9
Operation Method
Sequence of Operation
1. Fix the sample containers.
•
Before turning on the power switch, fix the sample containers firmly.
•
The quantity of sample containers to be installed is as shown in the
following table.
100 200 300 500 1000
Separating
Funnel
Holder
5 4 4 4 3
Erlenmeyer Flask Holder
25 16 16 9 4
2. Turn on the power.
•
Before turning on the power switch, check if the timer knob is OFF
and if the speed adjusting knob is at the lowest speed.
•
When the power switch is turned on, the power lamp is ON.
3. Set the shaking time.
•
When the sample container is installed, turn the timer dial, and set it
to the required time (min).
•
In continuous operation, adjust the dial position to "CONT".
4. Adjust the shaking times.
•
Shaking times shall be set by the speed adjusting knob.
Since the scale has its variable range divided evenly, use this for
reference.
CAUTION!
If used at the max shaking times, such shaking times could exceed the
standard value in the area of commercial power frequency 60 Hz. If
shaking is performed in the too wide range, or if the main unit vibrates
widely, reduce the speed-regulating dial.
Shaking times could also rise according to the condition of the load. In
such cases, slightly reduce the speed-regulating dial similarly.
Shaking times for each scale (reference) are as shown below
approximately:
2 4 6 8
Max.
SA300 Start
115
185
250
300
SA320 Start
95
140
180
210
Capacity (mL)
Holder
Scale
Model