- 23 -
8. TROUBLESHOOTING
Troubleshooting Table
Symptom
Possible Causes
Temperature controller light does not
illuminate.
Power cable is not properly connected to outlet.
Power failure in progress.
Main power switch (ELB) is turned
“OFF”
Temperature does not fall.
Condenser fins need cleaning.
Thermal capacity of coolant has been exceeded.
External temperature exceeds 35°C.
Intake or heat vents obstructed.
Power supply voltage has dropped below rating.
Refrigeration system will not restart.
Refrigeration system is in overload. Turn off refrigeration
switch immediately, see possible causes for
“temperature
does not fall
”, above, and turn on refrigeration system again
after several minutes.
Coolant fluid does not circulate.
Outlet stop valve is closed.
Circulation path is closed, restricted or otherwise obstructed.
Coolant fluid is too heavy or viscous.
Error Table
Error Sign
Cause
Remedy
UUUU
Sensor error
Check whether sensor is disconnected or severed.
FALR
Temperature controller
malfunction
Terminate operation. Turn off power immediately, and call
for service
If problems persist, turn off power, disconnect power cable and call for service.