![Yamato WA500 Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/yamato/wa500/wa500_instruction-manual_905421037.webp)
-34-
Safety Devices and Error Codes
Displays and Details of Error Code
Safety device
Indicators
Indication
Cause
Effect on operation
Measure
Detecting
leakage
Indicator(1)
is lit
Water
leakage
Heaters, solenoid valves
and any other controllers
become off.
Turn the bleaker off
and check parts.
(See P36 for
troubleshooting)
Heater
overheat
protection
is lit
Heater
overheat
Heaters, solenoid valves
and any other controllers
become off.
Contact the seller or
Yamato Scientific
’s
Technical Service
Department.
Detecting
trouble
of
float switch
is lit
Trouble of the
float
switch
for
distilled
water tank
Heaters, solenoid valves
and any other controllers
become off.
Contact the seller or
Yamato Scientific
’s
Technical Service
Department.
Detecting
abnormal
water level
is lit
Abnormal
water level in
the
float
receptacle
Heaters solenoid valves
and any other controllers
become off.
Check if the boiler
water drain cock is
Indicated with the
cock closed, contact
the seller or Yamato
Sientific
’s Technical
Service Department.
Detecting
float switch
trouble
is lit
Trouble of the
float
switch
for
float
receptacle
Heaters solenoid valves
and any other controllers
become off.
Contact the seller or
Yamato Scientific
’s
Technical Service
Department.
Detecting
trouble of the
electric circuit
is lit
Trouble in the
electric circuit
Heaters, solenoid valves
and any other controllers
become off.
Contact the seller or
Yamato Scientific
’s
Technical Service
Department.
●
”TROUBLE” blinks to alarm abnormal situation while the indicator (1) shows a certain error code.