User’s
Manual

4th Edition: October 2017
All Rights Reserved. Copyright © 
2008, Yokogawa Meters & Instruments Corporation, 
2017, Yokogawa Test & Measurement Corporation
Printed in Japan

Store this manual in an easily  
accessible place for quick reference.

Safety Precautions

This product is designed to be used by a person with specialized  

knowledge. When operating the instrument, be sure to observe  

the cautionary notes given below to ensure correct and safe use  

of the instrument. If you use the instrument in any way other than  

as instructed in this manual, the instrument’s protective measures  

may be impaired. This manual is an essential part of the product;  

keep it a safe place for future reference.

YOKOGAWA is by no means liable for any damage resulting from  

use of the instrument in contradiction to these cautionary notes.

The following safety symbols are used on the instrument  

and in this manual:

Danger! Handle with Care

This symbol indicates that the operator must refer to  

an explanation in the instruction manual in order to avoid  

the risk of injury or death of personnel or damage to  

the instrument.

WARNING

 

Indicates a hazard that may result in the loss of life or  

serious injury of the user unless the described instruction is  

abided by. 

CAUTION

 

Indicates a hazard that may result in an injury to the user  

and/or physical damage to the product or other equipment  

unless the described instruction is abided by.

Model 30032A
Leakage Clamp-on Tester

IM 30032A-E

4th Edition: 

Oct. 2017 (YMI)

This symbol indicates that this instrument designed to 

be applied around or removed from HAZARDOUS LIVE 

conductors provided if the RATED circuit-to-earth voltage 

dose not exceed the value indicated in the measurement 

category

Double Insulation

This symbol indicates double insulation.

Alternating Current

This symbol indicates aiternating current (AC).

Direct Current

This symbol indicates direct current (DC).

Earth TERMINAL

This symbol indicates ground.

■ Strictly observe the following cautionary notes in order to 

 

avoid the risk of injury or death of personnel or damage to 

 

the instrument due to hazards such as electrical shock.

WARNING

 

• This instrument is a current measuring device (sensor). 

  Do not use for any other purpose.

• Do not use the instrument if there is a problem with its 

 

physical appearance. (Do not use the instrument if there is any 

damage to the casing, battery cover, display and labels  

or when the casing is removed.)

• The barrier is there to protect you from touching the conductor. 

 

Be careful not to reach cross the barrier when using the  

instrument.

• Disconnect the instrument from the measurable conductors 

 

under test before opening the casing to replace the battery.

• Do not use the product when there are raindrops or droplets 

 

of condensed water on its surface, or if your hands are wet.

• Safety protectors such as rubber-insulated gloves should be 

worn to prevent electrical shock when using the instrument.

• Do not use this product in a place where an explosive 

 

gas or vapor is present.

• Do not open the case except when replacing batteries. 

  Only Yokogawa service personnel are authorized to remove  

the casing or disassemble or modify the instrument.

  Do not attempt to repair the instrument yourself,  

as doing so is extremely dangerous.

  When the instrument needs an internal inspection or  

calibration, contact Yokogawa or the dealer from whom  

you purchased the instrument.

WARNING

 

The maximum allowed current is 62 Arms and 

the RATED circuit-to-earth voltage is 300 Vrms or less.

Do not apply input exceeding this value.  

Otherwise, it will not only damage the tester, but also pose  

a risk of damage to the human body.  

For safety standards, refer to the specifications.

CAUTION

 

The product is for domestic use (Class B) and meets  

the electromagnetic compatibility requirenents.

CAUTION

 

• The jaw section is a delicate, precision sensor. 

 

Do not subject the jaw to unreasonably strong shock,  

vibration, or force when using it.

• If dust gets into the tops of the jaws, remove it immediately. 

 

Do not close the jaws when dust is trapped in its joints as  

the sensor may break.

CAUTION

 

• Do not use the instrument near noise-emitting equipment 

 

or where there may be sudden changes in temperature.  

Otherwise, the instrument may produce an unstable readings  

or errors.

• Do not wipe the instrument using an organic solvent such as 

 

benzine or paint thinner. Otherwise, the front panel may be  

damaged or discolored. When cleaning the instrument,  

use a dry cloth.

• Do not leave the tester exposed to direct sunlight or in a 

 

hot and humid location such as the inside of a car,  

for any prolonged length of time.

• If the instrument will not be used for long periods, 

 

remove the battery.

■ Measurement Category 

WARNING

 

The 30032A is designed for measurement category III.

Do not use the 30032A for measurements in location  

that fall under measurement category IV.

Measurement 

category

Description

Remarks

O

(None, Other)

For measurements performed on  

circuits not directly connected to  

MAINS.

Circuits not connected to  

a mains power source.

CAT II

For measurements performed on  

circuits directly connected to the  

low-voltage installation.

Appliances,portable 

equipment,etc.

CAT III

For measurements performed in  

the building installation.

Distribution board, 

circuit breaker,etc.

CAT IV

For measurements performed at  

the source of the low-voltage  

installation.

Overhead wire, 

cable systems,etc.

1. Components

1.1  Display (LCD)

1.2  Main unit (30032A)

1) Jaw Section

   Is a precision sensor for detecting currents.

2) Open/Close Lever

   Opens and closes the jaws.

3) Display

   Shows the measured value (digital reading or bar graph),  

unit, function and low-battery symbol ( 

 ), etc.

4) mA/A Selector Switch

   Selects the unit of AC current (either “mA” or “A”)  

to be measured.

5) Data Hold Switch

   Retains the measured data.  

If you press this switch, the 

 symbol appears and  

the data is retained.  

If you press this switch once again,  

data holding is canceled (the 

 symbol disappears).

6) Power Switch

   Turns on the power to the instrument. 

7) Filter Switch

   Turns on/off the filter. 

   Press the FILTER switch for more than 2 seconds.

   When the filter is on (used), 

 appears.

8) Battery Housing (Battery cover)

   Contains the battery.

9) Barrier

   Prevents contact with the wires.

Thank you for purchasing our Leakage Clamp-on Tester.

Before using this product, thoroughly read this manual  

to understand how to use it properly.

Contact information of Yokogawa offices worldwide is

provided on thefollowing sheet.

  PIM 113-01Z2:  Inquiries  List of worldwide contacts

1) Jaw

section

2) Open/

close lever

8) Battery

housing

(cover)

4) mA/A 

selector

switch

6) Power

switch

5) Data hold

switch

3) Display

Hand strap

9) Barrier

7) Filter

switch

2.  AUTO POWER OFF Function

2.1  When the AUTO POWER OFF Function is Used 

• AUTO OFF appears. 

• The instrument automatically turns off 10 minutes after 

 

the last switch operation. A buzzer starts to beep 15 seconds  

before the automatic power-off. 

• Pressing any switch after the beep restarts the auto power-off 

 

function.

2.2  When the AUTO POWER OFF Function is not Used 

(Cancellation of the AUTO POWER OFF Function) 

• Turn off the power. 

• With the DATA HOLD switch held down, press the POWER 

 

switch for more than 2 seconds to turn on the power.  

This causes the buzzer to sound, canceling the AUTO  

POWER OFF function (the AUTO OFF display goes off).  

If the instrument is used with the AUTO POWER OFF function  

cancelled, take care not to let the battery run down. 

2.3  Recovering the AUTO POWER OFF Function 

• Turn off the power. 

• Turn on the power. This enables the AUTO POWER OFF 

 

function to recover. (AUTO OFF appears.) 

3.  Measuring Instructions

3.1  Before measurement 

a)  Examine the casing, battery cover, display, and labels of  

the instrument for abnormalities. 

b)  Make sure that the battery cover is firmly closed. 

3.2  AC Current Measurement (unit: mA/A) 

a)  Press the POWER switch to turn on the power. 

b)  Squeeze the open/close lever to open the jaws.  

Insert a wire from the measurable conductors under test  

through the jaws, making sure the tops of the jaws are  

tightly shut. 

c)  When the reading stabilizes, read the value.  

If you have difficulties in reading the value, 

 

use the DATA HOLD function. 

d)  When measuring Load Current, press the mA/A switch to  

change to A range.

e)  When you have finished measurement, 

 

press the POWER switch to turn off the tester.

CAUTION

 

The correct measured values are not displayed if the jaws  

do not fit precisely. Make sure that they are shut tightly.

3.3  Using the DATA HOLD switch 

Pressing the DATA HOLD switch retains the measured data  

and displays 

 . 

The mA/A selector switch and the FILTER switch cannot be used  

in this situation. 

The only switches that can be used are the DATA HOLD switch 

(to cancel data holding) and the POWER switch.