Operation Method
Apparatus/Liquid Sterilization Course
Follow the procedures below for the setting of apparatus sterilization and liquid sterilization courses.
Turn the cover CCW and separate the upper
and lower handles completely (Figure 1).
Remove the cover and place it on a flat
surface (Figure 2). When placing the cover,
the pressure gauge should be placed upward
to avoid collision and deformation with the
case, resulting in the error of the pressure
gauge pointer.
Connect to the external power supply of AC
220V±10%, 50Hz±1Hz and power
≥
2KW,
plug into the sterilizer special power socket,
turn on the power switch to the ON position
(Figure 3), the required setting can be made
on the control panel (an electromagnetic
switch is set in the handle, and the heater will
not work when the cover is not closed or
there is water shortage in cylinder). The
blinking water level lamp indicates that there
is water shortage in the cylinder (Figure 4),
and the display screen (Figure 4) indicates
the current temperature in the cylinder.
3. Take out the inner barrel, inject the tap water
directly into the cylinder and add water until
the water level indicator lamp on the control
panel is normally ON (Figure 5). If there is too
much water in the cylinder, turn on the drain
valve (Figure 7) to drain the excess water (the
standard water level should reach the lower
edge of the three brackets in the cylinder
(Figure 6). Turn the drainage knob to the OFF
position after the water level is confirmed
(Figure 8).
Figure 1
Figure 2
Figure 3
Figure 5
Figure 6
Figure 7
Figure 8
Temp. in cylinder
Inner barrel
Bracket*3
Heating
pipe
Standard
water level
Liquid level
sensor
Temp. sensor
Figure 4
NO
OFF