3-12

IM 04P02B01-01E

3.6

Switching the Display Screen

This section describes the procedure for switching the display screen.
To change the displayed content, see chapter 8.

Procedure

Switching the Display Screen

The display screen switches each time the 

DISP

 key is pressed.

Screen 01 through 15 are switched in order.  Screens that are set to “Skip” are skipped.

Channel Auto Switching

On screens that show the measured values and computed values, the displayed channel
is automatically switched in ascending order.  The switching interval can be set to 1 s, 2
s, 3 s, 4 s, or 5 s.
Measurement channels set to Skip and computation channels set to OFF are not
displayed.

Switching the Displayed Channel Using Keys

If auto switching is not specified, the channel switches each time the 

CH UP

 key is pressed

in ascending order.  All channels are displayed in order.

Note

The operation is different between auto channel switching and manual switching using keys

on the 2-channel digital display.

03

04

05

06

01

02

02

03

03

04

01

02

During auto switching

When switching using keys

Содержание 437124

Страница 1: ...User s Manual Y okogawa Electric Corporation Models 437101 437102 437103 437104 437106 437112 437118 437124 µR20000 Recorder IM 04P02B01 01E 1st Edition ...

Страница 2: ... to the instrument s performance and functions Every effort has been made in the preparation of this manual to ensure the accuracy of its contents However should you have any questions or find any errors please contact your nearest YOKOGAWA dealer as listed on the back cover of this manual Copying or reproducing all or any part of the contents of this manual without the permission of Yokogawa Elec...

Страница 3: ... explanation in the manual Protective ground terminal AC High temperature To avoid injury caused by hot surface do not touch locations where this symbol appears For the protection and safe use of the product and the system controlled by it be sure to follow the instructions and precautions on safety that are stated in this manual whenever you handle the product Take special note that if you handle...

Страница 4: ...nnect the protective grounding before connecting to the item under measurement or to an external control unit Damage to the Protective Structure Operating the recorder in a manner not described in this manual may damage its protective structure Exemption from Responsibility YOKOGAWA makes no warranties regarding the product except those stated in the WARRANTY that is provided separately YOKOGAWA a...

Страница 5: ...der µR20000 12 dot recorder µR20000 18 dot recorder µR20000 24 dot recorder Japanese English deg F DST Alarm output relay 2 points1 Alarm output relay 4 points1 Alarm output relay 6 points1 Alarm output relay 12 points1 2 Alarm output relay 24 points1 2 RS 422A 485 interface3 Ethernet 10BASE T interface3 Fail Chart end detection and output2 Clamped input terminal4 Non glare door glass Mathematical...

Страница 6: ...ely The optional accessories below are available for purchase separately If you make an order make sure that all contents are present and undamaged For information about ordering accessories contact the dealer from which you purchased the recorder Red Green Blue Violet Purple Disposable felt pen Item Z fold chart paper 3 pcs 10 pcs 3 pcs 3 pcs 3 pcs 3 pcs 250 Ω 0 1 100 Ω 0 1 10 Ω 0 1 250 Ω 0 1 100...

Страница 7: ...display and key panel section toward you The section opens upward Tab on the display Open Open Display Remove all packing materials Pen Model Dot Model Pen model Dot model Return the display and key panel section to its original position by pressing the section downward Checking the Contents of the Package ...

Страница 8: ... functions such as how to assign tags to channels and how to set message strings that are to be printed 7 Setup Operations for Changing Adding Functions Basic Setting Mode Describes the setup operations for changing or adding functions such as setting the recorder to detect sensor burnouts and changing the contents that are printed on the chart paper 8 Setup Operations for Changing the Displayed C...

Страница 9: ...AUTION Calls attentions to actions or conditions that could cause light injury to the user or damage to the instrument or user s data and precautions that can be taken to prevent such occurrences Note Calls attention to information that is important for proper operation of the instrument Subheadings On pages that describe the operating procedures in Chapter 3 through 9 the following symbols are us...

Страница 10: ... 1 Names of Parts 3 1 3 2 Installing or Replacing the Chart Paper 3 4 3 3 Installing Replacing Felt Pens or Plotter Pen Pen Model 3 7 3 4 Installing Replacing the Ribbon Cassette Dot Model 3 9 3 5 Starting Stopping the Recording 3 11 3 6 Switching the Display Screen 3 12 3 7 Printing Measured Values Manual Printout 3 13 3 8 Printing the Recorder Settings 3 14 3 9 Clearing the Alarm Printout Buffer...

Страница 11: ...ing the Channel Recording Color Dot Model 7 8 7 6 Recording by Compensating for the Pen Offset along the Time Axis Pen Model 7 9 7 7 Turning Printouts ON OFF 7 10 7 8 Setting the Periodic Printout Interval and the Type of Measured Values to Be Printed 7 12 7 9 Setting the Bar Graph Display Mode 7 15 7 10 Setting the Key Lock Function 7 16 7 11 Enabling the Moving Average Function Dot Model 7 19 7 ...

Страница 12: ...g the Procedure Taken When the Computed Result Is in Error 9 28 Chapter 10 Troubleshooting 10 1 A List of Error Messages 10 1 10 2 Troubleshooting Flow Charts 10 4 Chapter 11 Maintenance 11 1 Periodic Inspection 11 1 11 2 Cleaning the Recorder 11 2 11 3 Replacing the Internal Light LED 11 3 11 4 Calibrating the Recorder 11 4 11 5 Adjusting the Pen Position Pen Model 11 6 11 6 Adjusting the Dot Pri...

Страница 13: ... Internal Light A light is provided for easier viewing of the recording area of the chart paper Display Measured values can be displayed numerically or using bar graphs on the large display Also alarm status and chart speed can be displayed Communication Function Using the Ethernet communication interface C7 option or the RS 422A 485 communication interface C3 option the measured values on the rec...

Страница 14: ... V is ON 1 1 5V 1 5V is scaled to values in the appropriate unit to be used as measured values Also the low cut function input less than 0 is fixed to 0 scale left value can be used Current Input A shunt resistor is attached to the input terminal The current signal is converted to a voltage signal and measured The measurable range is the range equivalent to the DC voltage range indicated above aft...

Страница 15: ...e root computation cannot be designated Scaling The input values are scaled to values in the appropriate unit to be used as measured values 0 V 100 0 C 10 V 300 0 C Measured value Input value Square Root Computation When the input type is DC voltage the square root of the input value is calculated the result is scaled to a value in the appropriate unit and used as the measured value of the channel...

Страница 16: ... Input When measuring the temperature using a thermocouple the reference junction compensation on the recorder can be used When using external reference junction compensation you can set the reference voltage The reference junction compensation can be set for each channel By default the recorder is configured to use the internal reference junction compensation function Note When using external ref...

Страница 17: ... output for a step input Moving Average Dot Model The average value of the m most recent values acquired at the scan interval is used as the measured value of the channel The number of moving averaged data points m can be set in the range 2 to 16 The figure below shows an example indicating the operation of the buffer for the moving average computation when the number of moving averaged data point...

Страница 18: ...to the model from the table below Model Integration Time of the A D Converter Pen model Select 16 7 ms 60 Hz 20 ms 50 Hz or Auto Dot model Select 16 7 ms 60 Hz 20 ms 50 Hz 100 ms or Auto If Auto is selected the recorder detects the power supply frequency and automatically selects 16 7 ms or 20 ms Because 100 ms is an integer multiple of 16 7 ms and 20 ms this setting can be used to suppress the po...

Страница 19: ... to the specified value Difference Low Limit Alarm l An alarm occurs when the difference in the input values of two channels is less than or equal to the specified value Can be specified on channels set to delta computation High Limit on Rate of Change Alarm R The rate of change of the measured values is checked over a certain time interval An alarm occurs if the rate of change of the measured val...

Страница 20: ... the alarm is released Note The alarm detection operation is reset when a power failure occurs The operation restarts after the power recovers If the alarm value is changed while a delay alarm is occurring the alarm is released if the new alarm value does not meet the alarm condition Alarm Hysteresis Hysteresis can be specified to the values for activating and releasing the alarm The hysteresis ap...

Страница 21: ...m Output Relay A1 A2 A3 A4 and A5 Options Contact signals can be generated from alarm output relays when alarms occur The number of output relays is 2 A1 4 A2 6 A3 12 A4 or 24 A5 The alarm output relays are denoted as I01 to I06 I11 to I16 I21 to I26 and I31 to I36 on the recorder The following functions can be assigned to the alarm output relay Diagnosis Output The diagnosis output can be assigne...

Страница 22: ...default the reflash alarm is disabled Alarm Alarm output relay Reflash alarm ON Alarm output relay Reflash alarm OFF Channel 1 Channel 3 Channel 2 Approx 500 ms Approx 500 ms Note If the reflash alarm is enabled I01 to I03 are set to reflash alarm operation In this case I01 to I03 are set to OR operation and non hold operation regardless of the settings specified in AND OR Operation of Alarm Outpu...

Страница 23: ...is output is enabled I01 is set to diagnosis output Specifying AND produces no effect Related Topics Setting the AND operation Section 7 1 Energized De energized Operation of Alarm Output Relays You can select whether the alarm output relay is energized or de energized when an alarm occurs If de energized is selected the status of the alarm output relay when an alarm occurs is the same as the stat...

Страница 24: ...ld Activated Released Note If the reflash alarm is enabled I01 to I03 are fixed to non hold operation Specifying Hold produces no effect If diagnosis output is enabled I01 is fixed to non hold operation Specifying Hold produces no effect Related Topics Setting the non hold hold operation of alarm output relays Section 7 1 Alarm ACK Operation The alarm acknowledge alarm ACK operation releases all a...

Страница 25: ...recording as well as various other types of information Recording Example on the Pen Model Recording Example on the Dot Model The recording examples may appear differently from the actual recording as a result of functional improvements made on the recorder after this manual was written Related Topics Starting Stopping recording Section 3 5 ...

Страница 26: ...he recording colors are assigned to each set channels 01 06 07 12 13 18 and 19 24 of six channels in ascending order as follows purple red green blue brown and black The recording color of each channel can be changed among these six colors For each channel trend recording can be enabled or disabled Related Topics Setting the dot printing interval Section 6 1 Changing the recording color Section 7 ...

Страница 27: ...ers there are offsets along the time axis phase difference between the pens This offset is corrected when pen offset compensation is used Same time Below is an explanation for the 2 pen model The recording of these two pens are offset by an amount of phase P If pen offset compensation is enabled the measured values of pen 1 are stored in the memory and recorded when the chart paper is fed by an am...

Страница 28: ...m3 2 390 6µS cm 3 H 598 4 C 4 d 0 222V 0 0 500 0 1CH mg cm3 RED 50mm h_ 08 00 25mm h 09 52 START 205 ABCDEF 1H3 10 09 1H3 10 05 50mm h 14 55 Time tick The time ticks are marks that indicate the positions of the date time on the chart paper Time tick cancel mark An exclamation point is printed when the periodic printout time tick was not printed at the correct position Related Topics Setting the ch...

Страница 29: ...mg cm3 02 437 2µS cm 03 H 591 6 C 04 0 222V 05 0 665V 06 L 0 448V _50mm h 14 55 _08 00 25mm h 09 52 START 205 ABCDEF 01H3 10 09 01H3 10 05 Nov 09 04 13 50 01 218 7mg cm3 02 390 6µS cm 03 H 598 4 C 04 d 0 222V 05 0 995V 06 L 0 448V 0 0 500 0 01CH mg cm3 50mm h_ Channel Printout Dot Model Only Prints the channel No or tag by the trend recording The channel No or tag is printed every approximately 25...

Страница 30: ...ils see appendix 1 Date time time ticks marks that indicate the positions of the date time on the chart paper measured values alarm status scale leftmost and rightmost values of span and recording color pen model for each channel and chart speed can be printed When pen offset is being executed on the pen model the pen offset marks are printed On the pen model if a time tick is not printed at the c...

Страница 31: ...en the buffer is full the message is not printed A buffer overflow mark is printed when there are messages that cannot be printed because the buffer is full The time printout format can be selected New Chart Speed Printout When the chart speed is changed the time tick dot model the date time of change and the new chart speed are printed An asterisk shows there are messages that cannot be printed T...

Страница 32: ...ettings such as the input range and alarm for each channel Setup List contains settings of basic specifications such as the alarm output relay operation and printout method Printout Example of List on the Pen Model The printout examples may appear differently from the actual printout as a result of functional improvements made on the recorder after this manual was written 1 4 Recording ...

Страница 33: ...l Description 1 Printout Example of List on the Dot Model The printout examples may appear differently from the actual printout as a result of functional improvements made on the recorder after this manual was written 1 4 Recording ...

Страница 34: ...turns ON when an alarm occurs and turns OFF when the alarm is released If the alarm indication is set to hold operation the indicator blinks when an alarm occurs When alarm ACK operation is performed the indicator turns ON or OFF MATH This indicator turns ON when computation on the computation function M1 option is started and turns OFF when stopped KEYLOCK This indicator turns ON when key lock is...

Страница 35: ...Channel on which an alarm is occurring Channel on which an alarm is not occurring Channel on which an alarm is not set 123456 789012 345678 901234 ABCDEF GJ P ABCDEF GJKMNP Date Time and Chart Speed The date format can be specified Date Tme Chart speed Aug 31 2005 11 26 37 12000mm h Status Display Remaining amount of chart paper is less than approx 2 cm F1 option Alarm status Related Topics Assign...

Страница 36: ... common logarithm y log10x Exponent EXP Determines the exponent y ex Relational computation LT LE GT GE Determines or of two elements and outputs 0 or 1 EQ NE Logical computation AND OR XOR NOT Determines the AND logical product OR logical sum XOR exclusive logical sum of two elements NOT negation of an element and outputs 0 or 1 TLOG computation TLOG SUM TLOG Determines the sum SUM the maximum MA...

Страница 37: ...lue of channel 02 is 20 V the computed result of 01 02 is 40 Recording Computation Channels Trend Recording Trend recording is possible by setting the recording span for each computation channel The recording span of computation channels can be set in the range of 9999999 to 99999999 excluding the decimal Zone recording and partial expanded recording are also possible Assignment of Computation Cha...

Страница 38: ...LOG timer see appendix 1 Related Topics Setting the periodic printout Section 7 8 9 6 and 9 15 Setting the printout using the TLOG timer Section 9 6 and 9 12 Alarms Up to 4 alarms can be assigned to each computation channel The alarm types are high limit alarm H low limit alarm L delay high limit alarm T and delay low limit alarm t Hysteresis can be specified to the values for activating and relea...

Страница 39: ...also when the power is turned OFF including a power failure This relay behavior cannot be changed FAIL Relay Output NO C NC Normal NO C NC Failure NO C NC Power OFF NO Normally Opened C Common NC Normally Closed Chart End Output When the remaining amount of chart paper falls to approximately 2 cm recording is stopped and the relay contact signal 1 relay is output The relay is energized in this cas...

Страница 40: ...l signal is applied to the terminal The secondary chart speed is set in advance Message 1 Printout to Message 5 Printout Remote input signal Trigger Prints message 1 to 5 on the chart paper Manual Printout Remote input signal Trigger Executes a manual printout Alarm ACK Remote input signal Trigger Alarm ACK is executed on all alarms that are occurring This is valid only when the alarm indication o...

Страница 41: ...gnal produces no effect Remote Signal Edge Trigger and Level The above actions are carried out on the rising or falling edge of the remote signal edge the ON signal lasting at least 250 ms trigger or the ON OFF signal level Rising Falling edge Trigger Level Rising Falling 250 ms or more Operates at the secondary chart speed 250 ms or more For contact inputs the remote signal rises when the contact...

Страница 42: ...pen replacement pen model ribbon cassette replacement dot model Related Topics Setting the key lock function Section 7 10 Using the key lock function Section 3 13 Language The display and recording language can be set to English or Japanese Related Topics Changing the language Section 7 14 DST If the recorder is used in a region that has DST time can be switched automatically between DST and stand...

Страница 43: ...eformation Keep electrically charged objects away from the recorder as this may cause malfunction Do not apply volatile chemicals to the door glass display panel keys etc Do not allow rubber and vinyl products to remain in contact with the recorder for long periods of time This may damage the recorder When not in use make sure to turn OFF the power switch If there are any symptoms of trouble such ...

Страница 44: ...anel mounting Note Condensation may occur if the recorder is moved to another place where both the ambient temperature and humidity are higher or if the temperature changes rapidly In addition measurement errors will result when using thermocouples In this case let the recorder adjust to the new environment for at least one hour before using it The chart paper may be adversely affected by a rapid ...

Страница 45: ...rder to the panel according to the procedure below First attach the two mounting brackets and temporarily fasten the attachment screws Next fix the recorder in place by tightening the attachment screws with the appropriate torque When the recorder is approximately perpendicular to the panel as you fasten the screws press the mounting bracket against the case so that they are in contact with each o...

Страница 46: ...herwise specified tolerance is 3 however tolerance is 0 3 mm when below 10 mm 39 5 1 56 180 5 7 11 279 6 11 01 180 7 09 108 4 25 180 7 09 108 4 25 0 94 1 0 279 6 11 01 1 0 54 6 2 15 288 11 34 294 6 11 6 288 11 34 294 6 11 6 220 8 66 180 5 7 11 39 5 1 56 9 4 0 37 2 to 26 panel thickness 7 5 0 3 24 360 MIN 14 17 281 11 06 11 06 2 0 281 2 0 361 MIN 14 21 Panel Cutout 2 2 Installation ...

Страница 47: ... on lug with insulation sleeves designed for 4 mm screws be used when connecting the input output signal wires to the terminals However this does not apply clamped terminals H2 Crimp on lug with insulation sleeves for 4 mm screws For clamped terminals H2 the following wire is recommended Conductive cross sectional area for single wire 0 14 mm2 to 1 5 mm2 stranded wire 0 14 mm2 to 1 0 mm2 Length of...

Страница 48: ...connected devices If you need to make a parallel connection then Turn the burnout detection function OFF Ground the instruments to the same point Do not turn ON or OFF another instrument during operation This can have adverse effects on the other instruments RTDs cannot be wired in parallel Arrangement of the Measuring Input Terminals A terminal cover is screwed in place on the measuring input ter...

Страница 49: ...ut terminals 1 Turn OFF the recorder and remove the terminal cover 2 Connect the signal wires to the terminals Clamped input terminal Screw input terminal Philips screwdriver Flat blade screwdriver Crimp on lug with insulation sleeves Signal wire Signal wire Note Input signal wires of diameter less than or equal to 0 3 mm may not be secured firmly for clamped terminals H2 Fold over the conducting ...

Страница 50: ...current input Leadwire resistance 10 Ω max wire The resistance of the three wires should be equal Example For a 4 to 20 mA input a shunt resistor of 250 Ω 0 1 can be used to convert to 1 5V input A B b Note RTD input terminals A and B on the dot model are isolated on each channel Terminal b is shorted internally across all channels However for 3 legs isolated RTDs N2 option input b is also isolate...

Страница 51: ...sure not to touch the terminals Crimp on lug with insulation sleeves for 4 mm screws CAUTION To prevent fire use signal wires having a temperature rating of 70 C or more If a strong tension is applied to the cable wired to the recorder the terminals of the recorder and or the cable can be damaged In order to prevent tension from being applied directly on the terminals fasten all wiring cables to t...

Страница 52: ...O C NC NO C NC NO C NC I01 I02 I03 I04 I05 I06 NO C NC NO C NC NO C NC NO C NC NO C NC NO C NC A3 F1 I01 I02 I03 I04 I05 I06 I01 I02 I03 I04 I05 I06 NO C NC NO C NC R1 CE FAIL 1 2 3 4 5 C 1 2 3 4 5 C R1 NO C NC NO C NC NO C NC NO C NC NO C NC NO C NC A4 R1 I01 I02 I03 I04 I05 I06 1 2 3 4 5 C R1 R1 F1 NO C NC NO C NC CE FAIL F1 NO C NC NO C NC NO C NC NO C NC NO C NC NO C NC A4 I01 I02 I03 I04 I05 ...

Страница 53: ... NC NO C NC I21 I22 I23 I24 I25 I26 NO C NC NO C NC NO C NC NO C NC NO C NC NO C NC I31 I32 I33 I34 I35 I36 NO C NC NO C NC NO C NC NO C NC NO C NC NO C NC A5 R1 I01 I02 I03 I04 I05 I06 1 2 3 4 5 C R1 Alarm Output Relay Terminals and FAIL Chart End Output Relay Terminals NC Normally Closed C Common NO Normally Opened Remote Control Input Terminals 1 to 5 remote control input terminals C Common Ala...

Страница 54: ... of the remote control input terminals Connect the shield to the ground terminal of the recorder Relay Contact Output Specifications Output type Relay Contact rating 250 VAC 50 60 Hz 3 A 250 VDC 0 1 A resistive load Dielectric strength 1500 VAC at 50 60 Hz for one minute between output terminals and the ground terminal Relay Contact Input Transistor Input Specifications Input signal Voltage free c...

Страница 55: ... power switch double pole type on the power supply line in order to separate the recorder from the main power supply Put an indication on this switch as the breaker on the power supply line for the recorder and indications of ON and OFF Switch specifications Rated power current 1 A or more Rated rush current 60 A or more Use a switch complied with IEC 60947 1 3 Connect a fuse between 2 A and 15 A ...

Страница 56: ...minal cover in place Power terminal block Power terminal cover 2 Wire the power cord and the protective ground cord to the power supply terminals Use ring tongue crimp on lugs designed for 4 mm screws L N Philips screwdriver Crimp on lug with insulation sleeves Power cord Protective ground cord 3 Replace the power terminal cover and fasten it with screws The proper torque for tightening the screws...

Страница 57: ...wer switch of the recorder or another device during operation This can have adverse effects on the measured values If nothing is displayed when the power switch is turned ON turn OFF the power switch and check the points listed above one more time After checking the points turn ON the power switch again If the unit still does not work it is probably malfunctioning Contact your nearest YOKOGAWA dea...

Страница 58: ...one hole on each of the top bottom left and right panels The hole is covered with a seal Plotter pen Prints various types of information Pen model Dot model Display and key panel see the next page There are internal lights on the bottom section of the display and key panel They lights up the recording area of the chart paper Recording pen Records the measured value Printer carriage Records measure...

Страница 59: ...everse order LEFT RIGHT Key Moves the cursor to the right when entering a value or character Press this key while holding down the SHIFT key to move the cursor to the left SHIFT Key Used with the key key or the CHARACTER key ENTER Key Confirms the setup item or value CH UP key Switches the displayed channel when manual switching is specified During normal operation Status display Displays the foll...

Страница 60: ...input terminals Heatsink Dissipates the internal heat Optional terminal block This is where terminals or ports used by options such as alarm output relays and communication interface are installed Ethernet port C7 option Power terminal block The power terminal and protective ground terminal 3 1 Names of Parts ...

Страница 61: ...e to the chart cassette platen the cylindrical section that holds the paper during the recording operation Be sure to replace the chart paper ahead of time Loading the Chart Paper 1 Open the door If recording is in progress press the RCD key to stop the recording Open 2 Remove the chart cassette Gently press the center stopper inward The bottom section of the chart cassette comes out Gently lift t...

Страница 62: ...g the stopper on either side Chart holder Black Stopper Front cover Chart holder Transparent Stopper Chart holder Transparent 1 2 3 2 4 Load the chart paper Riffle the chart thoroughly before loading Make sure that the sprocket teeth of the chart drives are properly engaged in the chart paper perforations Make sure not to load the chart paper backwards Z fold chart paper Sprocket teeth 3 2 Install...

Страница 63: ...k into the recorder case Align the left and right projections with the guide grooves of the recorder and press the entire chart cassette into the recorder case The chart cassette is fixed in place with the stoppers Stopper Feeding the Chart Paper 7 Press the FEED key to assure that the chart moves two or more folds smoothly into the chart receiver If it moves unsteadily do the installing procedure...

Страница 64: ...color Replacing Felt Pens 1 Open the door If recording is in progress press the RCD key to stop the recording 2 Open the display and key panel section Hold the left and right tabs and pull the display and key panel section toward you The section opens upward 3 Hold the felt pen cartridge and pull it out from the pen holder If the pen pen holder is at a position that is not easily accessible see Wh...

Страница 65: ...er position and the Pen exchange End appears Note When the pen moves a line is drawn on the chart paper 4 Replace the pen 5 Press the CH UP key The screen returns to the data display screen Replacing the Plotter Pen 1 Open the door If recording is in progress press the RCD key to stop the recording 2 Open the display and key panel section 3 Hold the plotter pen cartridge and pull it out from the p...

Страница 66: ...top the recording 2 Press the FUNC key 3 Press the DISP key several times to display Ribbon exchange Func Ribbon exchange 4 Press the CH UP key The printer carriage moves near the center position and Ribbon exchange End is displayed 5 Open the display and key panel section Hold the left and right tabs and pull the display and key panel section toward you The section opens upward Note If the record...

Страница 67: ...shaft of the cassette with the ribbon feeding shaft hole of the holder Groove for the ribbon cassette stopper Ribbon feed shaft Ribbon feeding knob 8 Turn the ribbon feeding knob in the direction of the arrow a half turn or more to check that the ribbon is feeding properly If the ribbon is loose turn the knob in the direction of the arrow to tighten it 9 Return the display and key panel section to...

Страница 68: ...ord RECORD Stopping the Recording While recording is in progress press the RCD key to stop recording The word RECORD on the status display clears Feeding the Chart Paper The chart paper is fed while the FEED key is held down Viewing the Recorded Results Pull the front cover tab of the chart cassette to open the front cover The recorded chart paper can be pulled out for viewing Front cover tab left...

Страница 69: ...puted values the displayed channel is automatically switched in ascending order The switching interval can be set to 1 s 2 s 3 s 4 s or 5 s Measurement channels set to Skip and computation channels set to OFF are not displayed Switching the Displayed Channel Using Keys If auto switching is not specified the channel switches each time the CH UP key is pressed in ascending order All channels are dis...

Страница 70: ... the recorder continues the measurement and alarm detection in the background When manual printout is complete trend recording resumes If an alarm occurs during the manual printout the alarm is printed after the trend recording resumes Aborting the Manual Printout The following procedure is for aborting the manual printout that is in progress 1 Press the FUNC key 2 Press the key to select Print ou...

Страница 71: ...st Start and then press the key The List printout starts The screen returns to the data display screen Print List Start Aborting the List Printout 1 Press the FUNC key 2 Press the key to select Print out and then press the key 3 Press the key to display List Stop and then press the key The List printout stops The screen returns to the data display screen Print List Stop Starting the Setup Printout...

Страница 72: ...o prevent unneeded alarm printouts from being executed Procedure 1 Press the FUNC key 2 Press the key to select Buffer clear and then press the key Func Buffer clear 3 Press the key to select Alarm and then press the key The data in the alarm printout buffer is cleared The screen returns to the data display screen Buffer clear Alarm When the key or key is pressed while holding down the SHIFT key t...

Страница 73: ...ng a Message 1 Press the FUNC key 2 Press the key to select Message and then press the key Func Message 3 Press the key to select message number and then press the key The message printout starts The screen returns to the data display screen Displays the preset message Message Message1 Clearing the Message Printout Buffer Messages waiting to be printed are temporarily stored in the buffer memory T...

Страница 74: ...iodic printout Procedure Resetting the Report Data 1 Press the FUNC key 2 Press the key to select Periodic and then press the key Func Periodic 3 Press the key to select Reset and then press the key The report data is reset and the calculation of the report data starts again from that point The screen returns to the data display screen Log data Reset When the key or key is pressed while holding do...

Страница 75: ...e alarm is released The alarm output is released when the alarm ACK operation is carried out When an alarm is occurring The alarm output is held even after the alarm ACK operation If the alarm ACK operation is carried out again after the alarm is released the alarm output is released Indication Status Display Blinks when an alarm occurs When the alarm is released The alarm indication turns OFF whe...

Страница 76: ...t Keylock and then press the key The key lock is activated The screen returns to the data display screen Func Keylock Releasing the Key Lock Note A password is required to release the key lock 1 Press the FUNC key 2 Press the key to select Keylock and then press the key 3 Enter the password for releasing the key lock The password values are shown with asterisks Press the key to increment the value...

Страница 77: ...utation channels cannot be changed while computation M1 option is in progress The operations that can be carried out in this mode are described in chapter 5 and 6 Basic Setting Mode This mode is used to set the basic specifications of the recorder such as the thermocouple burnout detection function and the alarm output relay operation This mode cannot be entered while the recorder is recording or ...

Страница 78: ... from Setting mode Hold down the MENU key for 3 seconds to enter Setting mode Next hold down both the DISP key and the FUNC key for 3 seconds The Basic Setting mode display appears The top and bottom lines are the setup item and comment respectively The section that is blinking in the setup item that you change In this manual the section that you change appears shaded The item to be controlled bli...

Страница 79: ...creen appears When the Setting complete screen appears the changed item is applied 01 01 Channel Setting complete Using the ESC Key If you press the ESC MENU key the operation is cancelled and the display returns to a higher level menu In other words if you do not show the Setting Complete screen the changes you made up to that point are discarded Press the ESC MENU 1 key while holding down the SH...

Страница 80: ...RACTER key to show Ins DISP and then press the key A space for one character is inserted Enter the character Position to insert the character Display Ins DISP Unit ppm CHR Ins DISP Deleting a Character Use the key or SHIFT key to move the cursor to the character to be deleted Press the CHARACTER key or SHIFT CHARACTER key to show Del DISP and then press the key The character is deleted Deleting an...

Страница 81: ...on 3 12 section 9 1 section 3 10 sections 3 7 3 8 sections 3 9 3 10 section 3 13 section 3 11 section 3 3 Pen model Dot model section 3 4 Math key FUNC Alarm ACK Alarm output release operation This operation is valid only when the alarm indication or relay output is set to hold operation Math Models with the Computation Function M1 Option Starts stops the computation Print out Starts Stops the man...

Страница 82: ...g pen to a position that is easily accessible for replacement on the pen model Ribbon exchange Dot Model Moves the printer carriage near the center position when replacing the ribbon cassette on the dot model 4 3 Menu Structure Settings and List of Default Values ...

Страница 83: ...ft Span right K Number CH CH CH Left Right CH CH Trend CH CH Periodic Tag Duration Level Timer number Alarm SUM scale Type Value Relay Relay number Zone Partial Partial Expand Boundary Print out Tag Alarm delay itme Display Number of samples Response Time Duration Light DST Start month DST Start time mm h Tag Periodic are not displayed in the default condition To display these items settings must ...

Страница 84: ...a length Parity Protocol Timer number Pen CH Error Over Mode Interval Ref Time Reset Print Color Output pen Bar graph Error data Host Local IP DNS Login Login set Timeout K Alive A M G DNS P S Domain Suffix_P Login Register User Password Suffix_S P_Adj Pen number Value End End Timeout Duration Keep alive Host Level Suffix primary IP address Subnet mask Primary server Secondary server Gateway Suffi...

Страница 85: ...al Expand 1 to 99 50 Aux Partial Boundary Within the recording span range Aux Print out CH 01 to 24 01 Aux Print out Trend Dot Model On Off On Aux Print out Periodic On Off On Aux Tag CH 01 to 24 01 Aux Tag Tag 7 characters or less 01 to 24 Aux Message Message No 1 to 5 1 Aux Message Message 16 characters or less Blank Aux Chart2 Pen Model 82 types pen model 25 mm h Dot Model 1 to 1500 mm h dot mo...

Страница 86: ...On Off On Math Aux Tag CH 0A 0B 0C 0D 0E 0F 0G 0J 0K 0M 0N 0P 0A 1A 1B 1C 1D 1E 1F 1G 1J 1K 1M 1N 1P Math Aux Tag Tag 7 characters or less 0A to 1P Math Aux Alm delay T CH 0A 0B 0C 0D 0E 0F 0G 0J 0K 0M 0N 0P 0A 1A 1B 1C 1D 1E 1F 1G 1J 1K 1M 1N 1P Math Aux Alm delay T Duration 0 to 3600s 10s Setup Items in Basic Setting Mode and Their Default Values Setup Item Pen Dot Selectable Range or Selections...

Страница 87: ...ylock Periodic Free Lock Free Keylock Pen exchange Pen Model Free Lock Free Keylock Ribbon exchange Dot Model Free Lock Free Moving_AVE Moving_AVE Dot Model Not Use Not Filter Filter Pen Model Not Use Not Partial Partial Not Use Not Language Lang English Japanese English Date format Type Y M D M D Y D M Y D M Y M D Y M D Y Temperature Temp C F C Personalize Add function Bias Not Use Not Personaliz...

Страница 88: ...r data Over Skip Limit Skip RS 422A 485 communication interface function C3 option Setup Item Pen Dot Selectable Range or Selections Default Value RS 422 485 Address 1 to 32 1 RS 422 485 Baud rate 1200 2400 4800 9600 19200 38400 9600 RS 422 485 Data length 7 8 8 RS 422 485 parity Odd Even None Even RS 422 485 Protocol NORMAL MODBUS NORMAL Ethernet communication interface function C7 option Setup I...

Страница 89: ...t Selectable Range or Selections Default Value P_Adj P_Adj Dot Model Hysteresis Zero Full Hysteresis Pen Model Zero Full Zero P_Adj Pen No Pen Model 1 2 3 4 1 Store Setup Item Pen Dot Selectable Range or Selections Default Value End End Store Abort Store 4 3 Menu Structure Settings and List of Default Values ...

Страница 90: ...et the unit after scaling Low cut 7 15 Use Personalize Add function SQRT low cut in Basic Setting mode and select Use or Not If Use is selected set the low cut value using Range in Setting mode 5 1 If Not is selected the Low cut item does not appear in the Range setting Bias Use Personalize Add function Bias in Basic Setting mode and select Use 7 15 or Not 6 12 If Use is selected set the bias valu...

Страница 91: ... alarm Use Alarm Hysteresis in Basic Setting mode to set the hysteresis to be applied 7 1 to the high limit alarm and low limit alarm of measurement channels Likewise use Alarm M_Hysteresis to set the hysteresis to be applied to the high limit alarm and low limit alarm of computation channels M1 option Change the alarm output relay operation If Alarm Act in Basic Setting mode set to De_energize al...

Страница 92: ...he alarm occurrence and release print only the alarm occurrence or not print Time printout format 7 16 Use Personalize Time print Alarm in Basic Setting mode to set the time printout format when printing alarm occurrence release Recording Start Printout Turn printout On Off 7 7 Use Print Record On in Basic Setting mode to enable disable the recording start printout Time printout format 7 16 Use Pe...

Страница 93: ...et the timer used on each 9 6 computation channel Unit Use Math Unit in Setting mode to set the unit for each channel 9 3 Constants Use Math Constant in Setting mode to define constants to be used in the computing equations 9 4 Alarms Use Math Alarm in Setting mode to set the alarm for each channel 9 5 Delay high limit alarm delay low limit alarm Alarm type 7 15 Use Personalize Add function Alarm ...

Страница 94: ...out interval 9 12 Use Math Timer TLOG Print in Basic Setting mode to turn the printout On 9 12 Use Math Aux Print out Periodic in Setting mode to enable disable the 9 9 periodic printout for each computation channel Set Math TLOG Timer No in Setting mode to set the timer used on each 9 6 computation channel Scale printout and recording color pen model Common with periodic printout Alarm occurrence...

Страница 95: ... select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 4 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 5 Hold down the MENU key ...

Страница 96: ... 328 0 to 1472 0 F J 200 0 to 1100 0 C 328 0 to 2012 0 F T 200 0 to 400 0 C 328 0 to 752 0 F N 0 0 to 1300 0 C 32 to 2372 F W 0 0 to 2315 0 C 32 to 4199 F L 200 0 to 900 0 C 328 0 to 1652 0 F U 200 0 to 400 0 C 328 0 to 752 0 F WRe 0 0 to 2400 0 C 32 to 4352 F RTD Range Type Selectable Span Range C Selectable Span Range F PT Pt100 200 0 to 600 0 C 328 0 to 1112 0 F JPT JPt100 200 0 to 550 0 C 328 ...

Страница 97: ...t ESC Set Range CH 01 01 Mode Scale Type Volt Range 2V Span left 2 000 Span right 2 000 Scale left 0 00 Scale right 200 00 01 01 Channel Setting complete First channel Last channel Description Scale Left and Scale Right Selectable range mantissa 20000 to 30000 Decimal position XXXXX XXXX X XXX XX XX XXX X XXXX The decimal position is the position specified by scale left Note The displayable printa...

Страница 98: ...the same value Input Type Range Type Selectable Span Range C Selectable Span Range F DC voltage 20 mV 20 00 to 20 00 mV Volt 60 mV 60 00 to 60 00 mV 200 mV 200 0 to 200 0 mV 2 V 2 000 to 2 000 V 6 V 6 000 to 6 000 V 20 V 20 00 to 20 00 V 50 V 50 00 to 50 00 V Thermocouple R 1760 0 to 1760 0 C 3168 to 3168 F TC S 1760 0 to 1760 0 C 3168 to 3168 F B 1820 0 to 1820 0 C 3276 to 3276 F K 1570 0 to 1570...

Страница 99: ...annel ESC Set Range CH 01 01 MOde DI Range Level Span left 0 Span right 1 01 01 Channel Setting complete Description Selectable Range Type Span Left and Span Right The range type span left and span right can be set in the range shown below Span left and span right cannot be set to the same value Range Type Selectable Span Values Level 0 Less than 2 4 V 1 2 4 V or higher within 6 V Contact 0 Open 1...

Страница 100: ...pan left range 0 8 to 1 2 V Selectable span right range 4 8 to 5 2 V Scale Left and Scale Right Selectable range mantissa 20000 to 30000 Decimal position XXXXX XXXX X XXX XX XX XXX X XXXX The decimal position is the position specified by scale left Make sure that scale left is less than scale right Low cut On Cuts values below 0 of the recording span set to 0 scale left value Off Does not apply th...

Страница 101: ... complete First channel Last channel Description Input Type in Which Square Root Computation Can Be Performed Square root computation can be performed when set to DC voltage input Span Left and Span Right Same as the selectable span range of the DC voltage input Make sure that span left is less than span right Scale Left and Scale Right Selectable range mantissa 20000 to 30000 Decimal position XXX...

Страница 102: ...0000 the value when the low cut function is applied is displayed as Over Skip Unused Channels Select Skip Set the channel range The new setting takes effect First channel Last channel ESC Set Range CH 01 01 Mode Skip 01 01 Channel Setting complete Note Do not use channels set to skip as a reference channel of delta computations or in computing equations of the computation function M1 option 5 1 Se...

Страница 103: ...isplay returns to a higher level menu 4 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 5 Hold down the MENU key for 3 seconds to return to Operation mode Select the level number between 1 and 4 Select On to set the alarm When set to On the settings below can be entered Set the channel range Select the alarm ...

Страница 104: ...e reference channel Alarm on a channel set to delta computation Alarm H L R r T and t Alarm h and l Measured value If high limit on rate of change alarm or low limit on rate of change alarm is specified set the interval over which the rate of change is calculated If delay high limit alarm or delay low limit alarm is specified set the delay time Related Topics Setting the interval of the high limit...

Страница 105: ...4 High Limit on Rate of Change Alarm Low Limit on Rate of Change Alarm A value greater than or equal to 1 digit can be specified For example 1 digit corresponds 0 001 for the 2 V range measurable range 2 000 to 2 000 V The maximum value that can be specified is the width of the measurable range 4 000 V for 2 V range For ON OFF input only 1 can be specified Relay No The selectable relay numbers are...

Страница 106: ...the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 4 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 5 Hold down the MENU key for 3 seconds to return to Operation mode Set the unit Set the channel range The new setting takes effect First ch...

Страница 107: ...lete screen appears do either of the following To correct the setting press the key If you are done press the ESC key 5 Hold down the MENU key for 3 seconds to return to Operation mode Set the chart speed The new setting takes effect ESC Set Chart mm h 25 Chart speed Setting complete Explanation Chart Speed Pen Model The chart speed can be selected from 82 settings shown below 5 6 8 9 10 12 15 16 ...

Страница 108: ...n 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 4 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key 5 Hold down the MENU key for 3 seconds to return to Operation mode Set the time day month year The new s...

Страница 109: ... of the following To correct the setting press the key If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Select the trend recording interval mode The new setting takes effect ESC Set Aux Aux Trend Mode Auto Trend Interval Setting complete Explanation Mode Auto The trend recording interval is set according to the chart speed in the range of 10 s to...

Страница 110: ...ess the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel range Set the filter time constant Th...

Страница 111: ...iddle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel range Set the number of samples of the moving average Moving ...

Страница 112: ...ay returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel range Set the leftmost value of the recording zone Set the rightmost value of the recording zone The new setting takes effect First channel Las...

Страница 113: ...the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel range Select On to use the partial expand...

Страница 114: ...ress the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done pre...

Страница 115: ...ress the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel range Set the tag The new setting ta...

Страница 116: ...section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Select the message No Set messa...

Страница 117: ...ach item and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If y...

Страница 118: ...ears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel range Set the alarm delay time The new setting takes effect ESC First channel Last channel Set Aux Aux Alarm delay time CH 01 01 Duration 10 01 01 Channel Setting complete Explanation Duration The duration can be se...

Страница 119: ... the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Select the brightness of the display Select the brightness of the internal light The...

Страница 120: ...5 Hold down the MENU key for 3 seconds to return to Operation mode Select whether to use the bias Set the bias value Set the channel range The new setting takes effect First channel Last channel ESC When bias is On Set Bias CH 01 01 Bias Off Bias 0 000 01 01 Channel Setting complete Explanation Channel Range The only case when a bias can be set on multiple channels simultaneously is when the chann...

Страница 121: ...ENU key for 3 seconds to enter Setting mode 2 Press the key or SHIFT key to select Aux and then press the key 3 Press the key or SHIFT key to select DST and then press the key 4 Set each item and press the key Press the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entere...

Страница 122: ...Setting complete Month day of the week and time when the DST ends Month The n the day of the week of the month Time n Explanation DST Select Use to use DST Start month Start day and Start time Set the date time for switching from standard time to DST Specify the day as the n th day of the week of the month End month End day and End time Set the date time for switching from DST to standard time Spe...

Страница 123: ...ng is in progress or when computation is in progress on models with the computation function M1 option Procedure Changing the Settings 1 Hold down the MENU key for 3 seconds to enter Setting mode 2 Hold down the and keys simultaneously for 3 seconds to enter Basic Setting mode 3 Press the key with Alarm shown on the screen 4 Set each item and press the key Use the key or SHIFT key to select values...

Страница 124: ...ehavior Nonhold Indicator Nonhold Increase 01 Decrease 01 Hysteresis 0 5 Math hysteresis 0 0 Alarm Setting complete Applying the Changes and Returning to Operation Mode Press the key or SHIFT key to select End and then press the key Press the key to select Store and then press the key The changes are applied and the screen returns to Operation mode If you select Abort and press the key the changes...

Страница 125: ...er to set the alarm indication to hold operation or non hold operation Nonhold Releases the alarm indication at the same time the alarm is released Hold Holds the alarm indication until the alarm ACK operation is executed Increase Set the interval of the high limit on rate of change alarm to an integer between 1 and 15 The interval is set to scan interval 1 to 15 Pen model The scan interval is 125...

Страница 126: ...ey If you are done press the ESC key Select the integration time of the A D converter The new setting takes effect ESC A D Integrate Setting complete Basic A D Integrate Auto Applying the Changes and Returning to Operation Mode Press the key or SHIFT key to select End and then press the key Press the key to select Store and then press the key The changes are applied and the screen returns to Opera...

Страница 127: ...s entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key Set the channel range The new setting takes effect Select the recording operation that is carried out when a burnout is detected ESC First channel Last channel 01 01 Channel Sett...

Страница 128: ...le of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key Set the channel range Select the RJC mode of the thermocouple input Set the compensation voltage when using the external RJC function The new setti...

Страница 129: ... Uses the RJC function on the recorder External Uses an external RJC function If external is selected set the compensation voltage Volt Sets the compensation voltage when using an external RJC function The compensation voltage can be set in the range of 20000 µV to 20000 µV 7 4 Setting the RJC Function on Channels Set to TC Input ...

Страница 130: ...splay returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key Select the channel Select the recording color The new setting takes effect ESC 01 Channel Color Setting complete Basic Color Channel 01 Color Purple Applying the Changes and Returning to Operation Mode Press the key or SHIF...

Страница 131: ...lete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Select On to record by compensating for the pen offset along the time axis The new setting takes effect ESC POC Setting complete POC Off Basic POC Applying the Changes and Returning to Operation Mode Press the key or SHIFT key to select End and then press the key Press the key to s...

Страница 132: ...section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Select whether to print the channel number beside the trend recording Select whether to use the ch...

Страница 133: ...creen returns to Operation mode Press the ESC key to return to the Basic screen Explanation CH Tag Channel Prints the channel number Tag Prints the tag Channel Dot Model When On is selected channel No is printed by the trend recording Alarm On1 Prints the alarm information when an alarm occurs or releases On2 Prints the alarm information only when an alarm occurs Off Does not print alarm informati...

Страница 134: ...ddle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key When Period is set to Manual Set the reference time Select the periodic printout interval mode Select the periodic printout interval Set the typ...

Страница 135: ... level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key Select the type of report data to print Set the channel range Select the sum scale The new setting takes effect ESC First channel Last channel Printout2 Setting complete Basic Print2 CH 01 01 Mode AVE SUM scale Off Applying the Changes and Returnin...

Страница 136: ...IN Prints the minimum value over the interval MAX Prints the maximum value over the interval INST Prints the instantaneous value SUM scale Set this item only when SUM is selected SUM sums the data every computation interval For flow values that have units s min or h a simple summation results in the actual value not matching the computed result because the scan interval and the unit of the input v...

Страница 137: ...play returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key Set the channel range Select the display mode of the bar graph The new setting takes effect ESC First channel Last channel 01 01 Channel Setting complete Basic Bar graph Graph Normal CH 01 01 Applying the Changes and Returni...

Страница 138: ...er Setting mode 2 Hold down the and keys simultaneously for 3 seconds to enter Basic Setting mode 3 Press the key or SHIFT key to select Keylock and then press the key 4 Set each item and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to th...

Страница 139: ...k to apply the key lock function to the message printout operation Select Lock to apply the key lock function to the operation of clearing the printout buffer The new setting takes effect Select Lock to apply the key lock function to the operation of resetting the report data in periodic printout Select Lock to apply the key lock function to the operation of moving the carriage assembly to a posit...

Страница 140: ... operation of the keys to be key locked Select Lock or Free for each key operation RCD RCD key Feed FEED key Menu MENU key Disp Menu MENU 1 key Alarm ACK Alarm ACK operation under the FUNC key Math Computation start stop reset operation under the FUNC key Print out Printout start stop reset operation under the FUNC key Message Message printout operation under the FUNC key Buffer clear Clear operat...

Страница 141: ...SC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Enable Disable moving average The new setting takes effect ESC Move AVE Setting complete Basic Moving_AVE Moving_AVE Not Applyin...

Страница 142: ...e of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Enable Disable the filter function The new setting takes effect ESC Filter Setting complete Filter Not Basic Filter Applying the Changes and Return...

Страница 143: ...peration the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Enable Disable partial expanded recording The new setting takes effect ESC Partial Setting complete Basic Partial Partial Not Applying the Changes and Return...

Страница 144: ...y in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Select the display recording language The new setting takes effect ESC Language Setting complete Basic Language Language English Appl...

Страница 145: ...ers see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Enable Disable the bias function Enable Disable the low cut function on channels set to sq...

Страница 146: ...m does not appear in Setting mode 1 5V low cut Enables Disables the 1 5V low cut function Use Enables the setting of the low cut function when a channel is set to 1 5V in Setting mode Not The 1 5V low cut item does not appear in Setting mode Alarm delay Enables Disables the alarm delay function Use Enables the setting of delay high limit alarm or delay low limit alarm for the alarm type Not The de...

Страница 147: ...acters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Select the time printout format of the alarm printout Select the time printout format o...

Страница 148: ... D H M Month Day Hour Minute M D H M S Month Day Hour Minute Second YMD H M S Month Day Year Hour Minute Second None Does not print the date time RCD On Sets the time printout format of the record start time printout The format is the same as that of the alarm printout C Speed Selects the time printout format of the new chart speed printout The format is the same as that of the alarm printout Note...

Страница 149: ...s the key to select the settings to be initialized and then press the key Set Initializes the settings in Setting mode Setup Set Initializes the settings in Basic Setting mode and Setting mode 5 To execute the initialization press the key to select Yes and then press the key The settings are initialized and the recorder returns to Operation mode To cancel the initialization press the ESC key The s...

Страница 150: ...ter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Select the remote control input terminal number The new setting takes...

Страница 151: ...e internal clock to the nearest hour Manual print Executes manual printout Message Prints message where is a value between 1 and 5 Chart speed Changes the chart speed Math start stop Starts stops the computation on the computation function M1 option Math reset Resets the computed result of the computation function M1 option None No function is assigned Related Topics Setting the secondary chart sp...

Страница 152: ... the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Select the date format The new setting takes effect ESC Date format Setting complete Basic Date format Type M D Y Applying...

Страница 153: ...r characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Select the temperature unit The new setting takes effect ESC Temperature Setting ...

Страница 154: ...Hold down or 3 s to switch The panel keys are set to the functions marked above the keys as shown below FEED DISP MENU RCD MENU FUNC CH UP SHIFT ESC CHARACTER 1 Returning from the Data Display Setup Screen to Data Display Screen Hold the MENU 1 key for 3 seconds to return from the data display setup screen to data display screen Note When the data display setup screen is shown operations such as m...

Страница 155: ...r each channel End Channel alarm status display End Date Time Chart speed display End DI DO display when the R1 or A option is installed End Display in which different screens can be assigned to the top and bottom sections Channel No Display update rate Top section Channel No Display update rate End Bottom section 1 channel digital display tag display 2 channel digital display tag display 1 channe...

Страница 156: ...ching interval and then press the key 5 When the Setting complete screen appears press the key Hold the MENU 1 key down for 3 seconds to exit from the data display setup screen Select the channel switching interval Select the display type Select the screen No Current display type name The new setting takes effect Screen 01 Setting complete Screen number 01 Interval Auto2s 01 200 0 mV Explanation I...

Страница 157: ...the key to show Math and then press the key 3 Press the key with Stop shown on the screen The computation stops and the screen switches to the data display screen The word MATH in the status display clears Math Stop Resetting the Computed Values of All Computation Channels to 0 This operation can be executed only when the computation is stopped 1 Press the FUNC key 2 Press the key to show Math and...

Страница 158: ...hen are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode The new setting takes effect ESC Select On to use the computing equation Set the channel range Set the equation Set the left spa...

Страница 159: ...ntire string Copy DISP Store the entire string in the memory Past DISP Paste the string in the memory Inserting a Character Press the key or SHIFT key to move the cursor to the position where the character is to be inserted Press the CHARACTER key or SHIFT CHARACTER key to show Ins DISP and then press the key A space for one character is inserted Enter the character Deleting a Character Use the ke...

Страница 160: ...lusive logical sum OR XOR low order of precedence Data That Can Be Used in Equations In TLOG computation the measured values of measurement channels or computed values of computation channels can only be used For all other computations all of the data types below can be used Measured Values on Measurement Channels Specified by the channel No For channels that have linear scaling enabled the scaled...

Страница 161: ...d channel in this case channel 01 and make the result the computed value of computation channel 0A Computing equation of computation channel 0A 0A 01 Power and Other Computations The data that can be used in equations are measured values computed values constants communication input data and remote control input terminal status You can nest a computing element inside the parentheses of another com...

Страница 162: ... of channel 3 the computed result is 1 Otherwise the result is 0 02 GE 03 If the measured value of channel 2 is greater than or equal to the measured value of channel 3 the computed result is 1 Otherwise the result is 0 02 LE 03 If the measured value of channel 2 is less than or equal to the measured value of channel 3 the computed result is 1 Otherwise the result is 0 Logical Computation Checks w...

Страница 163: ...Syntax e1XORe2 Condition If the two data values e1 and e2 are zero and non zero or non zero and zero the computed result is 1 Otherwise it is 0 Explanation e1 0 e1XORe2 0 e2 0 e1 0 e1XORe2 1 e2 0 e1 0 e1XORe2 1 e2 0 e1 0 e1XORe2 0 e2 0 NOT Logical negation Syntax NOTe1 Condition The result is the inverse of the status of data e1 zero or non zero Explanation e1 0 NOTe1 1 e1 0 NOTe1 0 Equation Examp...

Страница 164: ...nel e1 TLOG MIN Minimum value Syntax TLOG MIN e1 Condition Determines the minimum value of channel e1 TLOG AVE Average value Syntax TLOG AVE e1 Condition Determines the average value of channel e1 TLOG SUM Sum value Syntax TLOG SUM e1 Condition Determines the sum of channel e1 TLOG P P Maximum minimum value Syntax TLOG P P e1 Condition Determines the maximum minimum value of channel e1 Equation Ex...

Страница 165: ...acters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel range...

Страница 166: ...the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Select the constant number The new setting takes effect E...

Страница 167: ... the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 6 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel range The new setting takes effect ESC Select the level number betwee...

Страница 168: ...lay high limit alarm or delay low limit alarm Section 9 11 Enabling the alarm delay function Section 7 15 Setting the alarm hysteresis Section 7 1 Alarm Value Can be set to a value in the range of 9999999 to 99999999 excluding the decimal 9999999 to 99999999 is the displayable range Relay No The selectable relay numbers are listed below I01 and I02 on the A1 option I01 to I04 on the A2 option I01 ...

Страница 169: ...G and then press the key 4 Set each item and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 5 When the Setting complete screen appears do either of the following Press the...

Страница 170: ...e time mode Yes No selectable Yes Timer Periodic Periodic Measurement and computation channels Absolute time mode Yes No For the timer 1 and 2 settings see section 9 12 Absolute time mode Keeps time from the reference time Relative time mode Starts the time in sync with the starting of the computation SUM scale Set the sum scale when using statistical calculation TLOG SUM For the definition of the...

Страница 171: ...the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 7 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel range The new setting takes effect ESC Set the leftmost value of the recording zone ...

Страница 172: ...haracters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 7 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel ra...

Страница 173: ...r SHIFT key to select Aux and then press the key 4 Press the key or SHIFT key to select Print out and then press the key 5 Set each item and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a h...

Страница 174: ...u press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 7 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel range The new setting takes effec...

Страница 175: ...play returns to a higher level menu 6 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key 7 Hold down the MENU key for 3 seconds to return to Operation mode Set the channel range The new setting takes effect ESC Set the alarm delay time First channel Last channel Set Math Math Aux Aux Alarm delay Time Duration 10...

Страница 176: ...tem and press the key Use the key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following To correct the setting press the key If you ar...

Страница 177: ...he interval The reference time is set on the hour 00 to 23 Example 1 Reference time 14 00 Interval 12h The timeout times are hour 2 and hour 14 Example 2 Reference time 00 00 Interval 10min The timeout times are 0 hour 0 hour 10 min 0 hour 20 min 23 hour 40 min and 23 hour 50 min For example if computation is started at 9 hour 36 min the timeout times are 09 hour 40 min 09 hour 50 min 10 hour and ...

Страница 178: ...n 1 2 3 Reset Reset Reset Reset On Timer timeout Reset Off When reset is ON the sum value is calculated over each interval When set to Off the sum value from computation start is calculated Printout On Prints the computed value every interval specified by the timer Computation channels set to Off do not apply Off Does not print 9 12 Setting the Timer Used in TLOG Computation and Periodic Printout ...

Страница 179: ...ngs entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Select the channel The new setting takes effect ESC Select the recording color 0A Channel color Setting complete Basic Math Math Color Channel 0A Color Purple Applying the Cha...

Страница 180: ...dle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Select the pen number and channel number The new setting takes effect ESC Pen 1 Setting complete Basic Math Math Output pen Pen 1 CH 01 Applying ...

Страница 181: ...key or SHIFT key to select values For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following To correct the setting press the key If you are done press the ESC key Set t...

Страница 182: ...NST Prints the instantaneous value SUM scale Set this item only when SUM is selected For the definition of the term sum scale see page 7 14 Off Simply sums the computed values s Sums by converting the computed values to a value over 1 second min Sums by converting the computed values to a value over 1 minute h Sums by converting the computed values to a value over 1 hour day Sums by converting the...

Страница 183: ...operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key Set the channel range The new setting takes effect ESC Select the display mode of the bar graph First channel Last channel 0A 0A Channel Setting complete Bas...

Страница 184: ... For the procedure on how to enter values or characters see section 4 2 If you press the ESC key in the middle of the operation the settings entered up to then are cancelled and the display returns to a higher level menu 6 When the Setting complete screen appears do either of the following Press the key to set other channels If you are done press the ESC key Select the computation error procedure ...

Страница 185: ...An over value on a computation channel refers to a value that is less than or equal to 100000000 or greater than or equal to 100000000 excluding the decimal Skip The over value is not used in the computation Limit The following limit value is used for the computation Channel Limit Value Measurement channel For scaling 1 5V scaling and square root Over 315000 Over 31500 excluding the decimal For in...

Страница 186: ... limits are equal This is not allowed 24 The lower span limit is greater than the upper span limit 25 The lower scale limit is greater than the upper scale limit 26 Bias cannot be set to the SKIPPED channel 27 Bias cannot be set to the DI channel 30 The partial boundary value exceeds the range of the span 31 Partial is invalid on the SKIPPED channel 35 The upper and lower limits of the printing zo...

Страница 187: ... is invalid during chart feed 169 This action is invalid during ribbon hold Operation Errors Code Message Explanation Countermeasures 232 There is no available data There is no data for periodic printout or data for calculating TLOG when the timer expired Communication Errors Code Message Explanation Countermeasures 300 Command is too long 301 Too many number of commands delimited with Keep the nu...

Страница 188: ... YOKOGAWA dealer 910 A D error Contact your nearest YOKOGAWA dealer 921 A D calibration value error Contact your nearest YOKOGAWA dealer 922 A D calibration is in the wrong order Contact your nearest YOKOGAWA dealer 930 Memory acquisition failure Contact your nearest YOKOGAWA dealer 940 The ethernet module is down Contact your nearest YOKOGAWA dealer 950 A D number error Contact your nearest YOKOG...

Страница 189: ... the switch work properly Connect properly Turn ON the power properly Contact your nearest YOKOGAWA dealer Section 2 5 Section 2 6 Supply the power at the specified voltage and frequency Does the input meet the specifications NO NO YES YES The reading error is large The indication is unstable The pen is off the scale on either the 0 or 100 side Adjust the span and range Change the input to meet th...

Страница 190: ...ctions 7 12 and 6 2 moving average dot model see sections 7 11 and section 6 3 Wire correctly see section 2 3 Install the terminal block correctly Fasten the terminal screws correctly Insulate the RTD from ground If burnt replace the TC Is the input wiring parallel with other equipment NO NO YES YES Is the RJC setting correct TC input Stop the TC burnout upscale downscale function of other equipme...

Страница 191: ...former into the power line Insert a power line filter into the power line If an inductive load is connected to an alarm contact output use a surge suppresser on that line Contact your nearest YOKOGAWA dealer Display and other functions do not work properly Is the key lock released NO YES Release the key lock using the FUNC key see section 3 13 Contact your nearest YOKOGAWA dealer Keys do not work ...

Страница 192: ... characters clear not blurred For the replacement procedure of felt pens and plotter pens see section 3 3 pen model For the replacement procedure of the ribbon cassette see section 3 4 dot model Is the chart paper feeding properly no paper jams If not see chapter 10 Is there enough chart paper remaining Remaining chart length is printed on the left margin of the chart at intervals of 20 cm For the...

Страница 193: ...er carriage shaft be cleaned once a year Cleaning Procedure Wipe the dust off the plotter carriage shaft using a soft lint free cloth or paper If the dirt does not come off easily apply ethyl alcohol to the cloth or paper Plotter carriage Shaft Dot Model To maintain smooth operation wipe the dust off the two shafts of the printer carriage using a soft lint free cloth or paper It is recommended tha...

Страница 194: ...l section 3 Pull out the internal light LED assembly The LED assembly is attached to the left and right of the display and key panel section Display and key panel LED assembly LED assembly 4 Remove the LED assembly from the connector of the flexible printed circuit board LED assembly Connector Flexible printed circuit board Flexible printed circuit board LED assembly Connector 5 Connect the new LE...

Страница 195: ...pecifications Standard temperature stability accuracy 0 05 C For information on purchasing the calibration instruments contact your nearest YOKOGAWA dealer Calibration Procedure 1 Wire the recorder and the calibration instrument as shown in the following figure and adequately warm up the instruments the warm up time of the recorder is at least 30 minutes 2 Check that the operating environment such...

Страница 196: ...force table based on 0 C The recorder performs compensation by measuring the temperature at the input terminal and adding the corresponding thermoelectromotive force to the actual output of the thermocouple Therefore when the measurement terminal is shorted equivalent to the case when the detector tip is 0 C the measured value indicates the temperature of the input terminal When calibrating the re...

Страница 197: ...art paper or Full right edge of the chart paper and then press the key Adjust Zero first and then Full Increase or decrease the setting to align the pen position with the scale line on the chart paper Then press the key To change the setting press the key to move the cursor and key to change the value The pen moves after the key is pressed A value change of 1 corresponds to 0 033 mm change in the ...

Страница 198: ...nter of the chart paper Zero left edge of the chart paper or Full right edge of the chart paper and then press the key Adjust Hysteresis first followed by Zero and then Full Adjusting the Hysteresis A line is drawn at the center of the chart paper If the line appears as shown in Figure a increase the displayed setting Then press the key If the line appears as shown in Figure b decrease the display...

Страница 199: ...mplete Hysteresis 2 Applying the Changes and Returning to Operation Mode Press the key or SHIFT key to select End and then press the key Press the key to select Store and then press the key The changes are applied and the screen returns to Operation mode If you select Abort and press the key the changes are discarded and the screen returns to Operation mode Press the ESC key to return to the Basic...

Страница 200: ...ED is the half life of the brightness The deterioration of brightness varies depending on the condition of use and its determination is subjective Consider these facts for determining the actual replacement period Pen Model Note 41 days 2 km B8802CA MOTOR ASSY 5 years Item Replacement Period Part Name Part No Z fold chart paper CHART B9573AN When used at 20 mm h 33 days When used at 25 mm h Pen se...

Страница 201: ...bbon cassette RIBBON CASSETTE B9906JA 5 years B8802CA 5 years 3 years PULLY B9901EY 5 years MOTOR ASSY B9906FX 5 years B8803BB MOTOR ASSY B9906GM 5 years GEAR B8801BW B8801BX 5 years MOTOR ASSY B9906GM Display DISPLAY ASSY MOTOR ASSY For the carriage Ribbon feed gear 5 years GEAR B9906GN B9901HM B9901HN Quantity Used 1 1 1 1 2 1 1 1 1 each 1 Internal light LED 2 years LED ASSY B8800CR 2 1 each The...

Страница 202: ... to 50 00 V 1 5V 0 800 to 1 200 V range to 4 800 to 5 200 V range TC R1 0 0 to 1760 0 C 32 to 3200 F S1 0 0 to 1760 0 C 32 to 3200 F B1 0 0 to 1820 0 C 32 to 3308 F K1 200 0 to 1370 0 C 328 to 2498 F E1 200 0 to 800 0 C 328 0 to 1472 0 F J1 200 0 to 1100 0 C 328 0 to 2012 0 F T1 200 0 to 400 0 C 328 0 to 752 0 F N1 0 0 to 1300 0 C 32 to 2372 F W2 0 0 to 2315 0 C 32 to 4199 F L3 200 0 to 900 0 C 32...

Страница 203: ...l set to delta computation the input value of the reference channel where reference channel NO channel No set to delta computation Computable input type DC voltage TC and RTD The channel set to delta computation and the reference channel is set to the same range type Linear scaling Scales the input value to a value with a unit suitable for the application Computable input type DC voltage TC RTD DI...

Страница 204: ...common within a channel 1 to 3600 s 1 hour However if the alarm delay time setting is not an integer multiple of the scan interval the alarm delay time is set to a value that is the first integer multiple of the scan interval exceeding the delay time setting Hysteresis Set a width to the value for detecting alarm occurrence release common to all channels and all levels Applied to high limit alarm ...

Страница 205: ...tion the other section is reduced Boundary position 1 to 99 Boundary value Within the recording span range Trend Recording Dot Model Item Specifications Recording method Trend recording by a 6 color wire dot printer Recording color Purple Red Green Blue Brown Black Trend recording interval Records at the following interval through a raster scan system Auto Synchronizes to the chart speed and autom...

Страница 206: ...ute second month day hour minute month day hour minute second year month day hour minute second and not print Message printout buffer Stores 5 messages waiting to be printed Recording start printout Prints at the start of recording Printout contents Time and chart speed printout buffer overflow mark Time printout format Selectable from hour minute hour minute second month day hour minute month day...

Страница 207: ...art printout Same as the pen model With time ticks New chart speed printout Same as the pen model With time ticks Manual printout Same as the pen model Setting printout List Same as the pen model Setting printout Setup Same as the pen model Items dependent on the chart speed 1 to 9 mm h 10 to 100 mm h 101 to 1500 mm h Channel printout Alarm printout Message printout New chart speed printout Printe...

Страница 208: ...play System display Lights off Skip Multiple display The following screens can be assigned arbitrarily to the top and bottom sections of the display 1 channel digital display 2 channel digital display 3 channel digital display Date Time Chart speed display Channel alarm status display Status display Lights off 1 channel digital display tag display Display update of measured values computed values ...

Страница 209: ...is shown Higher precedence H L T t h l R r Lower precedence Special values Measurement channel Status Display Description Over Over See below Over Over See below Skip SKIP A value for channels set to skip Error ERROR Values such as when both the reference channel and measurement channel are Over or Over in delta computation Burnout B out A value displayed when a burnout is detected on a channel se...

Страница 210: ...when an alarm occurs and clears when the alarm is released If the alarm indication is set to hold operation the indicator blinks when an alarm occurs When alarm ACK operation is performed the indicator turns ON stop blinking or OFF Date time and chart speed display Displays the date time on the left and chart speed on the right The date display format can be selected The format is common with the ...

Страница 211: ...display format as 4 channel digital display Value 5 or 6 digits 1 channels 1 to 12 2 channels 18 to 24 Displayable values are 19999 to 99999 and a decimal If this range is exceeded the value is displayed as or Channel 2 digits mode 1 digit alarm 1 digit value 7 digits unit 6 digits Channel 2 digits alarm 1 digit value 6 digits Measurement channels Channel 2 digits mode 1 digit alarm 1 digit value ...

Страница 212: ...nts switches every three sconds Flag and scale Alarm is not set Alarm is not occurring Channel No Alarm is occurring 123456 789012 345678 901234 ABCDEF GJKMNP ABCDEF GJKMNP Aug 31 2005 11 26 37 12000mn h 180mm Dot model Analog 24 Math 24 TI00001dH 1999 9ABCDE TI00002H9999999 9ABCDE TI00001dH 1999 9ABCDE TI00002H9999999 9ABCDE ABC DEF OFF ON TI00001dH 1999 9ABCD TI00001dH 1999 9ABCD ABC DEF TI00002...

Страница 213: ...999 9ABCDEF Computation channel 0AH9999999 9ABCDEF Uses the same digital display format as 2 channel digital display Uses the same digital display format as 4 channel digital display Uses the same digital display format as 6 channel digital display 01dH 1999 9ABC 0AH9999999 9AB 01H 1999 9 02H 1999 9 0AH9999 9 Date time and chart speed The date display format can be changed on the Basic Setting mod...

Страница 214: ... stop synchronization Baud rate 1200 2400 4800 9600 19200 or 38400 bps Start bit Fixed to 1 bit Data length 7 bits or 8 bits Stop bit Fixed to 1 bit Parity Odd even or none no parity Received buffer length 2047 bytes Escape sequence Open and close Protocol Dedicated protocol Modbus slave protocol Communication mode ASCII for input output for control and setting ASCII or binary for input output of ...

Страница 215: ...n input data Digital data via communication can be used in computations other than statistical computation TLOG Number of communication input data points 8 channels pen model 24 channels dot model Range 9 9999E 29 to 1 0000E 30 0 1 0000E 30 to 9 9999E 29 Symbol C01 to C24 Remote input Use remote input status 0 1 in equations Number of inputs 5 Symbol D01 to D05 Computation accuracy 32 bit floating...

Страница 216: ... Cu10 WEED Cu10 BAILEY Cu10 α 0 00392 at 20 C Cu10 α 0 00393 at 20 C Cu25 α 0 00425 at 0 C Measurement Accuracy Recording Accuracy Measurement accuracy 0 3 of recording span 0 4 of rdg 1 0 C 0 3 of rdg 0 8 C Input Type 3 Legs Isolated RTD N2 Item Specifications Input terminal isolation Applies to the dot model The RTD input terminals are isolated between channels on the standard dot model input te...

Страница 217: ...er built in isolated power supply for the input terminals and shared common Input type and signal level Voltage free contact Contact closed at 200 Ω or less and contact open at 100 kΩ or greater Open collector ON voltage 0 5 V or less sink current 30 mA or more leakage current when OFF 0 25 A or less Signal type Edge level or trigger 250 ms or more Operation Perform the following control operation...

Страница 218: ...z 2 60 Hz 2 Vibration 10 to 60 Hz 0 2 m s2 or less Shock Not acceptable Magnetic field 400 A m or less DC and 50 60 Hz Noise Normal mode 50 60 Hz Volt The peak value including the signal must be less than 1 2 times the rated value of the range Example 2 V for 2 V range TC The peak value including the signal must be less than 1 2 times the rated value of the range RTD 50 mV or less Common mode nois...

Страница 219: ...allation category II1 Measurement category II2 Pollution degree 23 UL Certified by UL61010B 1 CSA NRTL C CE EMC EN61326 Emission Class A Immunity Annex A EN61000 3 2 EN61000 3 3 EN55011 Class A Group 1 Low voltage directive EN61010 1 Installation category II Measurement category II Pollution degree 2 C Tick AS NZS CISPR 11 Class A Group 1 1 Installation category overvoltage category describes a nu...

Страница 220: ... 0 15 of rdg 0 5 C except 0 15 of rdg 0 7 C for J 200 to 100 C 0 15 of rdg 0 7 C 0 15 of rdg 1 C 0 15 of rdg 0 5 C except 0 15 of rdg 0 7 C for L 200 to 100 C Pen model Dead band 0 2 of recording span Dot model Resolution 0 1 mm 10 µV 100 µV 0 2 of rdg 1 C 0 15 of rdg 0 3 C DC voltage TC DC voltage excludes RJC accuracy RTD ON OFF input 20 mV 60 mV 200 mV 2 V 6 V 20 V R 1 5 V S B K E J N L W T U 5...

Страница 221: ... AC 50 60 Hz and DC 400 A m fields Measurement 0 1 of rdg 10 digit or less Recording 0 5 of the recording span or less Input source resistance DC voltage range With variation of 1 kΩ Ranges of 200 mV or less Within 10 µV Ranges of 2V or more 0 1 of rdg TC range With variation of 1 kΩ 10 µV or less RTD range With variation of 10 Ω per wire resistance of all three wires must be equal Measurement 0 1...

Страница 222: ...otherwise specified tolerance is 3 however tolerance is 0 3 mm when below 10 mm 39 5 1 56 180 5 7 11 279 6 11 01 180 7 09 108 4 25 180 7 09 108 4 25 0 94 1 0 279 6 11 01 1 0 54 6 2 15 288 11 34 294 6 11 6 288 11 34 294 6 11 6 220 8 66 180 5 7 11 39 5 1 56 9 4 0 37 2 to 26 panel thickness 7 5 0 3 24 Note When mounting the recorder to a panel use two brackets top and bottom or left and right ...

Страница 223: ...etter d for channels set to delta computation Alarm Time tick Unit Measured value Channel No or tag Date Time Channel data Computation channel Scale Recording color Chart speed Mar 31 2004 15 50 TAG 001 1 000V 2 1 000V 3 H 2 1050UNIT03 4 d 2 000V TAG 00A 1UNIT0A C L 1 00 D 10 00UNIT0D E 1 0000 J 1 00UNIT0J 0 000 2 000 TAG 001 V RED 50mm h_ Date Time The date time when the periodic printout was exe...

Страница 224: ... 15 00 TPR TAG 001 1 000V 2 1 000V 3 2 105500E 03UNIT03 4 2 000V TAG 00A 1UNIT0A C 1 00 D 10 00 150 00 100 00UNIT0D E 1 0000 J 1 00UNIT0J 0 000 2 000 TAG 001 V RED 50mm h_ Report start date time and report end date time The start date time and end date time of the report statistical calculation periodic printout interval Time tick and time tick cancel mark These are the same as with the instantane...

Страница 225: ...e Sum value Min Max Ave Time tick Unit Range change occurrence mark Measured value Channel No or tag Report start date time Report end date time Channel data Computation channel Scale Chart speed Event Mar 31 2004 15 00 Apr 01 2004 15 00 TPR TAG 001 1 000V 02 1 000V 03 2 105500E 03UNIT03 04 2 000V TAG 00A 1UNIT0A 0C 1 00 0D 10 00 150 00 100 00UNIT0D 0E 1 0000 0J 1 00UNIT0J 0 000 2 000 TAG 001 V 50...

Страница 226: ... A to J or tags computed values and unit Alarm This is the same as with the instantaneous mode of the periodic printout of the pen model Timer Prints the timer number being used Dot Model Event Unit Channel No or tag Measured value Date Time Channel data Scale Timer Alarm Mar 31 2004 15 50 TPR TAG 00A 1UNIT0A 0C L 1 00 0D 100 00UNIT0D 0E 1 0000 0G 1 0 0J 1 00UNIT0J 100 100 TAG 00A UNIT0A TLOG2 Dat...

Страница 227: ...on Operation After the printout time Statistical calculation data is created immediately after power is restored The measured computed data up to the time of the power failure is used At the next printout time data after the power recovery is used to create the statistical calculation data Before the printout time The data is created at the printout time after the power recovers The measured compu...

Страница 228: ...rt end output 1 27 chart paper 12 4 chart speed 1 14 5 13 cleaning 11 2 clearing alarm information 3 15 comment line 4 3 common logarithm 1 24 communication input data 9 4 compensation voltage 7 7 computation errors 9 29 computation dedicated channels 1 24 computing equation 9 2 computing equations 9 4 constants 9 4 9 10 construction 12 17 conventions used in this manual viii copying pasting a cha...

Страница 229: ...ntout 1 19 model iv mounting brackets 2 3 mounting hole 3 1 moving average 1 5 7 19 N Page names of parts 3 1 new chart speed printout 1 19 non hold 1 9 1 12 non hold operation 7 3 normal operating conditions 12 17 number of samples of moving average 6 3 O Page offset compensation mark App 1 ON OFF input 5 5 operation menus using the FUNC key 4 5 operation mode 4 1 operations on the panel keys 4 2...

Страница 230: ...y chart speed 6 9 Setting complete screen 4 3 setting mode 4 1 settings necessary to use various functions 4 14 Setup 3 14 setup guide 4 14 setup items basic setting mode 4 10 setup Items setting mode 4 9 shunt resistor 1 2 skip 5 8 square root 1 24 square root computation 1 3 5 7 square root low cut function 7 24 standard accessories v standard performance 12 19 standard temperature device 11 5 s...

Отзывы: