TAKING MEASUREMENTS
5. Turbidity Measurements
The turbidity of the solution is measured as the O.D. (Optical Density) of the R sensor.
The procedure is as follows:
(
Example of E.coli culture)
1. Run calibration;
2. Set the blank solution (Culture solution without E.coli), run blank measurement;
3. Set the cultured E.coli and measure.
Tap the
Turbidity Measure Button
on the top screen to move to the Turbidity Measurement Screen
(Illustration 5-1).
Illustration 5-1: Turbidity Measurement Screen
③
Measurement
Button
Operation is possible
after the Blank Measurement.
⑦
Clear Button
⑥
Conversion Factor
Input Sectinon
<P31>
②
Blank Measurement
Button
Red
:
unexecuted
Blue
:
executed
Operation is possible
after the Calibration.
①
Calibration Button
Red
:
unexecuted
Blue
:
executed
④
CSV File Button
⑤
Title Input Section
TAK
IN
G
ME
ASU
REME
NTS
29