4. Range Selection
4.1 Range Selection
Switching between the 3 mA and 30 mA ranges and between
the 30 A and 60 A ranges is performed by auto-ranging
(automatic).
For switching between the 3/30 mA and 30/60 A ranges,
the range must be switched using the mA/A selector switch.
4.2 Measurement Ranges
0.000 mA
0.010 mA<I
I
≤
2.88 mA
I
≥
3.270 mA
I
≥
32.70 mA
OL
3 mA range
30 mA range
I
≤
28.8 A
I
≥
32.70 A
I
≥
60.6 A
62.0 A MAX
OL
0.05 A<I
0.00 A
30 A range
60 A range
Zero correction
Zero correction
Range down
Range down
Range up
Range up
Overrange
Overrange
5. Battery Replacement
5.1 Battery Voltage
If the battery runs down and its voltage falls below the
operating voltage (effective range), the
symbol appears in
the LCD display; promptly replace the battery with a new one.
(Battery: CR2032, 1 piece)
CAUTION
• The specified accuracy is assured when the battery voltage
is in the effective range.
If the
symbol is displayed, promptly replace the battery.
• If the power does not come on even after pressing the
POWER switch, the battery voltage may have fallen too low.
Replace the battery with a new one.
5.2 Replacing the Battery
CAUTION
• Before replacing the battery, always disconnect the tester from
the measurable conductor under test because there is a risk
of electrical shock. After replacement, close the battery cover
securely and conduct measurements.
• Use the specified lithium battery (CR2032).
To replace the battery:
a) Press the POWER switch to turn off the power.
b) Turn the battery cover on the backside of the instrument in
the direction of the arrow using a coin, etc.
c) Remove the cover and the battery.
d) Insert a new battery, making sure that the polarities are
correct.
e) Close the cover back in place by turning it in the reverse
direction of the arrow.
Inserting a Battery
Closing the Battery Cover
Battery cover
Insert the battery being careful
that the polarities are correct.
Close the battery cover,
aligning the arrow's tail with
the dot and then turn clockwise.
6. Examples of Measurement
6.1 Examples of Leakage Current Measurement
Load
Load
Single-phase
Two-wire Circuit
Load
Grounding Wire
(1)
(3)
(2)
Single-phase Three-
wire or Three-phase
Three-wire Circuit
6.2 Example of Load Current Measurement
Load
Insert only one of the wires through the jaws.
7. Filtering Function
The instrument has a function for turning the filter on (use)/off.
Press the FILTER switch for more than 2 seconds to turn it on/off
(default: ON). When the filter is on, the
symbol appears.
When the FILTER switch is turned on, the sharp filter is enabled.
This filter attenuates 2nd-order and higher harmonic currents and
measures only the fundamental harmonics.
See the Filtering Characteristics below.
30032A Filtering characteristics
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Ampritude ratio
[d
B]
0
50
100
150
Frequency [Hz]
■ To Ensure Accurate Measurement
The maximum input current of the harmonic components in
the 3 mA and 30 mA ranges is 150 mA.
If the input exceeds 150 mA, the correct value is not displayed.
In some cases, “0L” does not appear and an error becomes
larger or readings do not change.
Therefore, follow the procedure below to check if the current
under measurement including 2nd-order and higher harmonic
components exceeds 150 mA.
Checking procedure:
1. Press the mA/A selector switch to choose the 30 A range.
2. Press the FILTER switch to select FILTER OFF and conduct
measurement.
3. Check that the measured value (reading) in the 30 A range is
0.15 A* or less.
4. When the measured value is 0.15 A or less, switch the range to
3 mA/30 mA and conduct measurement.
*: The accuracy of the 30 A range is 1.0% of r5 digits
(0.05A < I ≤ 50 A).
Five digits are equivalent to 0.05 A; when checking the current,
also consider the accuracy.
IM 30032A-E <P2>
8. Specifications (30032A)
8.1 Specifications (Accuracy)
Condition:
Temperature and humidity: 23 ˚C ±5 ˚C, 80% RH or less
Frequency: 50 Hz ±1.0 Hz, 60 Hz ±1.0 Hz
Battery voltage: Within the effective range
AC current measurement
Accuracy: ± (% of r digits)
Filter function OFF
Range Resolution
Accuracy
Maximum
Allowable Current
3 mA 0.001 mA
0.010 < I ≤ 32.70 mA:
1.0%+5
3.270 mA
30 mA
0.01 mA
32.70 mA
30 A
0.01 A
0.05 < I ≤ 50.0 A: 1.0% +5
32.70 A
60 A
0.1 A
50.0 < I ≤ 60.6 A: 5.0% +5
60.6 A
Filter function ON
Range Resolution
Accuracy
Maximum
Allowable Current
3 mA 0.001 mA
0.010 < I ≤ 32.70 mA:
1.5%+5
3.270 mA
30 mA
0.01 mA
32.70 mA
30 A
0.01 A
0.05 < I ≤ 50.0 A: 1.5% +5
32.70 A
60 A
0.1 A
50.0 < I ≤ 60.6 A: 5.5% +5
60.6 A
Note: Input current of 2nd-order and higher harmonics
: 150 mA rms maximum in the 3 mA/30 mA range
: 62 A rms maximum in the 30 A/60 A range
Filter specifications
(3 mA and 30 mA ranges and 30 A and 60 A ranges)
Amplitude ratio at 100 Hz: -38 dB (1.26%) or less (typical: -41 dB)
Amplitude ratio at 120 Hz: -53 dB (0.22%) or less (typical: -56 dB)
Zero correction
3 mA range: Displays 0.000 mA (zero) when 0.010 mA < I
30 A range: Displays 0.00 A (zero) when 0.05 A < I
8.2 General Specifications
• Method:
Mean-value detection and rms-value calibration
• Display (LCD): Digital reading
............... 3200 counts
Bar graph ...................... 32 segments
“OL” over-range indication
low-battery symbol
data hold symbol
“AUTO OFF” auto power off indication
filter ON
• Sampling:
Digital reading
............... 2 times/sec
Bar graph ...................... 12 times/sec
• Range:
3 mA, 30 mA, 30 A, 60 A
• Range selection:
Automatic (between 3 mA and 30 mA ranges and
between 30 A and 60 A ranges)
Manual (between 3 mA and 30 mA ranges and
between 30 A and 60 A ranges)
• Additional functions:
Data hold and Auto power-off
• Operating temperature and humidity range:
0 to 50°C, with a maximum humidity of 80% RH or less
(no condensation)
• Storage temperature and humidity range:
-20 to 60°C, with humidity range is 20 to 70%RH
(no condensation)
• Temperature coefficient:
The following values must be added in the temperature
range of either 0 to 18°C or 28 to 50°C.
0 ≤ I ≤ 50.0A: ± (0.08% of reading/˚C + 0.5 digits/˚C)
50.0 < I ≤ 60.6A: ± (0.3% of reading/˚C + 0.5 digits/˚C)
• Effect of external magnetic fields: 0.0005% typical value
(in terms of the magnitude of current in adjacent wires)
• Diameter of measurable conductors:
40 mm (Maximum)
• The RATED circuit-to-earth voltage: 300 Vrms or less
• Maximum allowed current: 62 Arms
• Withstanding voltage: 3.7 kV AC for one minute
(Tested between the cores of the jaw section and the case)
• Power consumption: 6 mW maximum
• Automatic power-off:
Automatically turns of the power about 10 minutes
after the last switch operation.
(The alarm buzzer begins beeping 15 seconds
before power-off.)
• Power supply: CR2032 lithium battery (cell)
.............1
• Battery life:
Approx. 90 hours (of continuous operation)
• Dimensions: Approx. 70 (W) x 178 (H) x 25 (D) mm
(excluding protrusions)
• Weight:
Approx. 200 g (including the battery)
• Accessories: Battery (housed in the instrument)
.....1
RB057 soft carrying case ...................1
User’s manual
Safety standards:
EN 61010-1, EN 61010-2-032
Measurement category III
(The RATED circuit-to-earth voltage: 300 Vrms AC)
Indoor use, 2000 m max. above sea level,
pollution degree 2
EMC standards:
EN 61326-1 Class B, EN 61326-2-2
EMC Regulatory Arrangement in Australia and
New Zealand
EN 55011 Class B Group 1
Korea Electromagnetic Conformity Standard
(한국 전자파적합성기준)
Environmental standard:
EN 50581
Monitoring and control instruments including
industrial monitoring and control instruments
9. Sales in Each Country or Region
9.1 Disposing the Product
Waste Electrical and Electronic Equipment (WEEE),
Directive
(This directive is valid only in the EU.)
This product complies with the WEEE directive
marking requirement.
This marking indicates that you must not discard
this electrical/electronic product in domestic
household waste.
Product Category
With reference to the equipment types in the WEEE directive,
this product is classified as a “Monitoring and control
instruments” product.
When disposing products in the EU, contact your local
Yokogawa Europe B.V. office.
Do not dispose in domestic household waste.
9.2 How to Replace and Dispose the Batteries
EU Battery Directive
(This directive is valid only in the EU.)
Batteries are included in this product.
When you remove batteries from this product and dispose them,
discard them in accordance with domestic law concerning
disposal.
Take a right action on waste batteries, because the collection
system in the EU on waste batteries are regulated.
Battery type: Lithium battery (cell)
Notice:
This marking indicates they shall be sorted out and
collected as ordained in the EU battery directive.
How to remove batteries safely:
For further details, see “5.2 Replacing the Battery.”
9.3 Authorized Representative in the EEA
Yokogawa Europe B.V. is the authorized representative of
Yokogawa Test & Measurement Corporation for this product in
the EEA. (EEA: European Economic Area)
To contact Yokogawa Europe B.V., see the separate list of
worldwide contacts, PIM 113-01Z2.