4.  Range Selection

4.1  Range Selection

Switching between the 3 mA and 30 mA ranges and between  

the 30 A and 60 A ranges is performed by auto-ranging  

(automatic). 

For switching between the 3/30 mA and 30/60 A ranges,  

the range must be switched using the mA/A selector switch.

4.2  Measurement Ranges

0.000 mA

0.010 mA<I

I

2.88 mA

I

3.270 mA

I

32.70 mA

OL

3 mA range

30 mA range

I

28.8 A

I

32.70 A

I

60.6 A

62.0 A MAX

OL

0.05 A<I

0.00 A

30 A range

60 A range

Zero correction

Zero correction

Range down

Range down

Range up

Range up

Overrange

Overrange

5.  Battery Replacement

5.1  Battery Voltage

If the battery runs down and its voltage falls below the  

operating voltage (effective range), the 

 symbol appears in  

the LCD display; promptly replace the battery with a new one. 

(Battery: CR2032, 1 piece) 

CAUTION

 

• The specified accuracy is assured when the battery voltage 

 

is in the effective range.  

If the 

 symbol is displayed, promptly replace the battery. 

• If the power does not come on even after pressing the 

 

POWER switch, the battery voltage may have fallen too low.  

Replace the battery with a new one. 

5.2  Replacing the Battery

CAUTION

 

• Before replacing the battery, always disconnect the tester from 

the measurable conductor under test because there is a risk 

of electrical shock. After replacement, close the battery cover 

securely and conduct measurements. 

• Use the specified lithium battery (CR2032). 

To replace the battery: 

a)  Press the POWER switch to turn off the power. 

b)  Turn the battery cover on the backside of the instrument in  

the direction of the arrow using a coin, etc. 

c)  Remove the cover and the battery. 

d)  Insert a new battery, making sure that the polarities are  

correct. 

e)  Close the cover back in place by turning it in the reverse  

direction of the arrow. 

Inserting a Battery

Closing the Battery Cover

Battery cover

Insert the battery being careful 

that the polarities are correct.

Close the battery cover, 

aligning the arrow's tail with 

the dot and then turn clockwise.

6.  Examples of Measurement

6.1  Examples of Leakage Current Measurement

Load

Load

Single-phase

Two-wire Circuit

Load

Grounding Wire

(1)

(3)

(2)

Single-phase Three-

wire or Three-phase 

Three-wire Circuit

6.2  Example of Load Current Measurement

Load

Insert only one of the wires through the jaws.

7.  Filtering Function

The instrument has a function for turning the filter on (use)/off. 

Press the FILTER switch for more than 2 seconds to turn it on/off  

(default: ON). When the filter is on, the 

 symbol appears. 

When the FILTER switch is turned on, the sharp filter is enabled. 

 

This filter attenuates 2nd-order and higher harmonic currents and 

 

measures only the fundamental harmonics. 

See the Filtering Characteristics below.

30032A Filtering characteristics

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Ampritude ratio 

[d

B]

0

50

100

150

Frequency [Hz]

■ To Ensure Accurate Measurement

The maximum input current of the harmonic components in  

the 3 mA and 30 mA ranges is 150 mA.

If the input exceeds 150 mA, the correct value is not displayed.  

In some cases, “0L” does not appear and an error becomes  

larger or readings do not change.

Therefore, follow the procedure below to check if the current  

under measurement including 2nd-order and higher harmonic  

components exceeds 150 mA.

Checking procedure:

1. Press the mA/A selector switch to choose the 30 A range.

2. Press the FILTER switch to select FILTER OFF and conduct  

measurement.

3. Check that the measured value (reading) in the 30 A range is  

0.15 A* or less.

4. When the measured value is 0.15 A or less, switch the range to  

3 mA/30 mA and conduct measurement.

*: The accuracy of the 30 A range is 1.0% of r5 digits  

(0.05A < I ≤ 50 A).

 

Five digits are equivalent to 0.05 A; when checking the current,  

also consider the accuracy.

IM 30032A-E <P2>

8.  Specifications  (30032A)

8.1  Specifications (Accuracy)

Condition: 

  Temperature and humidity: 23 ˚C ±5 ˚C, 80% RH or less

  Frequency: 50 Hz ±1.0 Hz, 60 Hz ±1.0 Hz

  Battery voltage: Within the effective range

AC current measurement

  Accuracy: ± (% of r digits)

  Filter function OFF

Range Resolution

Accuracy

Maximum 

Allowable Current

3 mA 0.001 mA

0.010 < I ≤ 32.70 mA: 

1.0%+5

3.270 mA

30 mA

0.01 mA

32.70 mA

30 A

0.01 A

0.05 < I ≤ 50.0 A: 1.0% +5

32.70 A

60 A

0.1 A

50.0 < I ≤ 60.6 A: 5.0% +5

60.6 A

  Filter function ON

Range Resolution

Accuracy

Maximum 

Allowable Current

3 mA 0.001 mA

0.010 < I ≤ 32.70 mA: 

1.5%+5

3.270 mA

30 mA

0.01 mA

32.70 mA

30 A

0.01 A

0.05 < I ≤ 50.0 A: 1.5% +5

32.70 A

60 A

0.1 A

50.0 < I ≤ 60.6 A: 5.5% +5

60.6 A

  Note: Input current of 2nd-order and higher harmonics

   

: 150 mA rms maximum in the 3 mA/30 mA range

   

: 62 A rms maximum in the 30 A/60 A range

Filter specifications

(3 mA and 30 mA ranges and 30 A and 60 A ranges)

Amplitude ratio at 100 Hz: -38 dB (1.26%) or less (typical: -41 dB)

Amplitude ratio at 120 Hz: -53 dB (0.22%) or less (typical: -56 dB)

Zero correction

3 mA range: Displays 0.000 mA (zero) when 0.010 mA < I

30 A range:  Displays 0.00 A (zero) when 0.05 A < I

8.2  General Specifications

• Method: 

Mean-value detection and rms-value calibration

• Display (LCD):  Digital reading

 ............... 3200 counts

 

Bar graph ...................... 32 segments

   

 

“OL” over-range indication

    

 low-battery symbol

    

 data hold symbol

   

 

“AUTO OFF” auto power off indication

    

  filter ON

• Sampling: 

Digital reading

 ...............   2 times/sec

 

Bar graph ...................... 12 times/sec

• Range: 

3 mA, 30 mA, 30 A, 60 A

• Range selection: 

   

Automatic (between 3 mA and 30 mA ranges and  

   

between 30 A and 60 A ranges)

   

Manual (between 3 mA and 30 mA ranges and  

   

between 30 A and 60 A ranges)

• Additional functions: 

Data hold and Auto power-off

• Operating temperature and humidity range: 

   

0 to 50°C, with a maximum humidity of 80% RH or less 

 

   

(no condensation)

• Storage temperature and humidity range: 

   

-20 to 60°C, with humidity range is 20 to 70%RH  

   

(no condensation)

• Temperature coefficient: 

   

The following values must be added in the temperature  

   

range of either 0 to 18°C or 28 to 50°C.

   

0 ≤ I ≤ 50.0A: ± (0.08% of reading/˚C + 0.5 digits/˚C)

   

50.0 < I ≤ 60.6A: ± (0.3% of reading/˚C + 0.5 digits/˚C)

• Effect of external magnetic fields: 0.0005% typical value 

 

   

(in terms of the magnitude of current in adjacent wires)

• Diameter of measurable conductors: 

 40 mm (Maximum)

• The RATED circuit-to-earth voltage: 300 Vrms or less

• Maximum allowed current: 62 Arms

• Withstanding voltage: 3.7 kV AC for one minute 

 

  (Tested between the cores of the jaw section and the case) 

• Power consumption: 6 mW maximum 

• Automatic power-off: 

   

Automatically turns of the power about 10 minutes  

   

after the last switch operation. 

   

(The alarm buzzer begins beeping 15 seconds  

   

before power-off.)

• Power supply:  CR2032 lithium battery (cell)

 .............1

• Battery life: 

Approx. 90 hours (of continuous operation) 

• Dimensions:  Approx. 70 (W) x 178 (H) x 25 (D) mm 

   

 

(excluding protrusions) 

• Weight: 

Approx. 200 g (including the battery)

• Accessories:  Battery (housed in the instrument)

 .....1

 

RB057 soft carrying case ...................1

 

User’s manual

Safety standards: 

 

EN 61010-1, EN 61010-2-032

 

Measurement category III

 

(The RATED circuit-to-earth voltage: 300 Vrms AC)

 

Indoor use, 2000 m max. above sea level, 

 

pollution degree 2

EMC standards: 

 

EN 61326-1 Class B, EN 61326-2-2 

 

EMC Regulatory Arrangement in Australia and  

 

New Zealand

 

EN 55011 Class B  Group 1

 

Korea Electromagnetic Conformity Standard

 

(한국 전자파적합성기준)

Environmental standard: 

 

EN 50581

Monitoring and control instruments including 

industrial monitoring and control instruments

9.  Sales in Each Country or Region

9.1  Disposing the Product

Waste Electrical and Electronic Equipment (WEEE), 

Directive

(This directive is valid only in the EU.)

This product complies with the WEEE directive  

marking requirement.

This marking indicates that you must not discard  

this electrical/electronic product in domestic  

household waste.

Product Category

With reference to the equipment types in the WEEE directive,

this product is classified as a “Monitoring and control

instruments” product.

When disposing products in the EU, contact your local

Yokogawa Europe B.V. office.

Do not dispose in domestic household waste.

9.2  How to Replace and Dispose the Batteries

EU Battery Directive

(This directive is valid only in the EU.) 

Batteries are included in this product. 

When you remove batteries from this product and dispose them,  

discard them in accordance with domestic law concerning  

disposal.

Take a right action on waste batteries, because the collection  

system in the EU on waste batteries are regulated.

Battery type: Lithium battery (cell)

Notice:

This marking indicates they shall be sorted out and  

collected as ordained in the EU battery directive.

How to remove batteries safely:

For further details, see “5.2 Replacing the Battery.”

9.3  Authorized Representative in the EEA

Yokogawa Europe B.V. is the authorized representative of  

Yokogawa Test & Measurement Corporation for this product in  

the EEA. (EEA: European Economic Area)
To contact Yokogawa Europe B.V., see the separate list of  

worldwide contacts, PIM 113-01Z2.