background image

Commands

Start communication: 10

  Starts communication with the MY600.

  Command format (reception and transmission)

Order Command ASCII Binary

Description

Size

1

STX

1

02H

Packet start (fixed value)

Write 02 in binary.

1

2

TYPE

1

‘0’

30H

Data packet type: small

1

3

SIZE

1

‘0’

30H

Packet size

“07” : 7 BYTES

2

4

2

‘7’

37H

5

CMD

1

‘1’

31H

Communication start command

“10”

2

6

2

‘0’

30H

7

CSUM

1

‘F’

46H

Checksum

“F8” 

2

8

2

‘8’

38H

9

ETX

1

03H

Packet terminator (fixed value)

Write 03 in binary.

1

End communication: 11

  Ends communication with the MY600.

  Command format (reception and transmission)

Order Command ASCII Binary

Description

Size

1

STX

1

02H

Packet start (fixed value)

Write 02 in binary.

1

2

TYPE

1

‘0’

30H

Data packet type: small

1

3

SIZE

1

‘0’

30H

Packet size

“07” : 7 BYTES

2

4

2

‘7’

37H

5

CMD

1

‘1’

31H

Communication end command

“11”

2

6

2

‘1’

31H

7

CSUM

1

‘F’

46H

Checksum

“F9” 

2

8

2

‘9’

39H

9

ETX

1

03H

Packet terminator (fixed value)

Write 03 in binary.

1

Start continuous data transmission: B1

  Retrieves measurement data in real-time until sending the continuous data transmission 

end command (B2).

  Command format (reception and transmission)

Order Command ASCII Binary

Description

Size

1

STX

1

02H

Packet start (fixed value)

Write 02 in binary.

1

2

TYPE

1

‘0’

30H

Data packet type: small

1

3

SIZE

1

‘0’

30H

Packet size

“07” : 7 BYTES

2

4

2

‘7’

37H

5

CMD

1

‘B’

42H

Data connection start command

“B1”

2

6

2

‘1’

31H

7

CSUM

1

‘0’

30H

Checksum

“0A” 

2

8

2

‘A’

41H

9

ETX

1

03H

Packet terminator (fixed value)

Write 03 in binary.

1

  Receive data in the following format

•  Voltage measurement:

  <model name>,VOLT,<site number 1>,<site number 2>,<voltage>,<voltage 

unit>,<DC±/AC/--><CRLF>

  If the voltage is less than 2 V, “--” is returned for <DC±/AC/-->.

 Example  MY600,VOLT,00,00,100,V,AC<CRLF>

  If the voltage is less than 2 V, “--” is returned for <DC±/AC/-->.

•  Insulation resistance measurement:

  <model name>,<range>,<site number 1>,<site number 2>,<insulation 

resistance>,<insulation resistance unit>,<elapsed time>,<1-minute value>,<1-minute 

value unit>,<DAR value>,<PI value>,<comparator><CRLF>

  Example  MY600,1000V,00,00,100.0,MΩ,00:10,----,--,----,----,PASS<CRLF>

  * The available six range are 50V, 100V, 125V, 250V, 500V, 1000V.

•  Low-resistance measurement 

  <model name>,CONT,<site number 1>,<site number 2>,<resistance>,<resistance 

unit>

  Example  MY600,CONT,00,00,100.0,Ω<CRLF>

End continuous data transmission: B2

  Ends the continuous retrieval of measurement data.

  Command format (reception and transmission)

Order Command ASCII Binary

Description

Size

1

STX

1

02H

Packet start (fixed value)

Write 02 in binary.

1

2

TYPE

1

‘0’

30H

Data packet type: small

1

3

SIZE

1

‘0’

30H

Packet size

“07” : 7 BYTES

2

4

2

‘7’

37H

5

CMD

1

‘B’

42H

Data connection end command

“B2”

2

6

2

‘2’

31H

7

CSUM

1

‘0’

30H

Checksum

“0B” 

2

8

2

‘B’

42H

9

ETX

1

03H

Packet terminator (fixed value)

Write 03 in binary.

1

Retrieve measurement data count in internal memory: BN

  Retrieves the number of measurement data entries stored in the MY600 internal 

memory.

  Transmission format

Order Command ASCII Binary

Description

Size

1

STX

1

02H

Packet start (fixed value)

Write 02 in binary.

1

2

TYPE

1

‘0’

30H

Data packet type: small

1

3

SIZE

1

‘0’

30H

Packet size

“07” : 7 BYTES

2

4

2

‘7’

37H

5

CMD

1

‘B’

42H

Data connection start command

“BN”

2

6

2

‘N’

4EH

7

CSUM

1

‘2’

32H

Checksum

“27” 

2

8

2

‘7’

37H

9

ETX

1

03H

Packet terminator (fixed value)

Write 03 in binary.

1

  Reception format

Order Command ASCII Binary

Description

Size

1

STX

1

02H

Packet start (fixed value)

Write 02 in binary.

1

2

TYPE

1

‘0’

30H

Data packet type: small

1

3

SIZE

1

‘0’

30H

Packet size

“0B” : 11 BYTES

2

4

2

‘B’

42H

5

CMD

1

‘B’

42H

Data connection start command

“BN”

2

6

2

‘N’

4EH

7

DATA

1

Data count

‘0000’ to ‘1000’

4

8

2

9

3

10

4

11

CSUM

1

Checksum

2

12

2

13

ETX

1

03H

Packet terminator (fixed value)

Write 03 in binary.

1

Retrieve measurement data from internal memory: BM

  Retrieves the measurement data stored in the MY600 internal memory.

  Transmission format

Order Command ASCII Binary

Description

Size

1

STX

1

02H

Packet start (fixed value)

Write 02 in binary.

1

2

TYPE

1

‘0’

30H

Data packet type: small

1

3

SIZE

1

‘0’

30H

Packet size

“0A” : 10 BYTES

2

4

2

‘A’

41H

5

CMD

1

‘B’

42H

Data connection start command

“BM”

2

6

2

‘M’

4DH

7

No.

1

Data number

‘000’ to ‘999’

3

8

2

9

3

10

CSUM

1

Checksum

2

11

2

12

ETX

1

03H

Packet terminator (fixed value)

Write 03 in binary.

1

  Reception format

Order Command ASCII Binary

Description

Size

1

STX

1

02H

Packet start (fixed value)

Write 02 in binary.

1

2

TYPE

1

‘0’

30H

Data packet type: small

1

3

SIZE

1

Packet size

Variable

2

4

2

5

CMD

1

‘B’

42H

Data connection start command

“BM”

2

6

2

‘M’

4DH

7

No.

1

Data number

‘000’ to ‘999’

3

8

2

9

3

DATA

Target internal memory data (variable)

N

CSUM

1

Checksum

2

2

ETX

1

03H

Packet terminator (fixed value)

Write 03 in binary.

1

  Data format

•  Voltage measurement:

  <model name>,<data number>,<save data>,<save time>,<range>,<site number 

1>,<site number 2>,<voltage>,<voltage unit>,<DC±/AC/-->

 Example  MY600,0000,2018/03/13,10:33:45,VOLT,00,00,100.0,V,AC

•  Insulation resistance measurement:

  <model name>,<data number>,<save date>,<save time>,<range>,<site number 

1>,<site number 2>,<insulation resistance>,<insulation resistance unit>,<elapsed 

time>,<1-minute value>,<1-minute value unit>,<DAR value>,<PI value>

  Example  MY600,0000,2018/03/13,10:33:45,1000V,00,00,100.0MΩ,00:10,----

,MΩ,--,----,----

  * The available six range are 50V, 100V, 125V, 250V, 500V, 1000V.

•  Low-resistance measurement

  <model name>,<data number>,<save date>,<save time>,<range>,<site number 

1>,<site number 2>,<resistance>,<resistance unit>

  Example  MY600,0000,2018/03/13,10:33:45,CONT,100.0,Ω

 

Checksum (CSUM)

  The checksum is the value of the last byte obtained by adding each byte from TYPE to 

the last DATA or number byte in each command format.

  Because the transmission/reception format of 10, 11, B1, B2 and the transmission format 

of BN are fixed, their checksums are also fixed.

 

IM 91030-01EN

2/2

Отзывы: