Load Impedance
1 Ω or less
Max. Output Voltage 30 V peak or less (restricted by the output protection circuit).
Working Voltage
Max. 600 Vrms
Influence of Adjacent Conductor When the frequency of the current in an adjacent conductor is 50 Hz, the
influence on the primary current is 0.5 mA/A or less.
Influence of Conductor Position in the Jaws 0.1% or less in the output signal for a frequency of 400 Hz or less.
Influence of Load Impedance r When 1 Ω < r ≤ 5 Ω, under 0.1% of output signal, and phase shift under 0.2 degrees
Influence of Frequency f
§
30 Hz ≤ f < 48 Hz: under 0.5% of output signal
65 Hz < f ≤ 1 kHz: under 1% of output signal
1 kHz < f ≤ 5 kHz: under 2% of output signal
Influence of Crest Factor
Under 1% of output signal, given a crest factor of 6 or less for a 2000 A peak
(333 Arms) current or less
Influence of DC Current Superimposed on Nominal Current
Under 1% of the output signal, assuming a current of DC 15 A or less.
Operating Temperature –10°C to +50°C
Storage Temperature –40°C to +70°C
Temperature Influence 0.02%/°C or less of the output signal
Operating Humidity
0 to 90% RH (no condensation) However, if 35°C is exceeded, humidity will impair the
primary functionality (by a factor of 0.5% RH/°C)
Influence of Humidity Under 0.1% of the output signal given 10% RH ≤ Humidity < 20% RH or
75% RH < Humidity ≤ 90% RH
Operating Altitude
2000 m or less above sea level
Max. Jaws Openin g
53 mm (open jaws height: 139 mm (W))
External Dimensions Approx. 111(W) x 216(H) x 45(D) mm
Weight
Approx. 620g.
Output
Safety jacks (Φ4 mm)
*
Measurement category
describes a number which defines a transient overvoltage condition. It implies the
regulation for impulse withstand voltage.
Measurement category III applies to measurement of the distribution level, that is , building wiring, fixed
installations. Measurement category IV applies to measurement of the primary supply level, that is,
overhead lines, cable systems, and so on.
†
Pollution degree
describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength
or surface resistivity is adhering.
Pollution Degree 2 applies to the normal indoor atmosphere. Normally, only non-conductive pollution
occures. Occasionally, however, temporary conductivity caused by condensation must be expected.
‡ Reference Conditions
Temperature: 23 °C±3°C
Humidity: 20 to 75% RH
External magnetic field < 40 A/m
No AC magnetic field
Conductor centered in jaws
Load impedance ≤ 1 Ω (≤ 1 VA)
No influence of current flowing in adjacent conductors
When the primary current is sinusoidal, the sinusoidal conditions are: frequency; 48 to 65 Hz, distortion
factor < 1%, no DC component
§ There is no frequency influence in the range 48 Hz ≤ f ≤ 65 Hz.
99
111
101
52
216
43.5
31
35
45
External Dimensions (Units: mm)
6. Maintenance
• Note the following when cleaning the probe.
• Do not clean the probe while clamped to a conductor. Likewise, do not clean while
connected to a measuring instrument.
• Do not allow water to contact the jaws.
• When opening the jaws, keep the exposed metal areas clean. If dust accumulates, wipe with a
clean dry cloth. To prevent rust, wipe metal surfaces with oil from time to time. Avoid getting oil on
non-metallic surfaces. Use only high quality, low-viscosity machine oil such as sewing machine oil.
• This product undergoes a 100% inspection at the time of shipment. If any layers of the core
come apart slightly during shipment, this will not affect the functioning of the product.
Four metal surfaces are
exposed with the clamp
is opened.
Metal surfaces
7. Servicing
If you encounter any problems during use, or if the device does not appear to be operating normal
-
ly, contact your dealer or nearest YOKOGAWA representative.
8. Appendix
Waste Electrical and Electronic Equipment
Waste Electrical and Electronic Equipment (WEEE), Directive
(This directive is valid only in the EU.)
This product complies with the WEEE directive marking requirement. This marking
indicates that you must not discard this electrical/electronic product in domestic
household waste.
Product Category
With reference to the equipment types in the WEEE directive, this product is classified
as a “Monitoring and control instruments” product.
When disposing products in the EU, contact your local Yokogawa Europe B.V. office.
Do not dispose in domestic household waste.
Authorized Representative in the EEA
Yokogawa Europe B. V. is Authorized Representative of Yokogawa Test & Measurement Corporation
in the EEA for this Product. To contact Yokogawa Europe B. V., see the separate list of worldwide
contacts, PIM 113-01Z2.
IM 751552-01E 2/2