<5. Basic Operating Procedures>
49
IM 01E24A01-01EN
(2) Status icons
Icon
Contents
Icon
Contents
Write protect Invalid
Write protect Valid
Device Busy
Device Fault
Ready for microSD
card
Accessing microSD
card
Disable to access
microSD card
Uploading parameters
Downloading
parameters
Trend graph executing
BRAIN communication
HART communication
Modbus
communication
F
OUNDATION
fi eldbus
communication
System alarm occurs
Setting alarm occurs
Process alarm occurs
Information occurs
Warning occurs
Operation level:
Operator
Display Damping Valid
Operation level:
Specialist
Operation level:
Maintenance
(3) Data indication part
The process values are available to select 8 items
maximum on the display. It is possible to indicate 4 items
maximum on the display at the same time, and the rest 4
items are able to show by scrolling.
Table 5.2.1 Abbreviation table of process values to be
indicated on the display.
Abbreviation
Contents
FLP(*1)
Flow rate %
PRV(*1)
Process value
VEL(*1)
Flow velocity
VFL(*1)
Volumetric fl ow
MFL(*1)
Mass fl ow
FLB
Flow rate in % bar graph
TL1(*1)
Totalization value 1
TL2(*1)
Totalization value 2
TL3(*1)
Totalization value 3
TAG
Tag No. (PD tag for F
OUNDATION
fi eldbus type)
LTG
Long Tag (for HART communication type only)
COM
Communication protocol
ADH
Adhesion diagnostic Level (Alarm at Level 4)
AO1(*1)
Analog output value 1
TC1
Count value of totalizer 1
TC2
Count value of totalizer 2
TC3
Count value of totalizer 3
MOD
Information of Modbus communication setting
Display example (*2): R/001/19200/E/1
FR
AI1 Flow rate
FRP
AI1 Flow rate (%)
FRB
AI1 Flow rate (%Bar)
AI1
AI1.OUT
AI2
AI2.OUT
AI3
AI3.OUT
IT1
IT1.OUT
IT2
IT2.OUT
AR
AR.OUT
*1:
Available to display the online trend graph.
*2:
Explanation of display example
F0502.ai
Device address
Baud rate
Transmission mode
•
R: RTU
•
A: ASCII
Stop bit
•
1: 1 bit
•
2: 2 bit
Parity
•
N: None
•
O: Odd
•
E: Even
Example: R/001/19200/E/1
5.2.2 Basic Confi guration for Display
For paramter setting from display panel, confi gurable
parameters differ by the three operational levels specfi ed
in Table 5.2.2, and a passcode is needed to enter into
Setting mode. No passcode requires for “Operator”, and
a passcode corresponding to each level requires for
“Maintenance” or “Specialist”.
For parameter in details, read the user’s manual of
applicable communication type as listed in Table 1.1.
Table 5.2.2 Parameter setting from display panel and
operation level
Operation
Level
Reading
parameters
Writing parameters
Operator
All parameters
Parameters related with basic display
settings including display language.
Maintenance All parameters
Parameters allowed for Operator level.
Parameters related with Zero adjustment.
Specialist
All parameters All Parameters
The following parameters are available to “Operator” level
without passcode.
(1) Display Language Setting
Display Menu Path:
Device setup ► Language
The language on the display is set to “English” as default
at the factory shipment. Select the adequate language.
The selectable display language is different by the model
and suffi x code (display code) specifi ed when ordering.
Position of the display code:
Integral Type:
AXW
-
-
Remote Transmitter:
AXW4A-
AXG1A-G000
1
Display code
Selectable display language
1
English, French, German, Italian, Spanish,
Portuguese, Japanese, or Russian
2
English or Chinese
(2) Display Contrast Setting (shading)
Display Menu Path:
Device setup►Detailed setup►Display set►Optional confi g►Contrast
Available to change the contrast of the display.
Setting item
Contents
-5 to +5
Set the contrast of the display
(The value is small: Low, and the value is big: High)
Basic Operating Procedures