other equipment should be observed to verify that they are operating
normally.
2* WARNING: Use of accessories, transducers and cables other than
those specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper
operation.
3* WARNING: Portable RF communications equipment (including
peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the ME
equipment, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could
result.
Guidance and manufacture’s declaration – electromagnetic emission
The HX-400 is intended for use in the electromagnetic environment specified below. The
customer or the user of the HX-400 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The HX-400 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The HX-400 Pupilometer is suitable for
use in all establishments, including
domestic establishments and those
directly connected to the public
low-voltage network that supplies
building used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable