![Yamato IC103C Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/yamato/ic103c/ic103c_manual_905240029.webp)
27
* You can set either of + or – side for the offset compensation temperature.
When compensation is set for the – side, the measured temperature display decreases by the
compensation temperature while the temperature in the chamber increases by the same
amount.
When compensation is set for the + side, the measured temperature display increases by the
compensation temperature while the temperature in the chamber decreases by the same
amount.
* Since too large a compensation value may result in larger difference between the actual and
indicated temperatures and may present a danger, consult our nearest sales office before
entering a large compensation value.
* The device has, in addition to the calibration offset function, the two-point compensation function
that adjusts offset for the lower temperature range and higher temperature range, for which
adjustment temperatures have been input on shipping from the factory.
* Consult the nearest sales office before attempting validation work for the temperature adjusting
device.