instrument Double Insulation This symbol indicates double insulation AC Voltage Current This symbol indicates AC volta...

Страница 4: ...d to the circuit under test Do not install substitute parts or make any modification to the instrument Return the instrument to Yokogawa Meters Instruments or your distributor for repair or re calibra...

Страница 5: ...iate terminal on the instrument Make sure to remove the test leads from the instrument before making current measurement Be sure to set the Function Selector switch to the OFF position after use When...

Страница 6: ...n 4 Data Hold Button Freezes the display reading is shown on the display when Data Hold is enabled Note When the plug is inserted into the output terminal Data Hold Switch operates as range selection...

Страница 7: ...the selected function and measuring mode 8 Terminal Cover Slides over Hi and Lo Terminals to prevent access to them when OUTPUT terminal is in use 9 OUTPUT Terminal For AC or DC current range only Pro...

Страница 8: ...slipping off the hand during use 13 Test Leads Model 98072 Connected to Lo and Hi terminals for voltage or resistance measurement 14 Barrier It is a part providing protection against electrical shock...

Страница 9: ...r equipment under test Do not make current measurement with the test leads connected to the Hi and Lo terminals Keep your fingers and hands behind the barrier during measurement 1 Set the Function Sel...

Страница 10: ...o a position other than DCA cancels the zero adjustment 2 3 AC Current Measurement WARNING Never use the instrument on a circuit above 750VAC This may cause electrical shock hazard and damage to the i...

Страница 11: ...ING Never use the instrument on a circuit above 1000VDC This may cause electrical shock hazard and damage to the instrument or the circuit under test Keep your fingers and hands behind the barrier dur...

Страница 12: ...shown on the upper left corner of the display 2 Slide the terminal cover to the left to disclose the Hi and Lo terminals Plug the red test lead into Hi terminal and the black test lead into the Lo ter...

Страница 13: ...leads to the circuit under test and take the reading on the display 2 7 Continuity Check 400 range fixed The continuity check mode is enabled by pressing the button on resistance range and is indicate...

Страница 14: ...ns that measurement cannot be made Because range is fixed at 400V for AC voltage and at 400A for AC current WARNING Never use the instrument on a high voltage circuit above 750VAC This may cause elect...

Страница 15: ...plugged into the instru ment 1 The PEAK mode is available on DCA ACA DCV and DCA ranges Set the Function Selector switch to the A A V or V position Note Only on DCA range press the button for about o...

Страница 16: ...tion Selector switch to the A A V or V position 2 Press the button once to enter from the normal mode to the Average mode should be indicated on the display 3 Follow instructions for ACV DCV ACA or DC...

Страница 17: ...disabled in the PEAK measurement mode 3 2 Data Hold Function This is a function used to freeze the measured value on the display Press the Data Hold button to freeze the reading The reading will be he...

Страница 18: ...Switch Data Hold Switch OFF 400A range Data Hold Switch ON 2000A range NOTE After measurement be sure to return Data Hold switch to OFF position 1 Attach the output plug cable sold separately 98076 o...

Страница 19: ...e of the OUTPUT terminal use an Alkaline battery which will extend continuous recording time up to about 24 hours 3 5 Optional Accessories Clamp Adapter Model 99025 For AC current measurement only The...

Страница 20: ...If the battery voltage becomes too low for the instrument to operate normally is shown on the display Then replace the battery Note that when the battery is completely exhausted the display blanks wi...

Страница 21: ...ranging Accuracy 5 0 to 399 9A 1 5 rdg 3dgt 50 60Hz 3 0 rdg 4dgt 30 to 1kHz 150 to 1700A 1701 to 2000A 3 5 rdg 3dgt 50 60Hz Conversion method AC coupled true rms responding calibrated to the rms Auto...

Страница 22: ...nce Measuring Range Auto ranging Accuracy 0 to 399 9 1 5 rdg 2dgt 150 to 3999 Auto ranging Lo 0 to 399 9 Shifts to Hi at 400 0 Hi 150 to 3999 Shifts to Lo at 149 At 4000 or above OL is shown Continuit...

Страница 23: ...clamp meter is re sponding to the average value The calibration is done with a conversion factor of 1 111 for sine wave which is found by dividing the effective value by the average value These instr...

Страница 24: ...n Approx 15mA max Sleep function Automatically powered down in Approx 30 minutes after the last switch operation power consumption Approx 200 A Withstanding Voltage 6300V AC 50 60Hz for 5sec between e...

Страница 25: ...fails to operate correctly and needs repair contact the vendor from whom you purchased the instrument or the nearest Yokogawa Meters Instruments sales office Turn off the POWER switch once then turn...

Страница 26: ...IM CL255 21 This manual is valid only in China GB T 26572 GB T 26572...

Страница 27: ......

Отзывы: