10
Press
key determines the current settings and returns.
(c) Press
key to enter into next parameter setting or press
key to return
to measuring mode.
4.4.3. Resolution selection (P2)
(a) Press
key in the mode P1 choose P2 , press
key enters into mode
P2, see picture (4-3).
(b) Press
or
to choose resolution :
0.1→0.01;
Press
key determine the choice.
(c) Press
key to enter into next parameter setting or press
key to
return to measuring mode.
4.4.4. pH standard buffer solution selection (P3)
(a) press
key in mode P2 to choose mode P3,press
key enters into
mode P3 see picture (4-4).
(b) Press
key or
change buffer solution selection;there are Europe &
U.S.A series,NIST series and China series;press
key determines the current
settings and returns;
(c) press
key enter into next parameter setting or press
key to
return to measuring mode.
Three series standard buffer solution values:
CH(China series): 1.68ph 4.00ph 6.86ph 9.18ph 12.46ph
NIS(NIST series):1.68ph 4.00ph 6.86ph 9.18ph 12.46ph
USA(Europe & U.S.A series):1.68ph 4.00ph 7.00ph 10.01ph 12.46ph