3. Normal Illuminance Measurement
Automatic Zero Adjustment
1.
Cover the light-detecting surface with the cap.
(Check that the HOLD switch is not locked.)
2.
Press the
POWER
key to turn on the power.
After all the elements light, automatic zero adjustment is
executed ([--CAL--] is displayed).
When the adjustment is finished, [--CAL--] disappears.
(Initial display end)
The illuminance measurement [0.00 lx] display appears.
([0.0 lx] appears on Model 51011 and 51012.)
Note:
If the [--CAP--] display persists, the cap may not be on
correctly. Put the cap on correctly. (The cap may be broken.)
Starting to Measure
3.
Remove the cap, and start measuring.
F
RSPS
AUTO POWER OFF
lx
4.
When you are finished measuring, press the
POWER
key
to turn off the power. Cover the light-detecting surface for
protection.
Note:
If [Err] appears during measurement, check that the light
detector is connected correctly to the main unit and that the
cap is off, and then start from the beginning.
(See section 17.1, “Error Messages.”)