YCO-1081-07-01
WARNING AND ERROR INDICATION
The scale can check various functions automatically by itself. When a problem is detected, the following
messages will be displayed. If you take the steps described below but the problem is not fixed, the
scale needs to be serviced by a qualified scale technician.
1) The batteries are running low.
When the batteries are running low, this indicator will come on.
Replace all of the batteries with new ones.
2) The batteries are low.
When the batteries are too low, this indicator will be displayed and
you cannot use the scale until you replace the batteries.
3) The weight out of range.
If the item is 6 or more divisions heavier than the scale’s capacity
this message will be displayed.
When the weighed value is 6 or more divisions (increments)
below zero, this message will be displayed. It is usually displayed
when the platform is removed.
4) Initial error on startup
When the initial weight is heavier than the permissible zeroing
range and the
key is pressed, this message will
be dis-
played. Remove whatever is on the platform and press the
key again.
When the initial weight is lighter than the permissive zeroing
range and the
key is pressed, this message will be dis-
played. Install the platform or remove any objects under the
platform, and press the
key again.
5) The weighing sensor error
When the weight detecting section has a problem, this message
will
be displayed. Turn off the power. Then turn the power on
again.
6) The electronic circuitry error
When the electronic circuit has a problem, this message may be
displayed. Turn off the power. Then turn the power on again.
7) The electromagnetic disturbance error
This Indicator is occasionally displayed due to strong static
electricity and/or electro-magnetic disturbance, but it is not a
failure. Turn off the power. Then turn the power on again.
PPC-101
Operation Manual
5