3. ASSEMBLY & SETUP
Cooling Condenser
V. Condenser retainer ring, coil ring and vacuum seal placement.
1. Remove the condenser retainer ring, coil ring and
vacuum seal, if they have not already been removed
from the left side (side tilted upward) of the motor.
2. Place the condenser retainer ring, then the coil
ring over the condenser mouth flange and place
the vacuum seal over the mouth with the seal
protruding inward, in whichever arrangement is
applicable (see images below).
* Apply silicon vacuum grease to rippled side of
vacuum seal, if desired.
Condenser Type A
Condenser Type B/C
VI. Install condenser.
While inserting the steam duct into the cooling condenser, place seal against the metal flange on motor
assembly and tighten the condenser retainer ring securely.
Be careful not to damage the vacuum seal when inserting the steam duct.
Condenser Type A
Condenser Type B/C
●
Removing coil ring from condenser flange
Grab the coil ring with the supplied Allen wrench, as shown in image below, in
order to remove the ring more easily.
Apply moderate pressure when removing ring. Using excessive force may
damage coil ring and/or glass components.
Coil Ring
Vacuum Seal
Coil Ring
Vacuum Seal