12
Handling Precautions
CAUTION!
Sample installation method
The sample container mounting holders for separating funnel, Erlenmeyer flask, centrifugation
tube and test tube are provided respectively as nonstandard options. Select them according to
the purpose, and mount them on the shaking table of the main unit.
To shake the separating funnel, set the shaking table at vertical position (vertical shaking type),
and enter the positioning pins of the holder into the two pipe holes provided in the table (see the
sketch drawing), and then fix them firmly.
When using Erlenmeyer flask, set the vinyl chloride box type holder firmly on the table of the
main unit placed at level (horizontal shaking type) (See the illustration on Page 7).
Also, when using centrifugation tube and test tube, place the main unit horizontal.
Preparing timer
This unit incorporates a 60-min meter for timer. If it is necessary to operate this unit for a longer
time, set the timer to CONT. Continuous operation is available.
Return after power failure
When power is supplied after a power failure, the device automatically starts operation again with
the same state as just before the power failure. It is danger that the device starts unattached
operation after a power failure. We recommend for you to turn off the switch of this unit if a
power failure occurs during operation.
When the main unit vibrate using it
Please use it as follows when the main unit vibrate or begin.
It is considered becoming unbalance of the unit to use for a long time.
①
Decrease the shaking time not to vibrate it.
②
Decrease the weight and not to vibrate position.
③
If not to decrease the shaking time or the weight, please fix the unit with anchor.