Setting and Executing Program Sweeps (CA500)
3.
Select
Program Sweep Setup
, and press
ENTER
.
CA500
Turn data save on or off.
Set the interval time.
Setting the Interval Time
4.
Select
Interval Time
, and press
ENTER
. A list of options appears.
5.
Use the arrow keys to set the interval time, and then press
ENTER
.
Turning Data Save On and Off
6.
Select
Data Save
. ON and OFF appear in the selection menu.
7.
Use the arrow keys to set data save to ON or OFF.
When you set this to on, the measurement values and source values are saved automatically
when the sweep is completed.
Setting the Source Value
8.
Use the cursor keys to display page 2/2 of Program Sweep Setup.
9.
Select No.1 under Output Data. ON and OFF appear.
10.
Use the arrow keys to select ON, and then press
ENTER
. The settings are displayed at the
bottom of the screen.
11.
Use the arrow keys to select the source value (%), and then press
ENTER
.
12.
Likewise, set the source values of Output Data No.2 to No.10.
For the numbers you will not use, use the arrow keys to select OFF. “-” is displayed for the
source value.
If you set a number to off, the instrument sweeps up to the last number that has a source value
assigned. Then, it returns to No.1, and continues the signal generation.
OFF
2.9 Sweep Source
2-23
IM CA500-01EN