Positive and negative pole installation

10.2 Temperature measurement steps

1. Point the thermometer at the center of the forehead(above 
    the center of the eyebrow) and keep it vertical, about 3-5 
    cm away. Press the measurement switch and the 
    temperature will be displayed immediately.

■ Note for battery installation and replacement: 

1) After the battery cover is opened, place the battery in the 
     direction of the positive and negative electrodes at the place 
     where the batteries are installed. Pay attention to the correct 
     orientation of the positive and negative electrodes.
2) Wait for 10 minutes to warm up for the first time or immediately 
      after inserting a new battery.
 3) When the battery power is low, the symbol will appear on the 
     display, prompting to replace the new battery, open the battery 
     compartment cover (see 6 product structure), pay attention to 
     the positive and negative polarity when replacing the new 
     battery. Improper placement may cause damage to the product.
 4) Remove the battery when it has not been used for a long time 
     to prolong its service life and prevent damage due to battery 
     leakage. 

2. When the measurement switch is pressed, the current 
measurement temperature value is displayed on the display, 
and the measurement data can be automatically stored Note:

1) Before measuring, make sure that there is no hair, sweat, makeup 

    or hats on it.

2) When forehead sweating or other reasons cause the forehead 

     temperature does not reflect the body temperature normally, please 

     measure after aiming at the earlobe, and make sure that there is no 

     hair, sweat, cosmetics or hat covering.

3) The non-contact infrared forehead thermometer has not been used 

    for a long

    time, the first time the machine is opened for environmental 

    temperature detection, and the boot time is extended by 1 to 2 

    seconds.

4) A person's body temperature changes at different times of the day 

    and is also affected by other external conditions, such as age, 

    gender, and skin color.

06

07

Содержание JT-E020

Страница 1: ...20 Yichang Jiangjing Optoelectronic Co Ltd Manufacturer After sales service Yichang Jiangjing Optoelectronic Co Ltd After sales service Tel 86 0717 299555 Production Address No 66 Changning Third Road Maoping Town Zigui County Yichang City Hubei Province China ...

Страница 2: ...part is more affected by the ambient temperature 2 Use safety manual Please read this manual carefully before use The operating temperature of this product is 10 40 the optimal temperature is 25 Do not use this product in an environment above 40 or below 0 Do not place this product near charged objects to avoid electric shock Please do not use this product in an environment with relative humidity ...

Страница 3: ...an forehead It is widely used in home use This product cannot replace the diagnosis of doctors 7 LCD display description 5 Precautions Before Use Power On Self Test Aim at the measured object press the measurement switch the LCD will display all the numbers and characters of the self test and the self test screen appears This screen is displayed for about 1 second 6 Product Structure Model JT E020...

Страница 4: ...elative humility 93 Non corrosive gas well ventilated room The transportation requirements are stipulated in the order contract and the severe impact vibration and rain and snow splashing during transportation must be prevented 9 Technical index 9 1 Basic parameters 8 2 Product contraindications 1 Conditions such as birth defects congenital malformations septic shock and circulatory failure may se...

Страница 5: ...ve polarity when replacing the new battery Improper placement may cause damage to the product 4 Remove the battery when it has not been used for a long time to prolong its service life and prevent damage due to battery leakage 2 When the measurement switch is pressed the current measurement temperature value is displayed on the display and the measurement data can be automatically stored Note 1 Be...

Страница 6: ...select Fahrenheit Short press the setting key to save the current selection and enter the next option long press 3 seconds to save the current setting and exit the setting function 10 3 5 Warning threshold setting the back lit displays red color when the measurement exceeds this threshold Press the setting key to select the LCD display as F2 Press the key to modify the alarm temperature value the ...

Страница 7: ...biological comprehensive system Body temperature is an important data indicating whether human life activities are normal Usually we measure our health by measuring the temperature of the forehead cochlea anus mouth and armpit It is measured in different parts Body temperature will vary please refer to the table below for specific differences 10 3 7 Error correction settings Press the setting key ...

Страница 8: ... it has exceeded the measurement range or the temperature is higher than 42 9 in human temperature measurement mode 2 LCD display message Hi Using a non contact electronic body temperature timer the LCD displays the message Lo indicating that the measurement temperature is lower than the measurement range or the temperature is lower than 32 C in the human body temperature measurement mode 3 LCD di...

Страница 9: ...ference when using such as near mobile phones microwave ovens etc Guidelines and manufacturer s declarations are detailed in the attachment The equipment or system should not be used close to or stacked with other equipment If it must be used close to or stacked it should be observed to verify that it can operate normally in the configuration in which it is used 1 thermometer 1 manual 1 certificat...

Страница 10: ... immunity NOTE UT is the a c mains voltage prior to application of the test level The JT E020 is intended for use in the electromagnetic environment specified below The customer or the user of the JT E020 should ensure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Portable and mobile RF communications equipment shoul...

Страница 11: ... considered If the measured field strength in the location in which the JT E020 is used exceeds the applicable RF compliance level above the JT E020 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the JT E020 b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m...

Отзывы: