Safety Precautions

The handling precautions, component names, and specifications are covered in the separate Getting 

Started Guide (IM CA310-02EN).

Be sure to read the Getting Started Guide before using this instrument.

Basic Setup

1. 

Turn the instrument on, and set the rotary switch to 

SETTING

.

2. 

Use the left-most up and down keys to select the setting and the right-most up and down keys to 

switch the options.

3. 

Press OUTPUT ON/OFF to confirm the setting.

 

POWER

Selects the setting Selects an option

Confirms the settings

Setting

Auto power-off

Backlight

Sweep time

HART/BRAIN 

Communication 

resistance

Initialization of 

span settings

Initialization of 

settings

Options

ON

*

/OFF

ON/OFF

*

/TIMER

15s

*

/30s/45s/60s

ON/OFF

*

Display

*: Default value

OUTPUT

ON/OFF

Auto power-off

  If this is set to ON when the instrument is running on batteries, the power automatically turns off 

if you do not operate the instrument for approx. 20 minutes. This is invalid when the instrument is 

running off of the AC adapter.

Backlight

  Turns the backlight on and off. If set to Timer, the backlight turns on for approx. 2 minutes every 

time you operate a key.

Sweep time

  When sweeping in steps, set the output duration of each step. When sweeping linearly, set the 

output duration for the value to change from 0% to 100%.

HART/BRAIN communication resistance

  If you want to connect the 250 Ω communication load resistance to a HART or BRAIN 

communication system, turn this on.

Initialization of Span Settings

  Initializes the span settings.

Initialization of Settings

  Initializes settings.

Measurement

 WARNING

•  Always turn off the power of the device under measurement before connecting 

it. Connecting or disconnecting lead cables without turning off the device under 

measurement is extremely dangerous.

•  Do not connect current circuits to the voltage input terminals. Likewise, do not 

connect voltage circuits to the current input terminals. Improper connection may 

cause electric shock or burns or may damage the device under measurement 

or the instrument.

French

 AVERTISSEMENT

  Avant de brancher l’appareil à mesurer, mettez-le hors tension. Il peut être très dangereux de 

brancher/débrancher les câbles de mesure sans mettre l’appareil à mesurer hors tension.

  Évitez en particulier de brancher un circuit de courant sur les bornes d’entrée de tension ou 

un circuit de tension sur les bornes d’entrée de courant. Un branchement inadéquat risquerait 

non seulement d’endommager le circuit ou l’appareil à mesurer ainsi que l’équipement, mais il 

pourrait également représenter un risque pour les personnes.

1. 

Set the CA310’s measure/source switch to 

MEASURE

.

2. 

Follow the instructions in the Getting Started Guide to connect the CA310 to the device to be 

measured using the included lead cable.

3. 

Use the rotary switch to set the range. In measurement mode, mA SIMULATE is invalid.

4. 

Turn on the device to be measured. The measurement appears on the CA310.

This manual explains how to use the Volt mA Calibrator.

Keep this manual in a safe place for quick reference in the event that a question arises.

LOOP POWER

  Apply a constant 24 VDC voltage to the transmitter to perform a loop test. The displayed value is 

current.

  Set the rotary switch to 

LOOP POWER

 and take measurements.

 Press 

OUTPUT ON/OFF

. 24 VDC voltage is applied. Press 

OUTPUT ON/OFF

 again to stop 

the application.

  

4-20mA

+

Two-wire transmitter

Distributor

+

HART Communicator

Sourcing

1. 

Set the CA310’s measure/source switch to 

SOURCE

.

2. 

Follow the instructions in the Getting Started Guide to connect the CA310 to the target device 

using the included lead cable.

3. 

Use the rotary switch to set the range. In source mode, LOOP POWER, 50 mA, and 50 V are 

invalid.

4. 

Follow the instructions for common operations in the Getting Started Guide, and set the output 

value with the ▼ and ▲ keys.

5. 

Press 

OUTPUT ON/OFF

. ON appears on the screen, and the output is turned on.

 

Press 

OUTPUT ON/OFF

 again to turn off the output.

 

If you change the range, the output turns off.

Note

•  If the load current in a voltage range or load voltage in a current range exceeds the maximum 

value in the specifications, the protection limiter is activated, and the output is turned off.

  To restore, set the load back to a normal condition, and press OUTPUT ON/OFF again.

•  The CA310 is calibrated excluding the voltage dip in the lead cable.

  Load current readings include an error resulting from the lead cable resistance.

mA SIMULATE

  If the range is set to mA SIMULATE, the CA310 operates as a two-wire transmitter simulator and 

sinks current applied to the Hi terminal from an external voltage source (distributor).

  

4-20mA

+

Two-wire transmitter

Distributor

+

Changing the Output Mode

1. 

Refer to “Keys” in the Getting Started Guide, and set the output mode.

 

Normal output (no display): The set value is output. The output value can be varied using the ▼ 

and ▲ keys.

 

Span check (

): In addition to normal output, the output can be varied in units of 25% of the 

span.

 

Step sweep( ): The output is varied automatically in steps of 25% of the span.

 

Linear sweep ( ): The output is varied linearly from 0% to 100%.

 

Press simultaneously to set the output mode.

Pauses or resumes sweeping

Output setting for span check

Increase or 

decrease by 

1 count

Increase or decrease 

by 100 counts

Increase or decrease in 

25% steps of span

Set to 100% of 

span

Set to 0% of 

span

Pressing this when sweeping is 

stopped or paused sets the output 

to the span 0% value.

Sweeping

  The output value is varied automatically between 0% and 100% of the specified span. There are 

two sweep modes: step and linear.

 

Time

Output

0%

100%

Sweep time

Time

Output

0%

100%

Sweep time

Step sweep

Linear sweep

1. 

Set the 0% and 100% of the span, which is the sweep range.

 

Set the span 0% value using the ▼ and ▲ keys and then press simultaneously the first and the 

second ▼ keys from the right. SET is displayed, and 0% is set.

 

Likewise, set the span 100% value using the ▼ and ▲ keys and then press simultaneously the 

first and the second ▼ keys from the right.

 

For details on key operation, see “Keys” in the Getting Started Guide.

2. 

Set the output mode to step sweep or linear sweep.

3. 

Press 

OUTPUT ON/OFF

. Sweeping will start.

 

Pressing 

OUTPUT ON/OFF

 again stops sweeping and also the output.

 

Pressing 

OUTPUT ON/OFF

 yet again resumes sweeping from the last output value.

 

Pressing the left-most ▼ key while sweeping pauses sweeping. The displayed value continues 

to be output while sweeping is paused. Pressing the ▼ key again resumes sweeping.

 

When sweeping is stopped or paused, pressing the left-most ▲ key sets the output to the span 

0% value.

CA310
Volt mA Calibrator
User’s Manual

IM CA310-01Z2 

1st Edition

1st Edition: November 2015 (YMI)

All Rights Reserved, Copyright © 2015, Yokogawa Meters & Instruments Corporation

Printed in Japan

IM CA310-01Z2