Specifications
Product Specifications
Compatible sensors
700937 Current Probe
700939 FET Probe
701920 Differencial Probe
701930 Current Probe
701931 Current Probe
701932 Current Probe
701933 Current Probe
Number of power supply connectors
2
Output voltage
±12 V ±0.5 V
Operating temperature and humidity range
0 to 40°C (32 to 104°F), 80% RH or less (no
condensation)
Storage temperature and humidity range
-10 to 50°C (14 to 122°F), 80% RH or less (no
condensation)
Location for use
Indoor, altitude up to 2000 m (6562 feet)
Rated output current
600 mA (the total value of both outputs)
Rated supply voltage
100 V AC (120, 220, and 240 V requires specification)
(Voltage fluctuation of 10% from the rated supply
voltage are taken into account.)
Ripple supply voltage
3 mVp-p (at the rated output current)
Rated supply frequency
50/60 Hz
Maximum rated power
60 VA (at the rated output current)
External dimensions
Approx. 73(W) × 110(H) × 186(D) mm
Approx. 2.87”(W) × 4.33”(H) × 7.32”(D)
Mass
Approx. 1.1 kg
Approx. 38.8 oz.
Accessories
Power cord, Instruction manual,
Spare fuse (in the fuse holder)
100 V, 120 V : F1.0 AL/250 V, 20 mm × 5 mm dia.
220 V, 240 V : F0.5 AL/250 V, 20 mm × 5 mm dia.
Standards Applying
Safety
EN61010-1
Pollution Degree 2
EMC
Emission
EN61326-1 Class B
EN55011 Class B, Group 1
EN6100-3-2 Class B
EN6100-3-3
EMC standards of Australia and New Zealand
EN55011 Class B, Group 1
Immunity
EN61326-1 Table 1 (Basic immunity requirement)
Corrective Action to be Taken in the case of an Abnormality
• If an message appears on the screen, refer to the following pages.
• If maintenance service is required, or if the instrument still does not operate
properly even if the proper corrective action has been taken, contact your
nearest YOKOGAWA representatives, listed on the back cover of this manual.
Symptom
Possible Cause
Corrective Action
The power cannot be turned ON.
The fuse is blown.
Replace the fuse.