Installing and Connecting the Product
22
TUNICS T100S-HP User Manual
2.2
Connecting the TUNICS T100S-HP to a Power Source
Before Starting
Procedure
1.
Make sure the power adapter is not plugged to the wall socket.
2.
Make sure the power switch is set to
0
.
3.
Connect one end of the power supply cable to the rear panel of the product and plug
the other end to the proper voltage mains supply point. The TUNICS T100S-HP is
equipped with a self-regulating power supply that adapts to both AC 110V and 230V
voltages, and 50 Hz or 60 Hz frequencies.
2.3
Connecting External Optical Devices to the product
Subject
You can connect external optical components and instruments to the TUNICS T100S-HP.
Before Starting
Procedure
1.
Make sure fiber-optic connectors are clean.
2.
On the front panel, in the OUTPUT area, remove the protective cap and connect
external optical components and instruments to the TUNICS T100S-HP via the FC/
APC fiber-optic connector.
•
Make sure the product power source does not apply more than 265
Volts RMS between the supply conductors or between either of the
supply conductors and the ground.
•
To avoid the possibility of injury, make sure the socket outlet in which
the power supply cord will be plugged is equipped with a protective
ground contact, and that the electrical installation fulfills the local safety
requirements.
•
The TUNICS T100S-HP has a chassis connected to ground via the power
supply cable. A protective ground connection by way of the grounding
conductor in the power cord is essential for safe operation.
WARNING
Keeping fiber-optic connectors clean at all times is essential to achieve
optimum system performance. Refer to the cleaning instructions in section
Cleaning and Caring for Optical Connectors and Fibers, p. 55
. The standard
output connector is a FC/APC connector. Never connect another type of
connector to the optical output.
CAUTION
•
The use of optical instruments with this product will increase eye
hazard. Do not under any circumstances look directly into the fiber end
of an optical cable attached to the optical output while the device is in
use for this may cause permanent eye damage and possible loss of
eyesight. Note that the laser radiation is not visible to the human eye,
therefore, protective cap must always be replaced on the laser output
connector after use to avoid involuntary exposure to laser radiation.
•
Disable the laser output before connecting or disconnecting a fiber
optic cable on the product.
WARNING
•
To avoid damage to the laser output, always use a FC/APC type
connector. To connect other types of connectors such as FC/PC, make
sure to use the appropriate FC/APC converter.
•
To help protect the laser output, leave a fiber-optic cable connected to
the product while making connections to other devices. When the laser
output is not connected, always install the protective cap.
CAUTION
Содержание TUNICS T100S-HP
Страница 2: ......
Страница 6: ...About This Manual 6 TUNICS T100S HP User Manual...
Страница 24: ...Installing and Connecting the Product 24 TUNICS T100S HP User Manual...
Страница 36: ...Changing the Optical Output Power and Diode Current 36 TUNICS T100S HP User Manual...
Страница 50: ...Using the Auxiliary Inputs and Outputs 50 TUNICS T100S HP User Manual...
Страница 54: ...Verifying Performances 54 TUNICS T100S HP User Manual...
Страница 60: ...Certification and Compliance 60 TUNICS T100S HP User Manual...
Страница 66: ...Index 66 TUNICS T100S HP User Manual...