Thermoelectromotive Force of a Thermocouple
Thermoelectromotive force corresponding to the following temperatures is generated for each
thermocouple type.
Thermocouple
Source Range
Notes
K
-200.0°C to +1372.0°C
IEC 60584-1
*, **
E
-250.0°C to +1000.0°C
IEC 60584-1
*, **
J
-210.0°C to +1200.0°C
IEC 60584-1
*, **
T
-250.0°C to +400.0°C
IEC 60584-1
*, **
N
-200.0°C to +1300.0°C
IEC 60584-1
*
L
-200.0°C to +900.0°C
DIN 43710
U
-200.0°C to +600.0°C
DIN 43710
R
-20.0°C to +1767.0°C
IEC 60584-1
*, **
S
-20.0°C to +1768.0°C
IEC 60584-1
*, **
B
+600.0°C to +1820.0°C
IEC 60584-1
*, **
C
0.0°C to +2315.0°C
IEC 60584-1
*
XK
-200.0°C to +800.0°C
GOST R 8.585-2001
A
0.0°C to +2500.0°C
IEC 60584-1
D (W3Re/W25Re) 0.0°C to +2315.0°C
ASTM E1751/E1751M-09e1
G (W/W26Re)
+100.0°C to +2315.0°C
ASTM E1751/E1751M-09e1
PLATINEL II
0.0°C to +1395.0°C
ASTM E1751/E1751M-09e1
PR20-40
0.0°C to +1888.0°C
ASTM E1751/E1751M-09e1
*: Complies also with JIS C 1602
**: The setting can be changed to comply with IPTS-68 (JIS C 1602 1981).
Temperature Scale
This instrument complies with ITS-90 and IPTS-68.
Connection Terminals
This instrument is equipped with the following two types of terminals.
TC-A (TC mini plug)
A thermocouple is connected to the instrument using a thermocouple mini plug set,
sold separately. Reference junction compensation using an external RJ sensor (sold
separately) is not possible.
TC-B (banana plug)
Reference junction compensation using the internal RJ sensor or an external RJ sensor is
possible.
Reference Junction Compensation
The instrument measures the temperature of the contact using an RJ sensor and generates a
voltage based on the temperature component. This instrument can perform reference junction
compensation using the internal RJ sensor or an external RJ sensor.
When an instrument with a built-in reference junction temperature compensation is to be
calibrated, the reference junction temperature of the device to be calibrated is measured using
an external RJ sensor.
H
H
L
L
CA500/CA550
Calibration target
External RJ sensor
Source/measurement
lead cable
H
H
L
L
CA500/CA550 Calibration target
Thermocouple
Calibrating only the thermometer Calibrating including the thermocouple
Internal RJ sensor or
external RJ sensor
H
H
L
L
Thermocouple
Internal RJ sensor
1.2 Source Function