IM 51011-01EN
32
11. Comparator (COMP Key)
This function is available on Model 51021.
The comparator function determines whether the measured value
(displayed value) is within the specified range, greater than the
high limit, or less than the low limit.
Set the low limit (Lo) and high limit (Hi) to define the reference
range. The maximum number of digits available is 8.
Of these digits, 3 digits are significant, and the rest are place
holders (fixed at 0; including 2 decimal digits).
The meter displays the comparison result.
Display
Condition
Hi
high limit (Hi) < measured value
Go
low limit (Lo) ≤ measured value ≤ high limit (Hi)
Lo
measured value < low limit (Lo)
Setting the Comparator (Change)
1.
Press
SET
and then
COMP
.
[SET COMP Lo] appears. Lo indicates the low limit.
2.
Set the low limit. The top digit is displayed,
so use to set the number.
Press
. Set the remaining digits.
Use and to enter the digits one by one.
Example of how to set 1230
Display for setting
the top digit
Display for setting
the remaining (lowest) digits
3.
Press
COMP
. The display for setting the high limit appears.
Set the high limit in the same manner as in step 2.
(Press
COMP
to switch between high and low limits.)
4.
After entering the values, press
SET
. The setting is complete,
and [SET] disappears.
Содержание 51011
Страница 55: ...IM 51011 01EN 54 ...
Страница 57: ......